| Literature DB >> 31256020 |
Ariel Dora Stern1,2, William J Gordon3,4, Adam B Landman3, Daniel B Kramer4,5.
Abstract
OBJECTIVES: To more clearly define the landscape of digital medical devices subject to US Food and Drug Administration (FDA) oversight, this analysis leverages publicly available regulatory documents to characterise the prevalence and trends of software and cybersecurity features in regulated medical devices.Entities:
Keywords: FDA; cybersecurity; medical devices; software
Mesh:
Year: 2019 PMID: 31256020 PMCID: PMC6609083 DOI: 10.1136/bmjopen-2018-025374
Source DB: PubMed Journal: BMJ Open ISSN: 2044-6055 Impact factor: 2.692
Figure 1Assembly of analysis sample and results. CDRH, Center for Devices and Radiological Health; FDA, Food and Drug Administration; PMA, premarket approval.
Number of devices with machine-readable summaries by FDA/CDRH Advisory Committee and year, share with software and share of software sample with cybersecurity content by Advisory Committee
| Year | FDA/CDRH Advisory Committee | ||||||||
| Clinical chemistry | Cardiovascular | Dental | Gastroenterology, urology | General hospital | Orthopaedic | Radiology | General, plastic surgery | Totals | |
| (CH) | (CV) | (DE) | (GU) | (HO) | (OR) | (RA) | (SU) | ||
| 2002 | 216 | 436 | 318 | 215 | 328 | 403 | 290 | 367 | 2573 |
| 2003 | 192 | 441 | 295 | 233 | 329 | 389 | 329 | 357 | 2565 |
| 2004 | 204 | 395 | 284 | 195 | 270 | 464 | 345 | 319 | 2476 |
| 2005 | 155 | 389 | 245 | 166 | 262 | 480 | 310 | 331 | 2338 |
| 2006 | 197 | 412 | 293 | 142 | 244 | 442 | 338 | 362 | 2430 |
| 2007 | 153 | 358 | 283 | 160 | 257 | 444 | 271 | 319 | 2245 |
| 2008 | 149 | 387 | 279 | 139 | 207 | 477 | 325 | 370 | 2333 |
| 2009 | 130 | 442 | 268 | 155 | 254 | 432 | 290 | 316 | 2287 |
| 2010 | 121 | 390 | 245 | 157 | 280 | 428 | 235 | 312 | 2168 |
| 2011 | 163 | 428 | 258 | 141 | 241 | 542 | 347 | 285 | 2405 |
| 2012 | 155 | 426 | 240 | 166 | 282 | 551 | 344 | 302 | 2466 |
| 2013 | 185 | 428 | 235 | 153 | 202 | 554 | 346 | 301 | 2404 |
| 2014 | 130 | 400 | 225 | 199 | 245 | 583 | 385 | 342 | 2509 |
| 2015 | 108 | 392 | 244 | 179 | 174 | 575 | 340 | 322 | 2334 |
| 2016 | 95 | 368 | 230 | 171 | 204 | 464 | 375 | 354 | 2261 |
|
| 2353 | 6092 | 3942 | 2571 | 3779 | 7228 | 4870 | 4959 | 35 794 |
| Share with software (‘software sample’) | 9.14% | 18.99% | 4.59% | 8.01% | 4.97% | 1.36% | 52.28% | 6.96% | 13.79% |
| Share of software sample with cybersecurity content | 7.91% | 2.51% | 1.66% | 0.00% | 2.13% | 0.00% | 2.04% | 0.00% | 2.13% |
CDRH, Center for Devices and Radiological Health; FDA, Food and Drug Administration.
Figure 2Share of new devices with software (‘software sample’).