Miguel Aliste-Fernández1, Gemma Sole-Enrech1, Ruth Cano-Corres1, Silvia Teodoro-Marin2, Eugenio Berlanga-Escalera1.
Abstract
INTRODUCTION: Measurement of high-sensitivity troponin T (hs-TnT) has become an essential step in the diagnosis of acute myocardial infarction. This high-sensitivity method allows quantifying the concentration of troponin T in blood of healthy subjects with a lower inaccuracy compared to previous reagent generations. However, the presence of certain compounds in the sample may interfere with the result. We present a patient who had repeatedly high concentrations of hs-TnT in the serum sample that did not agreed with the signs and symptoms. In addition, ultrasensitive troponin I concentration was undetectable.Entities:
Keywords: case report; high-sensitivity troponin T; interference
Mesh:
Substances:
Year: 2019 PMID: 31223267 PMCID: PMC6559625 DOI: 10.11613/BM.2019.021001
Source DB: PubMed Journal: Biochem Med (Zagreb) ISSN: 1330-0962 Impact factor: 2.313
Measurements performed to elucidate the cause of the falsely elevated results of hs-TnT
| Serum indices: | No significant interference up to: | Cobas 8000 | |
| Creatine kinase: 152 U/L | 0-195 U/L | Cobas 8000 | |
| Alkaline phosphatase: 150 U/L | 2-300 U/L | Cobas 8000 | |
| Serum creatinine: 52 µmol/L | 34–65 µmol/L | Cobas 8000 | |
| Sample before reaction: 52 ng/L | - | Cobas e801 | |
| PTH: 14.4 ng/L | 10.0-65.0 ng/L | Cobas e801 | |
| Immunoglobulin G: 14.30g/L | 7.40–14.00 g/L | Immage 800 | |
| < 20 IU/mL | 70–200 IU/mL | Immage 800 | |
| ANA, ASMA, AMA and APCA: < 1/40 | < 1/40 | Quantalyser 160 | |
| Sample before incubation: 32 ng/L | - | Cobas e601 | |
| hs-TnT - high-sensitivity troponin T. HBT - heterophilic blocking tube. PTH - parathyroid hormone. CA - carbohydrate antigen. AFP - alpha-fetoprotein. TPSA - total specific prostate antigen. TSH – thyrotropin. FT4 - free thyroxine. ANA - anti-nuclear antibodies. ASMA - anti-smooth muscle antibodies. AMA - anti-mitochondrial antibodies. APCA - anti-parietal cells antibodies. ANCA - anti-neutrophil cytoplasmic antibodies. | |||
Evolution of the different myocardial damage markers in the patient; single reading
| 115 | 51 | - | - | |
| 49 | 98 | - | - | |
| 152 | 52 | 28 | < 10 | |
| 129 | - | 32 | < 10 | |
| CK - creatine kinase. hs-TnT - high-sensitivity troponin T. STAT - short turnaround time. hs-TnI - high-sensitivity troponin I. The 99th percentile for the specific test was: 0 - 195 U/L (CK), ≤ 14 ng/L (hs-TnT and hs-TnT STAT), ≤ 26 ng/L (hs-TnI STAT). | ||||
Results of dilution test (Cobas e801)
| 52 ng/L | 33 ng/L | 24 ng/L | 18 ng/L | |
| 52 ng/L | < 26 ng/L | < 17 ng/L | < 10 ng/L | |
| DF – dilution factor. | ||||
Figure 1Linearity test between expected and obtained values of the dilution test (ng/L).
Manufacturer’s study in which troponin T concentration were measured with different analysers with electrochemiluminescence immunoassay technology
| 34 | 41 | 32 | < 10 | 54 | |
| hs-TnT STAT - high-sensitivity troponin T short turnaround time. TnT STAT - troponin T short turnaround time. hs-TnT - high-sensitivity troponin T. | |||||