Fluoroscopy-Guided Suture Anchor Placement for Acetabular Labrum Repair.

Paul Kenneth Herickhoff1, Jason Eric Mascoe1.   

Abstract

A narrow safe angle exists for placing suture anchors into the acetabular rim during arthroscopic acetabular labrum repair. Intra-articular suture anchor placement has been reported, and cadaver studies have shown the highest risk of errant suture anchor placement at the anterior acetabular rim. We describe a technique using intraoperative fluoroscopy to guide suture anchor placement into the anterior and superior acetabular rim, which may reduce the risk of iatrogenic injury during arthroscopic acetabular labrum repair.

Entities:  

Year:  2019        PMID: 31194039      PMCID: PMC6551387          DOI: 10.1016/j.eats.2018.12.001

Source DB:  PubMed          Journal:  Arthrosc Tech        ISSN: 2212-6287


Hip arthroscopy is a technically demanding procedure with a steep learning curve and reoperation rates as high as 15.4%. For acetabular labrum repairs, improper suture anchor placement may result in damage to intrapelvic anatomical structures, iatrogenic cartilage injury, or intra-articular anchor placement. Previous publications have demonstrated fluoroscopy-guided suture anchor placement into the superior acetabulum. This surgical technique uses fluoroscopy to guide suture anchor placement in the anterior and superior acetabular rim, which may reduce the risk of iatrogenic injury during arthroscopic acetabular labrum repair.

Surgical Technique

The patient is positioned supine on a traction table (Steris OT 1000, Mentor, OH) with a large-diameter, padded perineal post. The dorsa of both feet are covered with abdominal pads wrapped in Coban (3M, St. Paul, MN) to reduce the risk of pressure sores and prevent slippage and secured in foam-lined boots. The nonsurgical leg is abducted 45° to 55°, allowing the C-arm to be brought between the legs. The surgical hip is flexed 10° to 15°, the foot is internally rotated approximately 25° to bring the axis of the femoral neck parallel to the floor, and the thigh is adducted to 0° against the perineal post. The surgical site is prepped, and the C-arm and surgical site are draped using sterile technique. A surgical time-out is called prior to making incision. Traction is applied to the operative extremity to open the superior joint space approximately 1 cm. The central compartment is accessed via standard anterolateral (AL) and midanterior (MA) portals, and an interportal capsulotomy is created using a straight beaver blade (Samurai; Stryker Sports Medicine, Greenwood, CO). Adjustable-length 8.0-mm Transport cannulas (Stryker Sports Medicine) are inserted through the AL and MA portals to facilitate instrument passage and switching the arthroscope between portals. A comprehensive diagnostic arthroscopy is performed. When a labrum tear is encountered, the location of the tear on the clockface of the acetabulum is recorded, and the tear is probed to evaluate whether the tear is amenable to repair (Fig 1, Video 1).
Fig 1

Right hip arthroscopic view from the anterolateral portal. The acetabular labrum is torn through the chondrolabral junction from the 12 o'clock to the 2 o'clock position.

Right hip arthroscopic view from the anterolateral portal. The acetabular labrum is torn through the chondrolabral junction from the 12 o'clock to the 2 o'clock position. Once the decision to repair the acetabular labrum has been made, a distal anterolateral accessory (DALA) portal is typically made to improve the angle of approach to the acetabular rim. The capsulolabral recess is developed with electrocautery (50° CoolCut wand, Arthrex, Naples, FL) and judicious use of the shaver, preserving the capsule for later repair. A 4.0-mm round, hooded burr can be used to remove any pincer lesions and gently debride the acetabular rim to stimulate a healing response. The C-arm is then brought in, and an anteroposterior (AP) image of the hip is taken. On an AP view of the hip, the x-ray beam is tangential to the superior acetabular dome, where the subchondral bone is easily seen as a curved, dark line, known as the sourcil. Depending on the patient's lumbar lordosis, pelvic tilt, and acetabular tilt, the most superior portion of the acetabulum may or may not be directly opposite the cotyloid fossa at the 12 o'clock position described by Philippon et al. For this example, assume the most cephalad portion of the acetabular rim is at the 12 o'clock position. To place a suture anchor at the 12 o'clock position, a straight drill guide for the Nanotack anchor (Stryker Sports Medicine) is placed through the DALA portal with the tip of the guide at the extracapsular labral insertion, 2.3 to 2.6 mm from the edge of the acetabular rim (Fig 2). The angle of the guide is adjusted so that the drill guide is parallel to or slightly divergent from the sourcil on fluoroscopy (Fig 3). A 1.4-mm pilot hole is then drilled while viewing the articular surface through the MA portal. Fluoroscopy is used to confirm the position of the drill bit in the subchondral bone (Fig 4). While keeping the drill guide in place, the drill is removed and a Nanotack suture anchor (Stryker Sports Medicine) is inserted and impacted past the laser line. The anchor is deployed by pulling on the purple tabs. The drill guide is removed, and the stability of the anchor is verified by pulling on the suture limbs. The suture limb closest to the labrum is then passed through the labrum using a curved suture passer (Nanopass; Stryker Sports Medicine), in either a labral base stitch or circumferential stitch pattern. The sutures are then tied and cut.
Fig 2

