El-Sayed M El-Emam1, Enas A Abd El Motlb1.
Abstract
CONTEXT: Some pediatric supraglottic airway devices have not been evaluated in a large perspective for blind intubation although they are validated as a conduit for fiberoptic-guided intubation.Entities:
Keywords: Air-Q; blind intubation; correction maneuvers; fiberoptic-guided intubation
Year: 2019 PMID: 31198243 PMCID: PMC6545951 DOI: 10.4103/aer.AER_42_19
Source DB: PubMed Journal: Anesth Essays Res ISSN: 2229-7685
Figure 1Consort flow diagram
Demographic data of the studied groups
| Group B ( | Group F ( | ||
|---|---|---|---|
| Sex (female/male) | 32/31 | 30/33 | 0.7 |
| Age (years) | 4.4±2.17 | 4.5±2.7 | 0.7 |
| Weight (kg) | 15.86±4.17 | 17.2±7.3 | 0.5 |
Data are presented as mean±SD or n. SD=Standard deviation
Pre- and postintubation hemodynamic variables of the studied groups
| Group B ( | Group F ( | ||
|---|---|---|---|
| HR (beat/min) | |||
| Preintubation | 128.66±5.45 | 137.5±6.13 | 0.0001** |
| Postintubation | 129.86±5.91 | 138.8±5.77 | 0.0001** |
| MBP (mmHg) | |||
| Preintubation | 64.73±6.19 | 66.46±8.72 | 0.2 |
| Postintubation | 66.73±6.60 | 68.53±8.1 | 0.2 |
| SpO2 (%) | |||
| Preintubation | 98.00±0.75 | 98.33±0.61 | 0.19 |
| Postintubation | 91.20±2.17 | 92.00±2.17 | 0.3 |
Data are presented as mean±SD. *P≤0.05 was considered statistically significant, SD=Standard deviation, HR=Heart rate, **P≤0.01 was considered as highly significant
Postextubation complications (croup/hoarseness)
| Group B ( | Group F ( | ||
|---|---|---|---|
| Post extubation croup | 9 (14) | 14 (22) | 0.2 |
| Post extubation hoarseness | 9 (14) | 17 (27) | 0.07 |
Figure 2Time to ventilate, time to intubate, and total time. Group B: Intubated through air-Q blindly. Group F: Intubated through air-Q using FOB. **P ≤ 0.01
Figure 3Success rates of intubation. Group B: Intubated through air-Q blindly. Group F: Intubated through air-Q using FOB