Literature DB >> 31231810

Increased risk of hematoma with uninterrupted warfarin in patients undergoing implantation of subcutaneous implantable cardioverter defibrillator.

Christopher Evenson1, Basil Saour1, Muhammad R Afzal1, Bradley Knight1, Toshimasa Okabe1, Raul Weiss1.   

Abstract

BACKGROUND: The perioperative management of anticoagulation with the use of subcutaneous implantable cardioverter-defibrillator (S-ICD) implantation is still evolving.
OBJECTIVE: The purpose of this study was to assess whether it is safe to perform S-ICD implantation with uninterrupted warfarin.
METHODS: This is a multi-center retrospective review of patients undergoing S-ICD implantation between October 1, 2012 and June 30, 2017. Forty-eight patients underwent successful S-ICD implantation during the study period. The most common indication for implantation was primary prevention of sudden cardiac death. In 23 (47.9%) patients, warfarin was continued without any interruption. In 25 (52.1%) patients, warfarin was interrupted prior to implantation. The incidence of clinically significant lateral pocket hematoma was compared in the two groups.
RESULTS: The mean international normalized ratio was 2.0 ± 0.4 in the uninterrupted group and 1.4 ± 0.4 for the interrupted group. A total of seven patients developed a hematoma at the lateral pocket. No patient developed a hematoma at the parasternal pockets. Six patients in the uninterrupted group (26.1%) and one patient in the interrupted group (0.04%) developed a significant lateral pocket hematoma (P = .04). The concomitant use of dual antiplatelet therapy did not increase the risk of hematoma. There was no significant difference between HASBLED and CHA2 DS2 VASc scores between the groups. None of the patients with a hematoma developed infection or required hematoma evacuation.
CONCLUSIONS: The uninterrupted use of warfarin in the perioperative period of S-ICD implantation is associated with an increased risk of lateral pocket hematoma.
© 2019 Wiley Periodicals, Inc.

Entities:  

Keywords:  anticoagulation; hematoma; lateral pocket; subcutaneous defibrillator; warfarin

Year:  2019        PMID: 31231810     DOI: 10.1111/pace.13754

Source DB:  PubMed          Journal:  Pacing Clin Electrophysiol        ISSN: 0147-8389            Impact factor:   1.976


  3 in total

1.  Ultrasound-guided intermuscular pocket creation for a subcutaneous implantable cardioverter-defibrillator.

Authors:  Yusuke Sonoda; Koji Fukuzawa; Yu Izawa; Jun Sakai; Ken-Ichi Hirata
Journal:  HeartRhythm Case Rep       Date:  2021-11-18

Review 2.  Process mapping strategies to prevent subcutaneous implantable cardioverter-defibrillator infections.

Authors:  Raul Weiss; George E Mark; Mikhael F El-Chami; Mauro Biffi; Vincent Probst; Pier D Lambiase; Marc A Miller; Timothy McClernon; Linda K Hansen; Bradley P Knight; Larry M Baddour
Journal:  J Cardiovasc Electrophysiol       Date:  2022-06-09       Impact factor: 2.942

3.  Approaches to Minimizing Periprocedural Complications During Subcutaneous Implantable Cardioverter-defibrillator Placement.

Authors:  Srikanth Vedachalam; Schuyler Cook; Tanner Koppert; Toshimasa Okabe; Raul Weiss; Muhammad R Afzal
Journal:  J Innov Card Rhythm Manag       Date:  2020-05-15
  3 in total