Clinical Evaluation of QuickNaviTM-Ebola in the 2018 Outbreak of Ebola Virus Disease in the Democratic Republic of the Congo.

Sheila Makiala1,2, Daniel Mukadi1,2, Anja De Weggheleire3,4, Shino Muramatsu5, Daisuke Kato5, Koichi Inano5, Fumio Gondaira5, Masahiro Kajihara6, Reiko Yoshida6, Katendi Changula7,8, Aaron Mweene7,8, Placide Mbala-Kingebeni1,2, Jean-Jacques Muyembe-Tamfum1,2, Justin Masumu1, Steve Ahuka1,2, Ayato Takada9,10,11,12.   

Abstract

The recent large outbreaks of Ebola virus disease (EVD) in West Africa and the Democratic Republic of the Congo (DRC) have highlighted the need for rapid diagnostic tests to control this disease. In this study, we clinically evaluated a previously developed immunochromatography-based kit, QuickNaviTM-Ebola. During the 2018 outbreaks in DRC, 928 blood samples from EVD-suspected cases were tested with QuickNaviTM-Ebola and the WHO-approved GeneXpert. The sensitivity and specificity of QuickNaviTM-Ebola, estimated by comparing it to GeneXpert-confirmed cases, were 85% (68/80) and 99.8% (846/848), respectively. These results indicate the practical reliability of QuickNaviTM-Ebola for point-of-care diagnosis of EVD.

Entities:  

Keywords:  DRC; EVD; Ebola virus; QuickNavi; immunochromatography; rapid diagnostic test

Year:  2019        PMID: 31261668      PMCID: PMC6669708          DOI: 10.3390/v11070589

Source DB:  PubMed          Journal:  Viruses        ISSN: 1999-4915            Impact factor:   5.048


1. Introduction

Ebola virus (EBOV) is known to cause severe hemorrhagic fever in humans and/or nonhuman primates with human case fatality rates of up to 90% [1]. Five distinct ebolavirus species are known: Zaire ebolavirus (i.e., EBOV), Sudan ebolavirus, Taï Forest ebolavirus, Bundibugyo ebolavirus, and Reston ebolavirus. Ebola virus disease (EVD) poses a significant public health threat as shown by the largest EVD epidemic during the years 2013–2016 in West Africa. The second largest EVD outbreak is currently ongoing in the Democratic Republic of the Congo (DRC) where a cumulative total of 2181 confirmed and probable EVD cases and 1459 deaths have been reported since the beginning of the outbreak (as of 17 June 2019) [2]. These large-scale outbreaks emphasize the need for sensitive, easy-to-use, and robust rapid diagnostic tests (RDT) to enable quick and reliable screening of suspected EVD cases at the point-of-care. By facilitating early detection, RDTs can contribute to controlling the spread of the virus, especially in a context with known resistance to centralized EVD care. The WHO-approved GeneXpert® (Cepheid) technology, which is currently used in the field to diagnose EVD, is a major step forward compared to the conventional reverse transcription-PCR (RT-PCR) methods used previously in terms of turnaround time, ease-of-use and performance. However, this method still requires trained technicians and an uninterrupted power supply and is therefore considered a ‘near’ point-of-care test. Immunochromatography (IC)-based RDT assays for EVD could be a complementary first-line screening strategy at decentralized ‘points-of-care’ in the community. We developed an IC-based RDT for EVD, QuickNaviTM-Ebola, using mouse monoclonal antibodies (mAbs) specific to the ebolavirus nucleoprotein (NP) [3]. This kit can detect ebolaviruses in the species Zaire ebolavirus, Bundibugyo ebolavirus, and Taï Forest ebolavirus with equally high sensitivity, but cannot differentiate these viruses. Since 2016, QuickNaviTM-Ebola has been continuously provided to the Institut National de Recherche Biomédicale (INRB), DRC, and experimentally used for early diagnosis of suspected EVD cases. It is particularly noted that QuickNaviTM-Ebola supported the diagnosis of initial EVD cases confirmed by INRB in May and August in the 2018 outbreaks in the DRC [4] (Table 1).
Table 1

Diagnosis of initial Ebola virus disease (EVD) cases in the two 2018 outbreaks in the Democratic Republic of the Congo (DRC).