Right hip arthroscopic view from the midanterior portal. A straight drill guide enters the hip through the distal anterolateral accessory portal. The tip of the drill guide is located at the 12 o'clock position, 2.3 to 2.6 mm from the acetabular rim.

Fig 3

Anteroposterior fluoroscopic view of the right hip shows a straight drill guide, inserted from a distal anterolateral accessory portal and positioned at the 12 o'clock position on the acetabular rim, slightly divergent from the subchondral bone.

Fig 4

Anteroposterior fluoroscopic view of the right hip shows a 1.4-mm diameter, 17 mm long drill bit inserted through a distal anterolateral accessory portal into the acetabular rim at the 12 o'clock position, slightly divergent from the subchondral bone.

Right hip arthroscopic view from the midanterior portal. A straight drill guide enters the hip through the distal anterolateral accessory portal. The tip of the drill guide is located at the 12 o'clock position, 2.3 to 2.6 mm from the acetabular rim. Anteroposterior fluoroscopic view of the right hip shows a straight drill guide, inserted from a distal anterolateral accessory portal and positioned at the 12 o'clock position on the acetabular rim, slightly divergent from the subchondral bone. Anteroposterior fluoroscopic view of the right hip shows a 1.4-mm diameter, 17 mm long drill bit inserted through a distal anterolateral accessory portal into the acetabular rim at the 12 o'clock position, slightly divergent from the subchondral bone. More anteriorly located suture anchors can be placed by rotating the C-arm backward approximately 30° for each clockface position. The subchondral bone is visualized at the location where the x-ray beam is tangential to the acetabular rim (Fig 5). For example, to place an anchor at the 1 o'clock position, the C-arm is rotated backward 30° from the vertical (Fig 6). The subchondral bone at the 1 o'clock position is visualized on fluoroscopy, the drill guide (Fig 7) is positioned parallel or slightly divergent from the subchondral bone, and the pilot hole is drilled (Fig 8). The acetabular subchondral bone at the 2 and 3 o'clock positions can similarly be visualized by rotating the C-arm backward an additional 30° for each clockface position (Fig 9) and used to guide anchor placement (Figs 10 and 11). To place suture anchors at the 11 o'clock position, the C-arm is rotated 30° forward from the vertical, and suture anchors are placed in a similar fashion. After the labrum repair is complete, traction on the hip is released and the repair is inspected from the peripheral compartment (Fig 12).
Fig 5

This pictorial representation of the acetabulum depicts x-ray beams tangential to the acetabular rim at the 11 o'clock to 3 o'clock positions, enabling the surgeon to fluoroscopically visualize the subchondral bone at these locations.

Fig 6

Right hip arthroscopy room setup. The C-arm is rotated backward 30° to view the subchondral bone at the 1 o'clock position on the acetabulum.

Fig 7

Right hip arthroscopic view from the anterolateral portal. A straight drill guide enters the hip through the distal anterolateral accessory (DALA) portal. The tip of the drill guide is located at the 1 o'clock position, 2.3 to 2.6 mm from the acetabular rim.

Fig 8

Fluoroscopic view of the right hip with the C-arm rotated 30° backward, showing the drill bit slightly divergent from the subchondral bone at the 1 o'clock position on the acetabulum.

Fig 9

Right hip arthroscopy room setup. The C-arm is rotated backward 60° to view the subchondral bone at the 2 o'clock position on the acetabulum.

Fig 10

Fluoroscopic view of the right hip with the C-arm rotated 60° backward, showing the straight drill guide parallel to the subchondral bone at the 2 o'clock position on the acetabulum.

