Sei Won Lee1, So Youn Shin2,3, Tai Sun Park1, Yoon Young Choi4, Jong Chun Park5, Jina Park1, Sang Young Oh2, Namkug Kim2, Se Hee Lee1, Jae Seung Lee1, Joon Beom Seo2, Yeon Mok Oh1, Sang Do Lee1, Sang Min Lee6.
Abstract
OBJECTIVE: The absence of collateral ventilation (CV) is crucial for effective bronchoscopic lung volume reduction (BLVR) with an endobronchial valve. Here, we assessed whether CT can predict the Chartis™ results.Entities:
Keywords: Chronic obstructive pulmonary disease; Emphysema; Tomography; X-ray computed
Mesh:
Year: 2019 PMID: 31270985 PMCID: PMC6609435 DOI: 10.3348/kjr.2018.0724
Source DB: PubMed Journal: Korean J Radiol ISSN: 1229-6929 Impact factor: 3.500
Fig. 1Semi-automatic lobe segmentation and fissure assessment.
A. Axial CT image with semi-automatic lobe segmentation result of left lung (green: left upper lobe, yellow: left lower lobe). B, C. Three-dimensional volume rendering image of left lung (green: left upper lobe, yellow: left lower lobe, red: left major fissure).
Fig. 2Flow chart of study patients.
BLVR = bronchoscopic lung volume reduction, CV = collateral ventilation, EBV = endobronchial valve
Baseline Characteristics
| Characteristics | Group 1 (n = 69) | Group 2 (n = 50) | Group 3 (n = 35) |
|---|---|---|---|
| Demographic characteristic | |||
| Age, years | 70.9 ± 6.6 | 71.1 ± 6.7 | 70.8 ± 7.4 |
| Male sex | 66 (95.7) | 47 (94.0) | 32 (91.4) |
| Body mass index (kg/m2) | 19.3 ± 3.1 | 18.8 ± 2.7 | 18.4 ± 3.0 |
| Smoking history | |||
| Former smoker | 62 (89.9) | 44 (88.0) | 30 (85.7) |
| Current smoker | 7 (10.1) | 6 (12.0) | 5 (14.3) |
| No. of pack years | 45.2 ± 21.7 | 46.2 ± 23.1 | 43.2 ± 15.9 |
| Medication use | |||
| Long-acting bronchodilator only | 17 (24.6) | 8 (16.0) | 4 (11.4) |
| Long-acting bronchodilator and inhaled corticosteroid | 49 (71.0) | 40 (80.0) | 30 (85.7) |
| Short-acting bronchodilator only | 3 (4.3) | 2 (4.0) | 1(2.9) |
| Lung function | |||
| Post-bronchodilator FEV1, L (% pred.) | 0.79 ± 0.38 (27.3 ± 11.9) | 0.76 ± 0.28 (26.7 ± 10.7) | 0.74 ± 0.29 (27.0 ± 11.6) |
| Post-bronchodilator FVC, L (% pred.) | 2.79 ± 0.76 (69.5 ± 17.2) | 2.81 ± 0.82 (70.8 ± 18.0) | 2.75 ± 0.85 (70.2 ± 18.9) |
| Carbon monoxide diffusing capacity, mL carbon monoxide/min/mmHg (% pred.) | 6.39 ± 2.95 (33.0 ± 12.6) | 6.23 ± 2.21 (32.1 ± 10.9) | 6.17 ± 2.35 (31.7 ± 11.5) |
| Residual volume, L (% pred.) | 4.74 ± 1.12 (194.7 ± 46.8) | 4.89 ± 1.11 (202.0 ± 46.5) | 5.05 ± 1.24 (211.0 ± 50.9) |
| Total lung capacity, L (% pred.) | 7.43 ± 1.07 (123.1 ± 18.1) | 7.56 ± 1.11 (126.1 ± 17.9) | 7.63 ± 1.26 (128.2 ± 20.1) |
| Exercise performance | |||
| 6-minutes walking distance, m | 242 ± 117 | 242 ± 115 | 241 ± 117 |
Data are presented as mean ± standard deviation or n (%), unless otherwise stated. FEV1 = forced expiratory volume in one second, FVC = forced vital capacity, pred. = predicted
Kappa Statistics between Two Readers with Criteria of 0% and 10%
