| Literature DB >> 31216146 |
John K Taylor, Stanley W Smith.
Abstract
Apparatus and techniques are described for the titration of acids and bases by constant-current coulometry. The precision and accuracy of the method are indicated by titrations of benzoic acid, potassium acid phthalate, adipic acid, hydrochloric acid, and sodium carbonate. Standard deviations of ±0.003 percent have been obtained which indicate that the reliability of the method is equal to or exceeds that of classical analytical procedures.Entities:
Year: 1959 PMID: 31216146 PMCID: PMC5287197 DOI: 10.6028/jres.063A.008
Source DB: PubMed Journal: J Res Natl Bur Stand A Phys Chem ISSN: 0022-4332
Figure 1Coulometric titration cell.
Figure 2Electrical circuit diagram for coulometric titrations.
Figure 3Typical end-point determination.
Summary of results
| Determination number | Potassium hydrogen phthalate | Potassium hydrogen phthalate | Benzoic acid | Benzoic acid | Hydrochloric acid | Adipic acid | Adipic acid | Sodium carbonate |
|---|---|---|---|---|---|---|---|---|
| assay | assay | assay | assay | assay | assay | assay | assay | |
|
|
|
|
|
|
|
|
|
|
| % | % | % | % | % | % | % | ||
| 1 | 99.976 | 99.979 | 99.985 | 99.991 | 99.982 | 99.935 | 99.985 | 99.909 |
| 2 | 99.980 | 99.976 | 99.991 | 99.993 | 100.004 | 99.937 | 99.987 | 99.901 |
| 3 | 99.978 | 99.981 | 99.989 | 99.996 | 100.014 | 99.937 | 99.980 | 99.915 |
| 4 | 99.970 | 99.976 | 100.001 | 99.991 | 100.003 | 99.947 | 99.981 | 99.910 |
| 5 | 99.974 | 99.974 | 99.985 | 99.986 | 100.002 | 99.942 | 99.989 | 99.920 |
| 6 | …………………………………… | …………………………………… | …………………………………… | 99.998 | 99.999 | 99.955 | …………………………………… | …………………………………… |
| 7 | …………………………………… | …………………………………… | …………………………………… | …………………………………… | 99.990 | 99.951 | …………………………………… | …………………………………… |
| 8 | …………………………………… | …………………………………… | …………………………………… | …………………………………… | …………………………………… | 99.948 | …………………………………… | …………………………………… |
|
|
|
|
|
|
|
|
| |
| Mean | 99.976 | 99.977 | 99.990 | 99.994 | 99.999 | 99.945 | 99.984 | 99.911 |
| Standard deviation | 0.004 | 0.003 | 0.006 | 0.004 | 0.010 | 0.007 | 0.004 | 0.007 |
Timed by electric clock.
Timed by time interval meter.
Assay based on calculated acidity by the pressure-distillation method.
Once recrystallized from melt.
Twice recrystallized from melt.