Literature DB >> 31239651

Exercise test is essential in LV-only fusion CRT pacing without right ventricle lead.

Cristina Vacarescu1, Dragos Cozma1,2, Lucian Petrescu1,2, Simona Dragan1,2, Cristian Mornos1,2, Simina Crisan1,2, Horea Feier1,2, Mihai-Andrei Lazar1, Ramona Alina Cozlac1,2, Constantin Tudor Luca1,2.   

Abstract

Purpose: Left ventricle (LV)-only pacing is non-inferior to biventricular pacing but permanent fusion pacing is needed to ensure cardiac resynchronization therapy (CRT) responsiveness. The role of systematic exercise testing (ET) in these patients has not been established. This study was designed to assess clinical and therapeutic implications (device programming/drugs) of systematic ET in patients requiring fusion-pacing CRT without an right ventricle (RV) lead.
Methods: Consecutive patients with a right atrium/LV-only dual-chamber (DDD) pacing system were included. Prospective data were obtained: device interrogation, ET, and echocardiography at every 6-month follow-up visit. CRT assessment during ET included maximal heart rate, beat-to-beat echocardiography analysis of LV fusion pacing, LV loss of capture, and improvement in exercise capacity. If LV loss of capture or unsatisfactory LV fusion pacing occurred, reprogramming was individualized for each patient and ET redone.
Results: A total of 55 patients (29 male) aged 62±11 years were included. During follow-up (39±18 months), a total of 235 ETs were performed, with mean exercise load 6.4±1.3 metabolic equivalents of task (118±35 W, maximal heart rate 119±17 beats/min). Twenty patients (36%) had inadequate pacing or loss of LV capture during ET, due to exceeding the maximum tracking rate (11%), chronotropic incompetence (7%), and LV pacing outside the fusion-pacing band (18%), caused by physiological shortening of the PR interval or exagerated LV preexcitation during maximum exercise. Post-ET CRT-device optimization included reprogramming of rate-adaptive atrioventricular interval (total decrease 23±8 ms), individualized programming of maximum tracking rate, or rate-response function. Drug optimization was performed in 32% of patients, and ET redone in 36%.
Conclusion: In one of three ETs, an intervention in device and medication optimization was done to ensure a better outcome. Routine ET should be a standard approach to maximize fusion-pacing CRT response during follow-up.

Entities:  

Keywords:  LV-only pacing; cardiac resynchronization therapy; constant fusion pacing; exercise test

Mesh:

Year:  2019        PMID: 31239651      PMCID: PMC6556564          DOI: 10.2147/CIA.S206251

Source DB:  PubMed          Journal:  Clin Interv Aging        ISSN: 1176-9092            Impact factor:   4.458


Introduction

Cardiac resynchronization therapy (CRT) using triple-chamber cardiac devices is the current standard treatment in heart-failure guidelines; nevertheless, left ventricle (LV)-only pacing is accepted as noninferior to classic biventricular pacing.1,2 Moreover, in a recent review prepared on behalf of the European Heart Rhythm Association Education Committee, Burri et al concluded that LV-only pacing may lead to better CRT outcomes and decrease the number of nonresponders.3 Fusion-pacing CRT in sinus rhythm has been studied in patients with biventricular pacemakers,4,5 and algorithms like adaptive CRT were designed to promote LV-only pacing in specific conditions.6 Using bicameral devices with the right atrium (RA)/LV achieves fusion pacing only in patients with normal atrioventricular (AV) conduction, and is an elegant and safe alternative to classic triple-chamber CRT, with an obvious condition of close to 100% fusion-pacing stability.7 CRT programming immediately after implantation may change during follow-up, and an optimal setting determined at rest may be different during exercise.8 How and when we need to optimize CRT is an area of ongoing debate, and even with the advancements achieved thus far, a gold standard is lacking. Optimal exercise CRT programming and tailoring to the individual patient have still not been described.9,10 Our assumption is that exercise testing (ET) can be an important follow-up investigation with powerful implications in achieving constant fusion pacing. The aim of this study was to assess and detect CRT-programming troubleshooting in a population with CRT indication and normal AV conduction in a real-world setting using ET as the principal tool in optimizing RA/LV-only CRT devices.