| Criteria | Reader 2 | Kappa | Criteria | Reader 2 | Kappa | ||||
|---|---|---|---|---|---|---|---|---|---|
| Complete (0%) | Incomplete (> 0%) | Total | Complete (≤ 10%) | Incomplete (> 10%) | Total | ||||
| Reader 1 | 0.786 | Reader 1 | 0.759 | ||||||
| Complete (0%) | 156 | 16 | 172 | Complete (≤ 10%) | 159 | 8 | 167 | ||
| Incomplete (> 0%) | 12 | 92 | 104 | Incomplete (> 10%) | 23 | 86 | 109 | ||
| Total | 168 | 108 | 276 | Total | 182 | 94 | 276 | ||
Kappa Statistics between Visual Analysis and Quantitative Analysis
| Criteria | Quantitative Analysis | Quantitative Analysis | Quantitative Analysis | ||||||
|---|---|---|---|---|---|---|---|---|---|
| Complete (≤ 5%) | Incomplete (> 5%) | Kappa | Complete (≤ 10%) | Incomplete (> 10%) | Kappa | Complete (≤ 15%) | Incomplete (> 15%) | Kappa | |
| Reader 1 | |||||||||
| Complete | 159 | 8 | 161 | 6 | 162 | 5 | |||
| Incomplete | 27 | 82 | 0.726 | 45 | 64 | 0.588 | 63 | 46 | 0.432 |
| Reader 2 | |||||||||
| Complete | 173 | 9 | 176 | 6 | 179 | 3 | |||
| Incomplete | 13 | 81 | 0.821 | 30 | 64 | 0.691 | 46 | 48 | 0.556 |
| Total | 186 | 90 | 206 | 70 | 225 | 51 | |||
Fig. 3Results of Chartis™ system (Pulmonx Inc.) according to fissure completeness.
In lobes without fissural defects, Chartis™ showed positive CV in 19.6% (11/56). Meanwhile, ratio increased to 79.5% (35/44) in lobes with any fissure defect, significantly higher than ratio for lobes with complete fissure (p < 0.001). In lobes with fissure defects, rates of positive CV did not depend on size or proportion of fissure defect.
Fig. 4Some patients showed discrepant results regarding fissure completeness and TLVR after EBV insertion.
A. Two experienced radiologists agreed that 77-year-old male had complete fissure around target lobe (RUL). B. However, TLVR was not achieved in this patient 3 months after BLVR; volume of RUL changed from 1146 mL to 808 mL (29.5% reduction). C. 58-year-old male had incomplete fissure around target lobe (RUL), with size of 43.4 cm2 (7.5 × 7.4 cm, 22.7% of total fissure area). D. Despite this incompleteness, TLVR was achieved in this patient; volume of RUL was reduced from 1012 mL to 207 mL (79.5% reduction). Chartis™ system showed negative CV in target lobes of both patients. Circles mean RUL. RUL = right upper lobe, TLVR = target lobe volume reduction
TLVR Achievement according to Fissure Completeness on CT
| Criteria | Lung Volume Reduction Results | Total | |||
|---|---|---|---|---|---|
| TLVR ≥ 350 mL | TLVR > 50% of Initial Volume | ||||
| Yes | No | Yes | No | ||
| Fissure status on CT | |||||
| Complete | 19 (95%) | 1 (5%) | 14 (70%) | 6 (30%) | 20 (83.3%) |
| Incomplete | 3 (75%) | 1 (25%) | 2 (50%) | 2 (50%) | 4 (16.7%) |
| Total | 22 (91.6%) | 2 (8.3%) | 16 (66.7%) | 8 (33.3%) | 24 (100%) |
All patients had negative collateral ventilation on Chartis™ except three patients with inconclusive results. TLVR = target lobe volume reduction