Sample IDOutbreaks (Month)QuickNaviTM-EbolaReal-Time PCR
18M-11st (May)NegativeNegative
18M-21st (May)PositivePositive (23.6/28.7)
18M-31st (May)NegativeNegative
18M-41st (May)NegativeNegative
18M-51st (May)NegativeNegative
18M-61st (May)PositivePositive (22.3/22.3)
18A-12nd (August)NegativePositive/Negative (39.6/NA)§
18A-22nd (August)NegativeNegative
18A-32nd (August)PositivePositive (21.6/25.0)
18A-42nd (August)PositivePositive (28.3/30.7)
18A-52nd (August)NegativeNegative
18A-62nd (August)PositivePositive (20.1/23.7)

† RealStar Filovirus Screen RT-PCR Kit (Altona Diagnostics) and LightMix® Ebola Zaire rRT-PCR Test (Roche Diagnostics). CT values of positive samples are indicated in parentheses (Altona/Roche). ‡ GeneXpert (Cepheid) to detect NP and GP genes. CT values of positive samples are indicated in parentheses (NP/GP). § Positive for NP, but negative for GP. NA: not applicable.

Currently, there are four RDTs approved by the FDA and/or WHO, ReEBOV (Corgenix), OraQuick Ebola (OraSure Technologies, Inc., Bethlehem, PA, USA), SD Q Line Ebola (SD Biosensor, Inc., Suwon, Korea), and DPP Ebola Antigen (Chembio, Medford, OR, USA) [5,6]. Of these, only OraQuick Ebola, which is based on the detection of the EBOV matrix protein, was available and used in the 2018 DRC outbreaks [4,7]. During the outbreak, QuickNaviTM-Ebola has also been experimentally used to assist screening of EVD-suspected patients at the outbreak sites. The RDT was used alongside confirmatory molecular testing, but its results were not used for clinical decision making. We report in this paper on the diagnostic performance of the QuickNaviTM-Ebola based on field data from the North-Kivu/Ituri EVD outbreak.

2. Materials and Methods

2.1. Devices

IC-based QuickNaviTM-Ebola devices were produced by using mouse mAbs ZNP105-7 and ZNP108-2-5 specific to the ebolavirus NP as described previously [3] (Figure 1a). Samples (10–30 μL) were applied onto the sample pad of the device, followed by the addition of 2 drops (approximately 40 μL) of the sample buffer supplied together with the devices. Results were interpreted 10–15 min later. Stability tests for QuickNaviTM were performed according to a standard procedure in Denka Seiken Co., Ltd. The test kit devices were stored in a storage room kept at 33 ˚C. Using virus-like particles consisting of the viral glycoprotein (GP), viral protein 40 (VP40), and NP, performance data were periodically obtained at the specified time and the sensitivity and specificity in the experiment condition were monitored up to 24 months.
Figure 1

(a) Illustration of QuickNaviTM-Ebola. A sample added to the sample window of the device migrates via capillary action. The ebolavirus NP antigens present in the sample bind to the latex-conjugated mAb on the conjugate pad. Another mAb is immobilized on a nitrocellulose membrane at the Test line position and captures the complexes of NPs and mAbs conjugated with latex. Those complexes deposit a visible blue line. (b) QuickNaviTM-Ebola used at a field laboratory in North-Kivu province.

2.2. Realtime PCR

RealStar Filovirus Screen RT-PCR Kit (Altona Diagnostics, Hamburg, Germany), LightMix® Ebola Zaire rRT-PCR Test (Roche Diagnostics, Basel, Switzerland), and GeneXpert® (Cepheid, Sunnyvale, CA, USA) were performed according to the procedures provided by the manufacturer.

2.3. Sample Collection and Detection of the Virus

Blood samples obtained by venous blood draw were collected from suspected EVD patients at Ebola Treatment Centers or health facilities in North-Kivu province (Mangina and Beni) and Ituri province (Tchomia, Bunia, and Komanda) in the DRC and subjected to rapid screening with QuickNaviTM-Ebola followed by testing with RT-PCR-based detection of the viral RNA genome (i.e., GeneXpert approved by WHO). Both tests were performed in INRB-managed field laboratories linked to the above-mentioned health facilities (Figure 1b). Most of the samples were analyzed within 5 h after sample correction. The technicians applied strict biosafety measures and adhered to the manufacturers’ instructions. Blood collection and clinical evaluation during the outbreak investigations were approved as standard care by the Ministry of Health of the DRC and oral consent was obtained from all patients before blood sampling.