Methods

This study was prospectively designed to quantify exercise functional capacity and to detect rhythm changes, such as loss of ventricular capture, sinus tachycardia above upper tracking limit, or shortening/lengthening of the intrinsic PR interval during ET. Patients with CRT-pacemaker indication and normal AV conduction (PR interval <250 ms) on optimal medical treatment for at least 3–6 months without significant coronary artery disease or atrial fibrillation susceptibility (severe biatrial dilatation, longer interatrial conduction time11) implanted with RA/LV-only DDD pacing systems at Timişoara Institute of Cardiovascular Diseases during 2012–2018 were included. The condition to perform ET was at least 1 month of stable medical state without changes in medication or hospitalization due to heart-failure (HF) decompensation. Exclusion criteria were inability to perform ET due to noncardiac physical limitation (eg, orthopedic pathology) or decompensated HF. At each follow-up, patients were checked for LV threshold, spontaneous AV interval, capture quality, adequate response of pacemaker functioning to exercise, percentage of pacing, electrocardiography (ECG) 12-lead pacing on/off, and review of pharmacological therapy. For all patients included in the study, at every device interrogation we certified that both atrial sensing and atrial pacing provided the same pattern of fusion pacing on surface ECG. Complete transthoracic echocardiography using standard views and techniques was also performed in all patients, with special focus during follow-up on LV systolic function parameters: LV end-systolic volume, LV end-diastolic volume, and LV ejection fraction. Cycle-ergometry ET was performed on a GE exercise system with an increase in workload of 25 W for each 3-minute exercise stage. Blood pressure and 12-lead ECG were continuously recorded. Exercise capacity was measured in metabolic equivalents of task at peak exercise. CRT assessment during ET included maximal heart rate, beat-to-beat ECG analysis of true LV-fusion pacing, loss of LV capture, and improvement in exercise capacity. Even if ET at baseline offered a clear advantage regarding PR interval behavior during ET, most patients were not able to exercise before implantation, and thus isolated ET sessions could not be analyzed. In patients with loss of LV capture during exercise by exceeding maximum tracking rate (MTR), drug optimization was performed (uptitration of β-blocker [BB]/adding ivabradine) associated with increasing MTR limit by 15 beats/min. When chronotropic incompetence was noted, rate-response function was programmed and ET redone immediately. All ET sessions were performed with live monitoring of the PM, and one of the main objectives was to maintain constant fusion pacing, both at rest and during maximal exercise. We defined the fusion-pacing band according to existing evidence12 as the range of AV intervals at which surface ECG (mainly in lead V1) has an intermediate morphology between native left bundle-branch block (upper limit of the band) and fully paced right BBB (lower limit). Two situations were noted outside the fusion pacing band: in patients with loss of capture by physiological shortening of PR interval or improper fusion pacing (rest-to-exercise variations regarding R-wave reduction in V1), the dynamic AV interval was progressively decreased by 20 ms, and in patients with important LV preexcitation during maximum exercise (right BBB appearance in V1), the dynamic AV interval was progressively increased by 20 ms. The AV interval was programmed individually and beat-to-beat ECG analysis performed to ensure adequate and constant fusion pacing. The stability of the spontaneous PR interval was obtained by BB ± ivabradine-dose titration, such that in shorter PR–interval patients the dose of BB alone was increased, while in longer PR–interval patients ivabradine without BB or in combination with lower doses of BB was given. Various BBs have been used; however, for study purposes carvedilol-equivalent doses were calculated using the algorithm described by Cohen-Solal et al.13 In patients where only device optimization was needed, ET was redone immediately after reprogramming to ensure adequate fusion pacing. In patients with drug-therapy changes, ET was redone at an interval no longer than 2 weeks. The study was conducted in accordance with the 1964 Declaration of Helsinki, and the protocol was approved by the ethical review committee of Timişoara Institute of Cardiovascular Diseases (1622/26.03.2014). Written informed consent was obtained from all patients included in the study.