Fig 11

Fluoroscopic view of the right hip with the C-arm rotated 60° backward, showing the 1.4-mm diameter, 17 mm long drill bit parallel to the subchondral bone at the 2 o'clock position on the acetabulum.

Fig 12

Right hip arthroscopic view from the anterolateral portal. Inspection of the acetabular labrum repair with traction released shows excellent restoration of the suction seal between the labrum and femoral head.

This pictorial representation of the acetabulum depicts x-ray beams tangential to the acetabular rim at the 11 o'clock to 3 o'clock positions, enabling the surgeon to fluoroscopically visualize the subchondral bone at these locations. Right hip arthroscopy room setup. The C-arm is rotated backward 30° to view the subchondral bone at the 1 o'clock position on the acetabulum. Right hip arthroscopic view from the anterolateral portal. A straight drill guide enters the hip through the distal anterolateral accessory (DALA) portal. The tip of the drill guide is located at the 1 o'clock position, 2.3 to 2.6 mm from the acetabular rim. Fluoroscopic view of the right hip with the C-arm rotated 30° backward, showing the drill bit slightly divergent from the subchondral bone at the 1 o'clock position on the acetabulum. Right hip arthroscopy room setup. The C-arm is rotated backward 60° to view the subchondral bone at the 2 o'clock position on the acetabulum. Fluoroscopic view of the right hip with the C-arm rotated 60° backward, showing the straight drill guide parallel to the subchondral bone at the 2 o'clock position on the acetabulum. Fluoroscopic view of the right hip with the C-arm rotated 60° backward, showing the 1.4-mm diameter, 17 mm long drill bit parallel to the subchondral bone at the 2 o'clock position on the acetabulum. Right hip arthroscopic view from the anterolateral portal. Inspection of the acetabular labrum repair with traction released shows excellent restoration of the suction seal between the labrum and femoral head.

Discussion

A narrow safe angle exists for placing suture anchors into the acetabular rim during arthroscopic acetabular labrum repair. Hernandez and McGrath calculated a safe angle of 20.1° to 27.6° at the anterior, superior, and anterosuperior acetabular rim. Cadaveric studies have reported a 50% risk of acetabular fossa perforation with suture anchor placement at the 4 o'clock position, 4% to 5% risk of articular perforation at the 3 o'clock position, and a 7.69% risk of psoas tunnel (intrapelvic) perforation between the 2 and 4 o'clock positions. Improperly placed suture anchors can be seen on postoperative MRI scans and can be a source of pain after surgery. Previous publications have illustrated the use of fluoroscopy to place suture anchors into the superior acetabulum; however, fluoroscopy-guided suture anchor placement at other locations around the acetabular rim, to our knowledge, has not previously been described. The surgical technique described herein provides a safety check when drilling and inserting suture anchors into the anterior and superior acetabular rim and relies on the simple concept of using tangential fluoroscopic images to visualize the subchondral bone at the locations where suture anchors are to be placed (Fig 5). Similar to the sourcil being visible on an AP fluoroscopic image of the hip because it is tangential to the x-ray beam, the subchondral bone at other locations of the acetabulum are seen by rotating the C-arm backward and forward. Matsuda et al. reported 5 cases of suture anchor–induced chondral damage in the hip. These authors suggested several techniques to prevent this complication, including using small-diameter, short suture anchors and visualizing the central compartment while drilling and placing suture anchors. In an anatomical study, Gereli et al. showed the risk of perforation to the pelvic cavity started with when drilling past 20 mm. We agree with these recommendations; we routinely view the acetabular cartilage while drilling, and our preferred Nanotack suture anchor has a small, 1.4-mm diameter drill bit, drilling to a depth of 17 mm. In a cadaveric study, Stanton and Banffy concluded that the DALA portal allows suture anchors to be placed at a greater distance from the articular surface than the AL and MA portals. Conversely, cadaveric studies by Degen et al. and Foster et al. reported no statistically significant differences in articular surface or psoas tunnel perforation when hip anchors were placed in the anterior acetabular rim with a straight drill guide from the MA or the DALA portals. Our fluoroscopy-guided technique can instruct surgeons when suture anchors can be safely placed through the MA and AL portals and when it is necessary to create a DALA portal to improve the angle of approach to the acetabular rim. The advantages of this surgical technique include that it is easy to learn, can be used with straight or curved drill guides, and does not require any additional equipment. A limitation of this technique is that radio-opaque metal hinges found in many commercially available hip arthroscopy traction tables may obstruct fluoroscopic views of the posterior acetabular rim. Therefore, for posterior labrum tears, fluoroscopy-guided suture anchor placement may not be possible. Since the acetabular labrum is most commonly torn in the anterosuperior quadrant, however, this technique is germane to most hip arthroscopy labrum repairs. A second limitation is the small amount of additional radiation to operating room staff resulting from the increased use of fluoroscopy. A summary of the advantages and limitations of this technique is listed in Table 1.
Table 1