3. Results and Discussion

During the outbreak caused by an ebolavirus (Zaire ebolavirus) in 2018–2019 [8], 928 whole blood samples collected from suspected EVD cases were tested with QuickNaviTM-Ebola and the WHO-approved GeneXpert. Whenever QuickNaviTM tests were available, they were used in a systematic manner on the first blood sample (and repeat if symptomatic less than 72 h at first blood draw) of each new EVD-suspect case. QuickNaviTM testing was always performed and read before GeneXpert testing. The definition of positive and negative cases is as follows. Negative: GP not detected/NP not detected, GP not detected/CT of NP ≥ 40, or GP detected/NP not detected (considered negative or vaccinated case depending on vaccination history). Positive: GP detected/NP detected (CT < 40) or GP not detected/NP detected (CT < 40). The sensitivity and specificity of the QuickNaviTM-Ebola assay were estimated by comparing its results to the GeneXpert-results and confidence intervals (CI) based on the F-distribution approximation were calculated (Table 2). Of the 80 GeneXpert-positive samples, 68 samples were positive with QuickNaviTM-Ebola, which represents a sensitivity of 85% (95% CI; 75.26–92.00). Most (10 of 12) were GeneXpert-positive, but QuickNaviTM-Ebola-negative samples had high CT values (30 <) both in NP and glycoprotein (GP) gene targets, and half of these samples were negative for detection of the GP gene (Table 3). Indeed, the distribution of CT values for NP gene detection indicated that the median, interquartile range, and mean of QuickNaviTM-Ebola-positive samples were remarkably lower than those of QuickNaviTM-Ebola-negative samples (Table 4). The QuickNaviTM-Ebola-negative samples that showed low CT values might result from a prozone effect or aggregation of too many antibody–antigen complexes which might restrict a flow on the membrane. Importantly, most of the GeneXpert-negative samples (846/848) were also negative with QuickNaviTM-Ebola, giving a specificity of 99.8% (95% CI; 99.15–99.97).
Table 2

Performance of QuickNaviTM-Ebola at the outbreak sites.

GeneXpert Ebola
PositiveNegativeTotal
QuickNaviTM-Ebola Positive68270
QuickNaviTM-Ebola Negative12846858
Total80848928

Sensitivity: 85% (68/80) (95% CI; 75.26–92.00); Specificity: 99.8% (846/848) (95% CI; 99.15–99.97); Negative predictive value: 98.6% (846/858); Positive predictive value: 97.1% (68/70); Agreement rate: 98.5% (914/928).

Table 3

Details of QuickNaviTM-Ebola-negative samples in PCR-confirmed cases.

SampleQuickNaviTM-EbolaGeneXpert Ebola
NPCT (NP)GPCT (GP)
1NegativePositive38.7NegativeNA
2NegativePositive39.9NegativeNA
3NegativePositive38.1Positive40.2
4NegativePositive38.7NegativeNA
5NegativePositive26.1Positive31.6
6NegativePositive34.2Positive42.3
7NegativePositive36.0Positive41.1
8NegativePositive33.6Positive37.9
9NegativePositive38.0NegativeNA
10NegativePositive36.8NegativeNA
11NegativePositive37.7NegativeNA
12NegativePositive13.9Positive19.1

† NA: not applicable.

Table 4

Distribution CT values of QuickNaviTM-Ebola positive and negative samples.