Statistical analysis

Data are expressed as means ± SD for continuous variables and proportions for categorical variables. Continuous variables were compared between groups using unpaired t-tests (variables with normal distribution) or Mann–Whitney U tests (abnormally distributed variables). Proportions were compared using χ2 and Fisher’s exact test. P<0.05 was considered significant. All analyses were carried out with the SPSS version 18.0 (SPSS Inc., chicago, IL, USA).

Results

A total of 55 patients with RA/LV DDD pacing systems were initially included. Seven patients had been lost by the end of the study (in five patients, nonsudden cardiac death occurred, and two patients developed orthopedic pathology). Basic demographic and echocardiography data are presented in Table 1. All patients were receiving optimal medical therapy, including HF medication (BB/ivabradine, ACE inhibitors/ARBs, antialdosteronics) individually titrated according to clinical and paraclinical variables (heart rate, blood pressure, renal function). During follow-up, we noted clinical and paraclinical improvement, and medication was uptitrated to target doses where possible; however, the main goal of BB and ivabradine titration was to obtain stable AV conduction to ensure constant fusion pacing both at rest and during exercise.
Table 1

Baseline demographic data

All patients (n=55)
Mean age (years)62±11
Male, n (%)30 (55%)
NYHA functional class(n, %)II24 (44%)
III31 (56%)
ElectrocardiographyPR interval (ms), mean ± SD186±32
QRS complex (ms), mean ± SD163±17
Mitral regurgitation (n, %)Mild6 (11%)
Moderate27 (49%)
Severe22 (40%)
Hypertension, n (%)20 (36%)
Diabetes mellitus, n (%)20 (36%)
Chronic kidney disease, n (%)*23 (42%)
Echographic dataLVEDD (cm), mean ± SD6.4±0.9
LVEF (%),** mean ± SD27±5.2
LVEDV (mL), mean ± SD243.2±82
LVESV (mL), mean ± SD182.4±73
LAV (mL), mean ± SD104.9±34
sPAP (mmHg), mean ± SD46.6±15

Notes: *Defined as reduction in creatinine clearance <90 mL/min. None of the patients in our cohort had creatinine clearance <30 mL/min. **Measured by modified Simpson’s method.

Abbreviations: LAV, left atrial volume; LVEDD, left ventricular end-diastolic diameter; LVEDV, left ventricular end-diastolic volume; LVEF, left ventricular ejection fraction; LVESV, left ventricular end-systolic volume; NYHA, New York Heart Association; sPAP, systolic pulmonary artery pressure.

Baseline demographic data Notes: *Defined as reduction in creatinine clearance <90 mL/min. None of the patients in our cohort had creatinine clearance <30 mL/min. **Measured by modified Simpson’s method. Abbreviations: LAV, left atrial volume; LVEDD, left ventricular end-diastolic diameter; LVEDV, left ventricular end-diastolic volume; LVEF, left ventricular ejection fraction; LVESV, left ventricular end-systolic volume; NYHA, New York Heart Association; sPAP, systolic pulmonary artery pressure. The following Pacing systems were implanted: Biotronik (eight patient), Medtronic (eight patients), St Jude Medical (34 patients), Boston Scientific (two patients), and Sorin Group (three patients). Bipolar RA/LV leads were used. No short- or long-term device-related complications were noted (eg, phrenic nerve stimulation, pocket infections). At baseline, all devices were programmed at a rest rate of 60 beats/min and MTR of 130 beats/min. Individualized AV intervals with AV pace of 147±22 ms and sensed AV of 119±25 ms (dynamic AV interval at baseline equal to sensed AV), allowing fusion pacing in all patients. During follow-up of 39±18 months, 235 ET sessions were done, with a mean of 118±35 W (6.4±1.3 metabolic equivalents of task) and a peak heart rate of 119±17 beats/min. Inadequate pacing or loss of LV capture was noted in 20 patients (36%), with causes identified being MTR exceeded (11% of patients), chronotropic incompetence (7% of patients; Figure 1), LV pacing outside the fusion-pacing band (18% of patients), including loss of capture by physiological shortening of PR interval or improper fusion pacing (rest-to-exercise variations in R-wave reduction in V1, 15% of patients) and important LV preexcitation during maximum exercise with right BBB appearance in V1 (3% of patients; Figure 2). Post-ET CRT optimization was performed by increasing MTR at 145 beats/min and reprogramming dynamic AV interval. After ET optimization, we registered a decrease in dynamic AV interval of 23±8 ms.
Figure 1

