| Literature DB >> 31278546 |
Andrea Coppadoro1, Nilde Eronia1, Giuseppe Foti1,2, Giacomo Bellani3,4,5.
Abstract
Electrical impedance tomography (EIT) is used for bedside ventilation monitoring; cardiac related impedance changes represent a source of noise superimposed on the ventilation signal, commonly removed by low-pass filtering (LPF). We investigated if an alternative approach, based on an event-triggered averaging (ETA) process, is more effective at preserving the actual ventilation waveform. Ten paralyzed patients undergoing volume-controlled ventilation were studied; 30 breaths for each patient were identified to compare LPF and ETA. For ETA the identified breaths were temporally aligned on the beginning of inspiration; the values of the thirty curves at each time point were averaged. The analysis was conducted on the global EIT signal and on four ventral-to-dorsal regions of interest. Global tidal variations by ETA resulted higher than LPF (average difference 139 ± 88 arbitrary units, p = 0.004). Both for global and regional waveforms, minimum and maximum EIT slopes were steeper by ETA as compared to LPF (average difference respectively - 57 ± 60 mL/s and 144 ± 96 mL/s for global signal, p < 0.05); ventilator inspiratory peak airflow correlated with maximum slope measured by ETA (r = 0.902, p < 0.001), but not LPF (p = 0.319). Beginning of inspiration identified on the ventilator waveform and on the global EIT signal by ETA occurred simultaneously, (+ 0.04 ± 0.07 s, p = 0.081), while occurred earlier by LPF (- 0.26 ± 0.1 s, p < 0.001). Removal of cardiac related impedance changes by ETA results in a ventilation signal more similar to the waveforms recorded by the ventilator, particularly regarding the slope of impedance changes and time at the minimum values as compared to LPF.Entities:
Keywords: Cardiac related impedance changes removal; EIT filtering; Electrical impedance tomography; Event-triggered average; Low-pass filter
Mesh:
Year: 2019 PMID: 31278546 PMCID: PMC7223993 DOI: 10.1007/s10877-019-00348-2
Source DB: PubMed Journal: J Clin Monit Comput ISSN: 1387-1307 Impact factor: 2.502
Fig. 1a Example of event-triggered averaging of EIT global impedance waveforms. Positive airflow at the ventilator was identified as the trigger event (dashed arrow) for all the selected breaths (light black lines); the resulting averaged waveform is depicted as a thick black line. b Exemplary low-pass waveform where the study variables were identified: EIT Minimum and Maximum values (Min, Max); time of minimum and maximum (Tmin, Tmax); minimum and maximum slope (Min and Max slope)
Fig. 2a Maximum slope (expressed in mL/s) of the global EIT signal and the 4 ROIs event-triggered averaging resulted higher than low-pass filtering (all p < 0.05). b Minimum slope (expressed in mL/s) of the global EIT signal and the 4 ROIs event-triggered averaging resulted lower than low-pass filtering (all p < 0.05 except ROI1)
Fig. 3Ventilator inspiratory flow and maximum slope (expressed in mL/s) of the global EIT signal measured by event-triggered averaging were closely related (p = 0.004, r = 0.821), while no significant correlation was present with maximum slope measured by low-pass filtering
Fig. 4Time of the minimum values (Tmin) of the global EIT signal and the 4 ROIs event measured by low-pass filtering occurred earlier than event-triggered averaging and earlier than ventilator beginning of inspiration (zero reference line, all p < 0.05). Tmin by event-triggered averaging did not differ from ventilator beginning of inspiration
Fig. 5Ventilator inspiratory time showed a tight correlation with EIT inspiratory time of the global EIT signal measured by event-triggered averaging (p < 0.001, r = 0.905), while no significant correlation was present with EIT inspiratory time measured by low-pass filtering