Literature DB >> 31237960

Technical feasibility of umbilical cannulation in midgestation lambs supported by the EXTra-uterine Environment for Neonatal Development (EXTEND).

Matthew A Hornick1, Ali Y Mejaddam1, Patrick E McGovern1, Grace Hwang1, Jiancheng Han1, William H Peranteau1, Emily A Partridge1, Marcus G Davey1, Alan W Flake1.   

Abstract

EXTEND (EXTra-uterine Environment for Neonatal Development) is a novel system for supporting extremely premature infants that replicates in utero conditions by maintaining a sterile fluid environment and providing gas exchange via a pumpless arteriovenous oxygenator circuit connected to the umbilical vessels. Target gestational age (GA) for EXTEND support in human infants is 23-27 weeks, when immature lungs are most susceptible to injury in the setting of air ventilation. We previously demonstrated physiologic support of premature lambs cannulated at 105-117 days GA (lungs developmentally analogous to the 23-27 week GA human infant) for up to 28 days on EXTEND. In the present study, we sought to determine the technical feasibility of umbilical vessel cannulation in 85-96 days GA lambs delivered to EXTEND at weights equivalent to the 23-27 week GA human infant (500-850 g). Five preterm lambs were cannulated at 85-96 days GA (term 145 days) and supported on EXTEND for 4-7 days. All lambs underwent umbilical artery and umbilical vein cannulation. Circuit flows and pressures were monitored continuously, and blood gases were obtained at regular intervals for assessment of oxygen parameters. Systemic pH and lactate were measured at least once daily. Mean body weight at cannulation was 641 ± 71 g (range 480-850 g). All lambs were cannulated successfully (cannula size varied from 8 to 12 Fr), and mean survival on EXTEND was 140 ± 7 hours. Mean circuit flow was 213 ± 15 mL/kg*min, mean pH was 7.37 ± 0.01, and mean lactate was 1.6 ± 0.2 mmol/L. During the initial 120 hours after EXTEND cannulation, there were no significant differences between 85-96 days GA lambs and 105-117 days GA lambs in weight-adjusted circuit flows, oxygen delivery, pH, or lactate levels. This study demonstrates successful umbilical cord cannulation and adequate circuit flows and oxygen delivery in midgestation lambs size-matched to the 23-27 week GA human fetus, which represents an important step in the translation of EXTEND to clinical practice.
© 2019 International Center for Artificial Organs and Transplantation and Wiley Periodicals, Inc.

Entities:  

Keywords:  EXTEND; midgestation; umbilical cannulation

Mesh:

Year:  2019        PMID: 31237960     DOI: 10.1111/aor.13524

Source DB:  PubMed          Journal:  Artif Organs        ISSN: 0160-564X            Impact factor:   3.094


  5 in total

Review 1.  Development of an artificial placenta for support of premature infants: narrative review of the history, recent milestones, and future innovation.

Authors:  Brian P Fallon; George B Mychaliska
Journal:  Transl Pediatr       Date:  2021-05

2.  The EXTrauterine Environment for Neonatal Development Supports Normal Intestinal Maturation and Development.

Authors:  Heron D Baumgarten; Christina M Wright; Avery C Rossidis; Kendall M Lawrence; Aimee G Kim; Ali Y Mejaddam; Patrick E McGovern; Melissa N Orr; Barbara E Coons; Zoya Butt; Haiying Li; Grace Hwang; Antoneta Radu; Lauren J Brown; Ronald C Rubenstein; William H Peranteau; Marcus Davey; Robert O Heuckeroth; Alan W Flake
Journal:  Cell Mol Gastroenterol Hepatol       Date:  2020-05-28

3.  Impact of the Addition of a Centrifugal Pump in a Preterm Miniature Pig Model of the Artificial Placenta.

Authors:  Alex J Charest-Pekeski; Steven K S Cho; Tanroop Aujla; Liqun Sun; Alejandro A Floh; Mark J McVey; Ayman Sheta; Marvin Estrada; Lynn Crawford-Lean; Celeste Foreman; Dariusz Mroczek; Jaques Belik; Brahmdeep S Saini; Jessie Mei Lim; Olivia J Moir; Fu-Tsuen Lee; Megan Quinn; Jack R T Darby; Mike Seed; Janna L Morrison; Christoph Haller
Journal:  Front Physiol       Date:  2022-07-22       Impact factor: 4.755

Review 4.  Advances in extracorporeal membrane oxygenator design for artificial placenta technology.

Authors:  David G Blauvelt; Emily N Abada; Peter Oishi; Shuvo Roy
Journal:  Artif Organs       Date:  2020-11-04       Impact factor: 3.094

Review 5.  Pediatric and neonatal extracorporeal life support: current state and continuing evolution.

Authors:  Brian P Fallon; Samir K Gadepalli; Ronald B Hirschl
Journal:  Pediatr Surg Int       Date:  2021-01-01       Impact factor: 1.827

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.