Histology laboratory deficiencies since 2014.
| Deficiency | Phase | Description |
|---|---|---|
| GEN.55900 | II | Lack of consistent adherence to semiannual competency training of new employees. |
| COM.30300 | II | Some reagents not labeled. |
| COM.30450 | II | Parallel testing for lot confirmation records are incomplete for special stains. |
| ANP.8216 | II | Formalin and xylene exposure monitoring was not done for new personnel. |
| ANP.21450 | II | No documentation or validation records for some special stains. |
| ANP.22956 | II | No validation records for in situ hybridization testing for kappa and lambda. |
| ANP.22964 | II | No documentation of controls used for EBV testing by in situ hybridization. |
| ANP.23410 | II | No decontamination policy for cryostat in histology lab. |
| ANP.21850 | II | No documentation of utilization of immunofluorescent control. |
| ANP.22979 | II | HER-2 testing is not being repeated on a separate block in cases of histologic discordance. |
| ANP.22500 | II | Missing quality control records for buffers for the months of July and August. |
| ANP.08216 | II | Formalin and xylene exposure monitoring was not done for rotating residents. |
| ANP.24300 | II | No policy for CJD special tissue handling. |
| ANP.21050 | II | Some blocks with handwritten labels did not have two patient identifiers. |
| ANP.23410 | II | No decontamination policy for cryostat in histology lab. |
| ANP.22985 | I | No procedure for decalcified tissue for predictive marker testing. |
| COM.04000 | II | One part of quality assurance policy has not been followed. No records reflecting conformance with the program as designed. |