StatisticsQuickNaviTM-Ebola Positive (n = 68) QuickNaviTM-Ebola Negative (n = 12)
Minimum14.2013.90
25 percentile17.9834.05
Median21.0037.25
75 percentile24.8838.25
Maximum35.6039.90
Mean 21.7634.31
Standard deviation4.697.40
Considering that PCR-based assays generally show higher sensitivity than IC-based RDTs, it was quite reasonable that most samples false-negative with QuickNaviTM-Ebola showed relatively high CT values in real-time PCR-based GeneXpert (Table 4). In experimental conditions, the limit of detection (LOD) of QuickNaviTM-Ebola was 103–104 focus forming units/mL for infectious Ebola virus and 33 ng/mL for the purified recombinant EBOV NP [3]. According to the manufacturer, the LODs of FDA/WHO-approved OraQuick Ebola and ReEBOV were 1.64 × 106 TCID50/mL and 106 plaque forming units/mL, respectively, and their sensitivities estimated with whole blood clinical samples were 84% (95% CI: 63.9–95.5) (OraQuick Ebola) and 78–96% (ReEBOV), respectively [5,9]. QuickNaviTM-Ebola had 99.8% specificity in this study, whereas the reported specificities of OraQuick Ebola and ReEBOV were 98% and 73–91%, respectively. Taken together, these data may implicate at least a comparable or even better performance of QuickNaviTM-Ebola than OraQuick Ebola and ReEBOV, as indicated by its lower LOD, similar sensitivity, and higher specificity, although this comparison should ideally made using the same set of clinical samples (e.g., head-to-head comparison on stored samples). It is also worth noting that our ongoing study suggests that QuickNaviTM-Ebola has a shelflife of at least 24 months at room temperature. Ebolavirus particles consist of seven structural proteins [1]. Of these, NP, VP 40, and GP are known to abundantly present in the viral particles. In general, viral envelope glycoproteins are antigenically variable and thus thought to be unsuitable for IC tests that require the capacity to widely detect virus variants. The four RDTs approved by the FDA and/or WHO (i.e., ReEBOV, OraQuick Ebola, SD Q Line Ebola, and DPP Ebola Antigen) have been designed to detect VP40 [5,6]. In contrast, we used NP-specific mAbs to produce QuickNaviTM-Ebola. The EBOV particle contains approximately 3200 NP molecules which form large complexes of the nucleocapsid consisting of the helical nucleoprotein–RNA complex [10,11]. The NP antigen appears to be one of the ideal targets for IC assays for EVD because of the presence of common epitopes among ebolavirus species, its strong antigenicity, and the large oligomeric structure of NP complexes providing multiple antibody binding sites, which may enhance the sensitivity of the test [12]. The use of mAbs is also one of the advantages of QuickNaviTM-Ebola since the antibodies can be stably produced and can avoid the use of live animals to produce polyclonal antisera. IC-based RDTs are currently used for various viral diseases such as influenza, human adenovirus infection, and norovirus infection with significant reliability [13]. Currently available EVD RDTs cost around 10–20 US dollars/test, while some other commercialized RDTs based on the immunochromatography generally cost around 1–10 US dollars depending on their targeted diseases. Since the clinical manifestations of EVD are usually non-specific and similar to those of other infectious diseases present in EVD endemic areas, IC-based RDTs can be powerful tools for diagnosis even in remote areas in African countries. Since it is not predictable where and when EVD outbreaks will occur in the future, it is important to quickly discriminate between EVD and other viral diseases for early cases of suspected EVD, which may enable us to respond immediately to potential outbreaks, followed by initiating EVD-specific countermeasures once it is confirmed [14]. Given the simplicity of the procedure of QuickNaviTM-Ebola, it would also be of benefit if repeated testing were done for patients who were QuickNaviTM-Ebola-negative in initial screening since increased plasma viral loads in such patients should ensure the detection in a few days. The present study demonstrates that QuickNaviTM-Ebola has good sensitivity and specificity in clinical field conditions. The results were similar to or tended to be even better than those obtained for other Ebola RDTs with WHO and/or FDA approval for emergency use. Owing to its significant practical utility, including simplicity, high stability, and the absence of requirements for special equipment and training, QuickNaviTM-Ebola is expected to be a useful tool for point-of-care screening of EVD.
  9 in total

1.  Structural dissection of Ebola virus and its assembly determinants using cryo-electron tomography.

Authors:  Tanmay A M Bharat; Takeshi Noda; James D Riches; Verena Kraehling; Larissa Kolesnikova; Stephan Becker; Yoshihiro Kawaoka; John A G Briggs
Journal:  Proc Natl Acad Sci U S A       Date:  2012-02-27       Impact factor: 11.205

Review 2.  Development and application of lateral flow test strip technology for detection of infectious agents and chemical contaminants: a review.

Authors:  Babacar Ngom; Yancheng Guo; Xiliang Wang; Dingren Bi
Journal:  Anal Bioanal Chem       Date:  2010-04-27       Impact factor: 4.142

3.  Speedy Ebola tests help contain Africa's latest outbreak.

Authors:  Declan Butler
Journal:  Nature       Date:  2018-06       Impact factor: 49.962

4.  Where are the Ebola diagnostics from last time?

Authors:  Lieselotte Cnops; Birgit De Smet; Placide Mbala-Kingebeni; Johan van Griensven; Steve Ahuka-Mundeke; Kevin K Ariën
Journal:  Nature       Date:  2019-01       Impact factor: 49.962

5.  Filovirus outbreak detection and surveillance: lessons from Bundibugyo.

Authors:  Adam MacNeil; Eileen C Farnon; Oliver W Morgan; Philip Gould; Tegan K Boehmer; David D Blaney; Petra Wiersma; Jordan W Tappero; Stuart T Nichol; Thomas G Ksiazek; Pierre E Rollin
Journal:  J Infect Dis       Date:  2011-11       Impact factor: 5.226