Advantages and Limitations of Fluoroscopy-Guided Suture Anchor Placement

AdvantagesLimitations
Reduces the risk of intrapelvic and intra-articular drilling and suture anchor placementMetal in traction table may obstruct fluoroscopic views of posterior acetabular rim
Applicable in most acetabular labral repairsAdditional radiation to operating room staff
No additional equipment required
Easy to learn
Can instruct the surgeon whether a DALA portal or curved drill guides are required
Advantages and Limitations of Fluoroscopy-Guided Suture Anchor Placement
  10 in total

1.  Are pelvic anatomical structures in danger during arthroscopic acetabular labral repair? Definition of safe bone depth.

Authors:  Arel Gereli; Baris Kocaoglu; Kerem Tekin Ulku; Abdul Veli Ismailoglu; Sena Silay; Mustafa Karahan
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2015-09-29       Impact factor: 4.342

Review 2.  Arthroscopic management of femoroacetabular impingement: osteoplasty technique and literature review.

Authors:  Marc J Philippon; Allston J Stubbs; Mara L Schenker; R Brian Maxwell; Reinhold Ganz; Michael Leunig
Journal:  Am J Sports Med       Date:  2007-04-09       Impact factor: 6.202

3.  Safe angle for suture anchor insertion during acetabular labral repair.

Authors:  Jaime D Hernandez; Brian E McGrath
Journal:  Arthroscopy       Date:  2008-10-10       Impact factor: 4.772

4.  Safety of Hip Anchor Insertion From the Midanterior and Distal Anterolateral Portals With a Straight Drill Guide: A Cadaveric Study.

Authors:  Ryan M Degen; Lazaros Poultsides; Stephanie W Mayer; Angela E Li; Struan H Coleman; Anil S Ranawat; Danyal H Nawabi; Bryan T Kelly
Journal:  Am J Sports Med       Date:  2016-12-07       Impact factor: 6.202

5.  Anatomy, histologic features, and vascularity of the adult acetabular labrum.

Authors:  R M Seldes; V Tan; J Hunt; M Katz; R Winiarsky; R H Fitzgerald
Journal:  Clin Orthop Relat Res       Date:  2001-01       Impact factor: 4.176

6.  Safe Angle of Anchor Insertion for Labral Repair During Hip Arthroscopy.

Authors:  Michael Stanton; Michael Banffy
Journal:  Arthroscopy       Date:  2016-04-29       Impact factor: 4.772

Review 7.  MRI Evaluation of Femoroacetabular Impingement After Hip Preservation Surgery.

Authors:  Angela E Li; Shari T Jawetz; Harry G Greditzer; Alissa J Burge; Danyal H Nawabi; Hollis G Potter
Journal:  AJR Am J Roentgenol       Date:  2016-05-25       Impact factor: 3.959

8.  Defining the Learning Curve for Hip Arthroscopy: A Threshold Analysis of the Volume-Outcomes Relationship.

Authors:  Nabil Mehta; Peter Chamberlin; Robert G Marx; Chisa Hidaka; Yile Ge; Danyal H Nawabi; Stephen Lyman
Journal:  Am J Sports Med       Date:  2018-01-16       Impact factor: 6.202

9.  Anchor-induced chondral damage in the hip.

Authors:  Dean K Matsuda; Srino Bharam; Brian J White; Nicole A Matsuda; Marc Safran
Journal:  J Hip Preserv Surg       Date:  2015-01-29

10.  Safe suture anchor insertion for anterior and posterior hip labral repair.

Authors:  Andrew D Foster; John Ryan; Thomas Ellis; James Flom
Journal:  J Hip Preserv Surg       Date:  2015-04-29
  10 in total
  1 in total

1.  Fluoroscopy-Guided Suture Anchor Placement Yields Excellent Accuracy for Arthroscopic Acetabular Labral Repair: A Cadaveric Study.

Authors:  Paul K Herickhoff; Matthew Widner; Jason Mascoe; Wayne J Sebastianelli
Journal:  Arthrosc Sports Med Rehabil       Date:  2021-08-19
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.