Purpose: To assess the value of intraoperative aberrometry (IA) in determining toric intraocular lens (IOL) power in eyes with no previous ocular surgery. Patients and methods: This was a retrospective data review at one US clinical site of eyes that underwent uncomplicated cataract surgery with toric IOL implantation where standard preoperative and IA measurements were available. Calculated IOL sphere and cylinder powers and orientation were compared based on the measurement method and the postoperative refraction, using both actual and simulated (back-calculated) results. Comparisons were between the surgeon's preoperative calculations, IA measurements, the actual IOL implanted and results from the Barrett toric calculator.Entities:
Keywords: IOL power calculation; ORA; astigmatism; prediction error; refraction; toric IOL
Year: 2019 PMID: 31239634 PMCID: PMC6551590 DOI: 10.2147/OPTH.S205352
Source DB: PubMed Journal: Clin Ophthalmol ISSN: 1177-5467
Residual spherical equivalent refraction by method (n=118)
| ±0.25D | ±0.50D | ±1.00D | Mean (D) | Mean absolute (D) | |
|---|---|---|---|---|---|
| Actual | 89 (75.4%) | 109 (92.4%) | 118 (100%) | −0.10 | 0.23 |
| Preop | 67 (56.8%) | 102 (86.4%) | 118 (100%) | −0.21 | 0.31 |
| IA | 85 (72%) | 110 (93.2%) | 117 (99.2%) | −0.08 | 0.24 |
Abbreviations: D, diopter; Preop, preoperative; IA, intraoperative aberrometry.
Best IOL sphere power by difference in suggested lens power
| Preop better | 5 | 3 | 1 | 1 | |
| Same | 108 | 52 | 50 | 6 | |
| IA better | 5 | 5 | |||
| All eyes | 118 | 8 | 53 | 50 | 7 |
Abbreviations: n, number of eyes; D, diopter; IA, intraoperative aberrometry; Preop, preoperative; IOL, intraocular lens.
Residual cylinder by method (n=118)
| ±0.50D | ±0.75D | ±1.00D | Mean (SD) | |
|---|---|---|---|---|
| Actual | 92 (78.0%) | 111 (94.1%) | 115 (97.5%) | 0.40 (0.32) |
| Preop | 63 (53.4%) | 90 (76.3%) | 101 (85.6%) | 0.64 (0.40) |
| IA | 89 (75.4%) | 107 (90.7%) | 111 (94.1%) | 0.45 (0.44) |
| Barrett | 89 (75.4%) | 104 (88.1%) | 111 (94.1%) | 0.46 (0.40) |
Abbreviations: n, number of eyes; D, diopter; IA, intraoperative aberrometry; Preop, preoperative.
Differences in IOL cylinder power and/or orientation
| IA vs Actual | IOL | Same | 83 (70%) | 5 (4%) | 88 (75%) |
| Different | 23 (19%) | 7 (6%) | 30 (25%) | ||
| Total | 106 (90%) | 12 (10%) | 118 (100%) | ||
| Same | Different | Total | |||
| IA vs Barrett | IOL | Same | 21 (18%) | 40 (34%) | 61 (52%) |
| Different | 20 (17%) | 37 (31%) | 57 (48%) | ||
| Total | 41 (35%) | 77 (65%) | 118 (100%) | ||
| Same | Different | Total | |||
| IA vs Preop | IOL | Same | 16 (14%) | 37 (31%) | 53 (45%) |
| Different | 21 (18%) | 44 (37%) | 65 (55%) | ||
| Total | 37 (31%) | 81 (69%) | 118 (100%) | ||
Abbreviations: vs, versus; IA, intraoperative aberrometry; Preop, preoperative.
Relatively lower expected residual cylinder by IOL calculation method
| IA vs Actual | Same | Same | 83 | 83 | ||
| Same | Different | 1 | 4 | 5 | ||
| Different | Same | 20 | 3 | 23 | ||
| Different | Different | 4 | 3 | 7 | ||
| Total | 5 | 110 | 3 | 118 | ||
| IOL | Orientation | Barrett Better | Same | IA Better | Total | |
| IA vs Barrett | Same | Same | 21 | 21 | ||
| Same | Different | 4 | 26 | 10 | 40 | |
| Different | Same | 7 | 8 | 5 | 20 | |
| Different | Different | 9 | 22 | 6 | 37 | |
| Total | 20 | 77 | 21 | 118 | ||
| IOL | Orientation | Preop Better | Same | IA Better | Total | |
| IA vs Preop | Same | Same | 16 | 16 | ||
| Same | Different | 4 | 19 | 14 | 37 | |
| Different | Same | 6 | 10 | 5 | 21 | |
| Different | Different | 6 | 16 | 22 | 44 | |
| Total | 16 | 61 | 41 | 118 |