6.  Development of an Immunochromatography Assay (QuickNavi-Ebola) to Detect Multiple Species of Ebolaviruses.

Authors:  Reiko Yoshida; Shino Muramatsu; Hiroshi Akita; Yuji Saito; Miwa Kuwahara; Daisuke Kato; Katendi Changula; Hiroko Miyamoto; Masahiro Kajihara; Rashid Manzoor; Wakako Furuyama; Andrea Marzi; Heinz Feldmann; Aaron Mweene; Justin Masumu; Jimmy Kapeteshi; Jean-Jacques Muyembe-Tamfum; Ayato Takada
Journal:  J Infect Dis       Date:  2016-07-26       Impact factor: 5.226

7.  Mapping of conserved and species-specific antibody epitopes on the Ebola virus nucleoprotein.

Authors:  Katendi Changula; Reiko Yoshida; Osamu Noyori; Andrea Marzi; Hiroko Miyamoto; Mari Ishijima; Ayaka Yokoyama; Masahiro Kajihara; Heinz Feldmann; Aaron S Mweene; Ayato Takada
Journal:  Virus Res       Date:  2013-05-20       Impact factor: 3.303

8.  Medical countermeasures during the 2018 Ebola virus disease outbreak in the North Kivu and Ituri Provinces of the Democratic Republic of the Congo: a rapid genomic assessment.

Authors:  Placide Mbala-Kingebeni; Amuri Aziza; Nicholas Di Paola; Michael R Wiley; Sheila Makiala-Mandanda; Katie Caviness; Catherine B Pratt; Jason T Ladner; Jeffrey R Kugelman; Karla Prieto; Joseph A Chitty; Peter A Larson; Brett Beitzel; Ahidjo Ayouba; Nicole Vidal; Stomy Karhemere; Mamadou Diop; Moussa M Diagne; Martin Faye; Ousmane Faye; Aaron Aruna; Justus Nsio; Felix Mulangu; Daniel Mukadi; Patrick Mukadi; John Kombe; Anastasie Mulumba; Christian-Julian Villabona-Arenas; Elisabeth Pukuta; Jeanette Gonzalez; Maggie L Bartlett; Shanmuga Sozhamannan; Stephen M Gross; Gary P Schroth; Roger Tim; Junhua J Zhao; Jens H Kuhn; Boubacar Diallo; Michel Yao; Ibrahima S Fall; Bathe Ndjoloko; Mathias Mossoko; Audrey Lacroix; Eric Delaporte; Mariano Sanchez-Lockhart; Amadou A Sall; Jean-Jacques Muyembe-Tamfum; Martine Peeters; Gustavo Palacios; Steve Ahuka-Mundeke
Journal:  Lancet Infect Dis       Date:  2019-04-15       Impact factor: 25.071

9.  Cryo-EM structure of the Ebola virus nucleoprotein-RNA complex at 3.6 Å resolution.

Authors:  Yukihiko Sugita; Hideyuki Matsunami; Yoshihiro Kawaoka; Takeshi Noda; Matthias Wolf
Journal:  Nature       Date:  2018-10-17       Impact factor: 49.962

  9 in total
  4 in total

Review 1.  Rapid diagnostic tests versus RT-PCR for Ebola virus infections: a systematic review and meta-analysis.

Authors:  Basilua Andre Muzembo; Kei Kitahara; Ayumu Ohno; Ngangu Patrick Ntontolo; Nlandu Roger Ngatu; Keinosuke Okamoto; Shin-Ichi Miyoshi
Journal:  Bull World Health Organ       Date:  2022-06-01       Impact factor: 13.831

2.  Ultrasensitive point-of-care immunoassay for secreted glycoprotein detects Ebola infection earlier than PCR.

Authors:  Cassio M Fontes; Barbara D Lipes; Jason Liu; Krystle N Agans; Aiwei Yan; Patricia Shi; Daniela F Cruz; Garrett Kelly; Kelli M Luginbuhl; Daniel Y Joh; Stephanie L Foster; Jacob Heggestad; Angus Hucknall; Maiken H Mikkelsen; Carl F Pieper; Roarke W Horstmeyer; Thomas W Geisbert; Michael D Gunn; Ashutosh Chilkoti
Journal:  Sci Transl Med       Date:  2021-04-07       Impact factor: 17.956

Review 3.  Aspects of Point-of-Care Diagnostics for Personalized Health Wellness.

Authors:  Sandeep Kumar; Monika Nehra; Sakina Khurana; Neeraj Dilbaghi; Vanish Kumar; Ajeet Kaushik; Ki-Hyun Kim
Journal:  Int J Nanomedicine       Date:  2021-01-14

Review 4.  Treatment of Ebola-related critical illness.

Authors:  Peter Kiiza; S Mullin; K Teo; N K J Adhikari; R A Fowler
Journal:  Intensive Care Med       Date:  2020-02-13       Impact factor: 41.787

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.