Heart-rate diagram during exercise (red line).

Notes: (A) Example of a patient with chronotropic incompetence; (B) heart-rate acceleration after reprogramming rate-response function.

Figure 2

LV pacing during exercise outside the fusion pacing band.

Notes: (A) Rest-to exercise-variations regarding R-wave reduction in V1; (B) important LV preexcitation during maximum exercise (RBBB appearance in V1). Incarca=exercise load; HR=heart rate; BP=blood pressure; Fază=phase; Etapă=stage; treaptă=level; RECUPERARE=recovery; vârf=peak; Referinţă=reference; bpm=beats per minute.

Abbreviations: LV, left ventricle; RBBB, right bundle-branch block.

Heart-rate diagram during exercise (red line). Notes: (A) Example of a patient with chronotropic incompetence; (B) heart-rate acceleration after reprogramming rate-response function. LV pacing during exercise outside the fusion pacing band. Notes: (A) Rest-to exercise-variations regarding R-wave reduction in V1; (B) important LV preexcitation during maximum exercise (RBBB appearance in V1). Incarca=exercise load; HR=heart rate; BP=blood pressure; Fază=phase; Etapă=stage; treaptă=level; RECUPERARE=recovery; vârf=peak; Referinţă=reference; bpm=beats per minute. Abbreviations: LV, left ventricle; RBBB, right bundle-branch block. Medication changes included BB uptitration with addition/uptitration of ivabradine in six patients (11%), due to LV-capture loss by exceeding MTR, and uptitration of BB in another nine patients (16%), due to loss of LV capture by physiological shortening of PR interval or improper fusion pacing. In three patients (5%) treated with BB/ivabradine, chronotropic incompetence was noted and ivabradine excluded from treatment. Comparative data representing baseline and follow-up demographic characteristic, detailed changes in medication, and device reprogramming after performing ET are presented in Table 2.
Table 2

Comparative data representing baseline and follow-up demographic characteristics, detailed changes in medication, and device reprogramming after ET (n=55)

BaselinePost-ET optimization(follow-up 39±18 months)Relative change, %*P-value
LVEF, mean ± SD27±5.237±7.927%<0.0001
LVEDV (mL), mean ± SD243.2±82193.7±8120%0.0028
LVESV (mL), mean ± SD182.4±73113±6338%<0.0001
LVESV reduction ≥15%, patients1629%
Increasing MTR, patients611%
Programming rate-response function, patients47%
Reprogramming dynamic AV interval, patients1018%
Dynamic AV interval, ms119±2596±1719%<0.0001
BB therapy, patients (%)45 (82%)54 (98%)16%
Maximal target-dose BB, patients (%)6 (11%)21 (38%)27%
Carvedilol-equivalent dose, mg (mean ± SD)6.7±717.5±1262%<0.0001
Ivabradine therapy, patients (%)24 (44%)27 (49%)5%
Ivabradine daily dose, mg, mean ± SD4.6±25.3±213%0.0692

Abbreviations: AV, atrioventricular; BB, β-blocker; ET, exercise testing; LAV, left atrial volume; LVEF, left ventricular ejection fraction; LVEDV, left ventricular end-diastolic volume; LVESV, left ventricular end-systolic volume; MTR, maximum tracking rate.

