Comparison between the new fully automated viscoelastic coagulation analysers TEG 6s and ROTEM Sigma in trauma patients: A prospective observational study.

Bernhard Ziegler1, Wolfgang Voelckel, Johannes Zipperle, Oliver Grottke, Herbert Schöchl.   

Abstract

BACKGROUND: Viscoelastic coagulation testing is increasingly used to diagnose trauma-induced coagulopathy. Two fully automated analysers, TEG 6s and ROTEM Sigma, were launched recently. No previous studies have compared these devices in trauma paients.
OBJECTIVE: The aim of this study was to evaluate whether both fully automatic devices deliver comparable results.
DESIGN: Prospective observational study.
SETTING: Level one trauma centre from August 2017 to September 2018. PATIENTS: A total of 105 blood samples from 67 trauma patients were analysed simultaneously on TEG 6s and ROTEM Sigma. MAIN OUTCOME MEASURES: TEG 6s assays kaolin (CK), RapidTEG (CRT), kaolin with heparinase (CKH) and functional fibrinogen were compared with ROTEM Sigma assays INTEM, EXTEM, HEPTEM and FIBTEM. TEG 6s functional fibrinogen level was compared with plasma fibrinogen concentration, measured using the Clauss method. Correlations were classified as weak (Spearman correlation coefficient 0.20 to 0.39), moderate (0.40 to 0.59), strong (0.60 to 0.79) or very strong (≥0.80).
RESULTS: The TEG 6s parameters reaction time, kinetic time and α-angle (CK, CRT and CKH assays) mostly showed strong correlations with the corresponding ROTEM parameters clotting time, clot formation time and α-angle (INTEM, EXTEM and HEPTEM assays). The exceptions were CRT reaction time vs. EXTEM clotting time, and CK α-angle vs. INTEM α-angle, which correlated moderately. Absolute values for many of these parameters showed significant differences between the two devices. Very strong correlations and similar absolute values were observed between TEG 6s maximum amplitude (CRT, CK and CKH assays) and ROTEM maximum clot firmness (EXTEM, INTEM and HEPTEM assays). Correlations were also very strong for functional fibrinogen maximum amplitude vs. FIBTEM maximum clot firmness and functional fibrinogen level vs. Clauss fibrinogen concentration, but absolute values were significantly different.
CONCLUSION: Strong to very strong correlations were observed between corresponding TEG 6s and ROTEM Sigma parameters. However, absolute values showed significant differences for most of the measurements.

Entities:  

Mesh:

Year:  2019        PMID: 31219873     DOI: 10.1097/EJA.0000000000001032

Source DB:  PubMed          Journal:  Eur J Anaesthesiol        ISSN: 0265-0215            Impact factor:   4.330


  8 in total

1.  A comparison between the TEG 6s and TEG 5000 analyzers to assess coagulation in trauma patients.

Authors:  Matthew D Neal; Ernest E Moore; Mark Walsh; Scott Thomas; Rachael A Callcut; Lucy Z Kornblith; Martin Schreiber; Akpofure Peter Ekeh; Adam J Singer; Lawrence Lottenberg; Michael Foreman; Susan Evans; Robert D Winfield; Michael D Goodman; Carl Freeman; David Milia; Noelle Saillant; Jan Hartmann; Hardean E Achneck
Journal:  J Trauma Acute Care Surg       Date:  2020-02       Impact factor: 3.697

Review 2.  Variations and obstacles in the use of coagulation factor concentrates for major trauma bleeding across Europe: outcomes from a European expert meeting.

Authors:  Vladimir Černý; Marc Maegele; Vanessa Agostini; Dietmar Fries; Santiago R Leal-Noval; Gábor Nardai; Giuseppe Nardi; Anders Östlund; Herbert Schöchl
Journal:  Eur J Trauma Emerg Surg       Date:  2021-01-05       Impact factor: 3.693

Review 3.  Thromboelastography for the Prevention of Perioperative Venous Thromboembolism in Orthopedics.

Authors:  Dejing Fan; Ziyao Ouyang; Yanping Ying; Shuangxia Huang; Pinyue Tao; Xiao Pan; Shuyu Lu; Qini Pan
Journal:  Clin Appl Thromb Hemost       Date:  2022 Jan-Dec       Impact factor: 2.389

4.  Operability of a Resonance-Based Viscoelastic Haemostatic Analyzer in the High-Vibration Environment of Air Medical Transport.

Authors:  Johannes Zipperle; Bernhard Ziegler; Herbert Schöchl; Wolfgang Voelckel; Christoph J Schlimp; Daniel Oberladstätter
Journal:  J Clin Med       Date:  2022-06-23       Impact factor: 4.964

Review 5.  Viscoelastic Hemostatic Assays: Moving from the Laboratory to the Site of Care-A Review of Established and Emerging Technologies.

Authors:  Jan Hartmann; Matthew Murphy; Joao D Dias
Journal:  Diagnostics (Basel)       Date:  2020-02-21

Review 6.  Point-of-Care Diagnostics in Coagulation Management.

Authors:  Sebastian D Sahli; Julian Rössler; David W Tscholl; Jan-Dirk Studt; Donat R Spahn; Alexander Kaserer
Journal:  Sensors (Basel)       Date:  2020-07-30       Impact factor: 3.576

Review 7.  Patient blood management during the COVID-19 pandemic: a narrative review.

Authors:  D M Baron; M Franchini; S M Goobie; M Javidroozi; A A Klein; S Lasocki; G M Liumbruno; M Muñoz; A Shander; D R Spahn; K Zacharowski; P Meybohm
Journal:  Anaesthesia       Date:  2020-05-06       Impact factor: 6.955

8.  The Role of TEG and ROTEM in Damage Control Resuscitation.

Authors:  Jason B Brill; Megan Brenner; Juan Duchesne; Derek Roberts; Paula Ferrada; Tal Horer; David Kauvar; Mansoor Khan; Andrew Kirkpatrick; Carlos Ordonez; Bruno Perreira; Artai Priouzram; Bryan A Cotton
Journal:  Shock       Date:  2021-12-01       Impact factor: 3.454

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.