Agreement of ocular biometry measurements between 2 biometers.

Elizabeth Yeu1.   

Abstract

PURPOSE: To establish the agreement of routine measurements taken before cataract surgery between the Aladdin biometer, which is based on optical low-coherence interferometry (OLCI), and the Lenstar biometer, which is based on optical low-coherence reflectometry (OLCR).
SETTING: Virginia Eye Consultants, Norfolk, Virginia, USA.
DESIGN: Prospective case series.
METHODS: Biometry parameters in the same eye were measured using the OLCI biometer and the OLCR biometer at separate preoperative visits. Measured parameters included axial length (AL), anterior chamber depth (ACD), keratometry (K) values (flat K, steep K, steep K axis), lens thickness, central corneal thickness (CCT), and the white-to-white (WTW) distance. Adverse events were monitored.
RESULTS: Data were collected for 101 eyes (65 right eyes) of 101 patients (mean age 67.9 years ± 7.2 [SD]). The mean values measured by the OLCI biometer and OLCR biometer, respectively, were as follows: AL, 23.88 ± 1.23 mm and 23.89 ± 1.23 mm; ACD, 3.24 ± 0.36 mm and 3.25 ± 0.37 mm; flat K, 43.63 ± 1.65 diopters (D) and 43.61 ± 1.67 D; steep K, 44.59 ± 1.76 D and 44.61 ± 1.77 D; lens thickness, 4.61 ± 0.39 mm and 4.45 ± 0.45 mm; CCT, 0.55 ± 0.04 mm and 0.55 ± 0.04 mm; and WTW distance, 11.63 ± 0.41 and 12.03 ± 0.49 mm. There was a high correlation between all parameter measurements obtained by the 2 devices. No adverse events occurred.
CONCLUSION: The measurements of the OLCI biometer were as reliable and reproducible as those of a leading biometer on the market, with the former having the advantage of an incorporated full corneal topographer.
Copyright © 2019. Published by Elsevier Inc.

Entities:  

Year:  2019        PMID: 31279621     DOI: 10.1016/j.jcrs.2019.03.016

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  1 in total

1.  Repeatability of automated measurements by a new anterior segment optical coherence tomographer and biometer and agreement with standard devices.

Authors:  Domenico Schiano-Lomoriello; Kenneth J Hoffer; Irene Abicca; Giacomo Savini
Journal:  Sci Rep       Date:  2021-01-13       Impact factor: 4.379

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.