Comparative data representing baseline and follow-up demographic characteristics, detailed changes in medication, and device reprogramming after ET (n=55) Abbreviations: AV, atrioventricular; BB, β-blocker; ET, exercise testing; LAV, left atrial volume; LVEF, left ventricular ejection fraction; LVEDV, left ventricular end-diastolic volume; LVESV, left ventricular end-systolic volume; MTR, maximum tracking rate. During ET, episodes of paroxysmal atrial fibrillation at maximum exercise were noted in two patients (4%), and amiodarone was introduced without recurrence at further examinations. In one patient (2%), after 3 years’ follow-up we noted at ET alternating fusion narrow QRS and wide complete LV-capture QRS. Pacemaker inhibition during exercise revealed Mobitz II second-degree AV block (Figure 3). After amiodarone and BB cessation, the AV block persisted and the patient upgraded to a classic triple-chamber CRT device, without complications.
Figure 3

Mobitz II second degree AV block occurrence during ET.

Notes: (A) LV-only fusion pacing at the beginning of ET; (B) at 100 W, fusion narrow QRS complexes and wide complete LV capture QRS complexes; (C) pacemaker inhibition with Mobitz II second-degree AV block.

Abbreviations: AV, atrioventricular; ET, exercise testing; LV, left ventricle.

Mobitz II second degree AV block occurrence during ET. Notes: (A) LV-only fusion pacing at the beginning of ET; (B) at 100 W, fusion narrow QRS complexes and wide complete LV capture QRS complexes; (C) pacemaker inhibition with Mobitz II second-degree AV block. Abbreviations: AV, atrioventricular; ET, exercise testing; LV, left ventricle. At the end of follow-up, 36% of patients needed CRT-device reprogramming, while in 32% of patients medication optimization was performed. We also found statistical significance for several echocardiography parameters: LV end-systolic volume (182.4±73 vs 113±63, P<0.0001), LV end-diastolic volume (243.2±82 vs 193.7±81, P<0.0028), and LV ejection fraction (27±5.2 vs 37±7.9, P<0.0001). The main outcome of this systematic follow-up was a 90% responder rate in our cohort.

