Literature DB >> 31209920

Echocardiographic mimicker of thrombus on a mechanical aortic valve prosthesis due to cavitation: A paradoxical phenomenon of pressure recovery.

Zhenzhen Wang1, Chadi Ayoub1, Jeremy J Thaden1, Said Alsidawi1, Fletcher A Miller1, Lawrence J Sinak1, Charanjit S Rihal1, Rowlens M Melduni1.   

Abstract

We describe a case of a mass-like echocardiographic density on a mechanical prosthetic aortic valve. We initially suspected a thrombus vs vegetation on transthoracic echocardiography, but after transesophageal echocardiography, the density was subsequently determined to be cavitation by reviewing the initial images in slow motion.
© 2019 Wiley Periodicals, Inc.

Entities:  

Keywords:  aortic valve prosthesis; cavitation; mass; microbubbles; thrombus; vegetations

Mesh:

Year:  2019        PMID: 31209920      PMCID: PMC6849478          DOI: 10.1111/echo.14403

Source DB:  PubMed          Journal:  Echocardiography        ISSN: 0742-2822            Impact factor:   1.724


  8 in total

1.  Mechanisms of cavitation and the formation of stable bubbles on the Björk-Shiley Monostrut prosthetic heart valve.

Authors:  Christopher Bachmann; Vinayak Kini; Steven Deutsch; Arnold A Fontaine; John M Tarbell
Journal:  J Heart Valve Dis       Date:  2002-01

2.  Cavitation versus degassing: in vitro study of the microbubble phenomenon observed during echocardiography in patients with mechanical prosthetic cardiac valves.

Authors:  Grégoire Girod; Andres Jaussi; Cédric Rosset; Philippe De Werra; Félix Hirt; Lukas Kappenberger
Journal:  Echocardiography       Date:  2002-10       Impact factor: 1.724

3.  Echocardiographic dilemma: misleading thrombus appearance due to cavitation with intravascular hemolysis on a mechanical prosthetic valve.

Authors:  Ibrahim Akpinar; Muhammet Rasit Sayin; Turgut Karabag; Mehmet Emin Kalkan; Sait Mesut Dogan; Mustafa Aydin; Ibrahim C Haznedaroglu
Journal:  Can J Cardiol       Date:  2012-05-26       Impact factor: 5.223

Review 4.  Mechanical heart valve cavitation.

Authors:  Peter Johansen
Journal:  Expert Rev Med Devices       Date:  2004-09       Impact factor: 3.166

5.  Misleading echocardiographic diagnosis of a prosthetic heart valve vegetation due to the cavitation phenomenon.

Authors:  Ramón Suárez Arias; Isabel Piñero-Uribe; Francesc Carreras; Sandra Pujadas; Rubén Leta; Guillem Pons-Lladó
Journal:  Exp Clin Cardiol       Date:  2009

6.  Potential fluid mechanic pathways of platelet activation.

Authors:  Shawn C Shadden; Sahar Hendabadi
Journal:  Biomech Model Mechanobiol       Date:  2012-07-11

7.  Spontaneous echocardiographic microbubbles associated with prosthetic mitral valves: mechanistic insights from thrombolytic treatment results.

Authors:  Cihangir Kaymaz; Mehmet Ozkan; Nihal Ozdemir; Cevat Kirma; Ubeydullah Deligönül
Journal:  J Am Soc Echocardiogr       Date:  2002-04       Impact factor: 5.251

8.  A misinterpreted case of aorta prosthesis endocarditis: remember the phenomenon of microbubbles.

Authors:  Christian Johann Lerche; Ketil Jørgen Haugan; Jesper Irving Reimers; Nikolaj Ihlemann
Journal:  Echocardiography       Date:  2013-04-25       Impact factor: 1.724

  8 in total