Discussion

This study shows that ET represents an important tool for follow-up and optimization in CRT devices that provide LV-only fusion pacing. The usefulness of ET in CRT follow-up was highlighted in patients with classic triple-chamber CRT devices and biventricular pacing, and even in small series, a significant percentage of patients need device reprogramming after ET.14,15 With the ultimate goal of maximizing CRT benefits, the current guideline recommends empirical programming of 100–120 ms sensed AV delay and simultaneous biventricular stimulation; however, the population of 30% nonresponders remains the Achilles heel of CRT, with no clear evidence of optimization or improvement.2 Leclercq identifies the most common reasons for loss of capture at exercise in biventricular devices: loss of atrial sensing, frequent premature ventricular complexes, atrial tachyarrhythmia, and spontaneous AV conduction that is more rapid than the programmed AV delay.16 All these issues come up in RA/LV-only CRT devices, with the addition of a specific problem: maintaining constant fusion pacing both at rest and during exercise. Physiological shortening of the AV interval with increasing heart rate has been well documented in normal subjects, but also in patients with New York Heart Association III/IV HF on β-antagonist or antiarrhythmic therapy, is mostly observed in earlier stages of exercise, and represents an established component of dual-chamber pacing for bradycardia as rate-adaptive or dynamic AV intervals.17,18 Although fusion pacing has clear evidence of substantially improving the structural responder rate,19–21 the variability of AV conduction still remains the main concern regarding LV-only fusion pacing, both in the short and long term, due to changes in disease state, exercise capacity, and medication.3 The algorithm described by the investigators of an adaptive-CRT trial,6 which promotes intrinsic conduction and reduces RV pacing, has been proven save and effective for heart rate <100/min (empirical limit based on a 22-patient hemodynamic study), and the only algorithm designed for DDD LV-only patients without an RV lead was described in 2007.7,22 Developing a device-based algorithm for beat-by-beat monitoring of effective LV pacing is feasible23 and still a matter of high interest; nevertheless, with a lack of large controlled randomized trials and guideline-based evidence, ET remains an important real-life investigation that can assess constant fusion pacing. A mainstay class of drug therapy in HF, β-blockers have important implications in providing fusion-pacing stability by their effect of slowing AV conduction. Moreover, a combination of early coadministration of BB–ivabradine for patients with acute HF with reduced ejection fraction showed 1-year improvement in LV systolic function.24 Nevertheless, recent data have highlighted that uptitration of HF medication is still a major drawback in real-world practice, and several studies have shown a general effort toward optimization of therapy after CRT with important benefits in reducing HF hospitalization and increasing survival, with a special focus on BB uptitration as an independent predictor of improved HF prognosis.25–28 Although such medications as ACE inhibitors/ARBs, loop diuretics, and antialdosteronics are of utmost importance in HF patients and changes pertaining to these drugs were also performed in our cohort during follow-up according to functional status and clinical/paraclinical variables, this study aimed to assess the importance of BB/ivabradine therapy and CRT-device reprogramming after systematic ET in maintaining constant CRT fusion pacing, which has proven to be a real need in clinical practice. The small, selective cohort and lack of randomization are important limitations; however, the study highlights the importance of finding better methods of evaluation and optimization of LV-only CRT fusion pacing. Although the end point of the study was not a clinical one in terms of HF hospitalizations or mortality, we succeeded in demonstrating that ET has powerful implications regarding device and medication optimization in fusion-pacing CRT patients. We found at the end of the study an important statistical significance regarding echocardiography parameters of LV systolic function, correlating with a 90% responder rate in our cohort; we can only assume this finding is based on a careful follow-up to permanently reassess treatment and device optimization.

Conclusion

A third of CRT fusion-pacing patients need device and medication optimization after ET. Routine ET has important implications in maintaining constant fusion pacing, and should be a standard approach to maximize CRT response during follow-up.
  3 in total

1.  Betablockers and Ivabradine Titration According to Exercise Test in LV Only Fusion CRT Pacing.

Authors:  Cristina Vacarescu; Constantin-Tudor Luca; Horea Feier; Dan Gaiță; Simina Crișan; Alina-Gabriela Negru; Stela Iurciuc; Emilia-Violeta Goanță; Cristian Mornos; Mihai-Andrei Lazăr; Caius-Glad Streian; Diana-Aurora Arnăutu; Vladiana-Romina Turi; Dragos Cozma
Journal:  Diagnostics (Basel)       Date:  2022-04-27

2.  A Novel and Simple Exercise Test Parameter to Assess Responsiveness to Cardiac Resynchronization Therapy.

Authors:  Alina-Ramona Cozlac; Lucian Petrescu; Simina Crisan; Constantin Tudor Luca; Cristina Vacarescu; Caius Glad Streian; Mihai-Andrei Lazar; Andra Gurgu; Angela Dragomir; Emilia Violeta Goanta; Radu Vatasescu; Kandara Mohamed Chahine; Ciprian Rachieru; Dragos Cozma
Journal:  Diagnostics (Basel)       Date:  2020-11-09

Review 3.  Arterial Hypertension: Individual Therapeutic Approaches-From DNA Sequencing to Gender Differentiation and New Therapeutic Targets.

Authors:  Constantin-Tudor Luca; Simina Crisan; Dragos Cozma; Alina Negru; Mihai-Andrei Lazar; Cristina Vacarescu; Mihai Trofenciuc; Ciprian Rachieru; Laura Maria Craciun; Dan Gaita; Lucian Petrescu; Alexandru Mischie; Stela Iurciuc
Journal:  Pharmaceutics       Date:  2021-06-09       Impact factor: 6.321

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.