Hiroshi Hashimoto1,2, Toshiki Kobayashi3, Fan Gao4, Masataka Kataoka1, Kuniharu Okuda1.
Abstract
INTRODUCTION: The alignment of a prosthesis is clinically determined based on observations by clinicians and the subjective perception of amputees during gait. However, this process has been reported to be unreliable. Socket reaction moment has been reported to be significantly impacted by alignment changes, but the impact of these alignment changes on other gait parameters is unclear. The aim of this study was to investigate the effects of coronal alignment changes of a transtibial prosthesis on socket reaction moment, spatiotemporal parameters, and perceptions of alignment during gait in amputees.Entities:
Keywords: Amputee; load cell; prosthesis; prosthetist; walk
Year: 2018 PMID: 31191953 PMCID: PMC6531796 DOI: 10.1177/2055668318795402
Source DB: PubMed Journal: J Rehabil Assist Technol Eng ISSN: 2055-6683
Demographic information of the participants.
| ID | Gender | Age | Height (cm) | Body mass (kg) | Cause of amputation | Amputation side | Length of residual limb (cm) | Duration of amputation (years) | Distance from the end of the socket to the floor (cm) | Prosthetic foot |
|---|---|---|---|---|---|---|---|---|---|---|
| 1 | M | 45 | 188 | 90 | Trauma | Right | 14 | 10 | 25 | Vari-Flex |
| 2 | M | 41 | 180 | 74 | Trauma | Left | 11 | 14 | 28 | LP Vari-Flex |
| 3 | M | 57 | 172 | 63 | Tumor | Right | 18 | 48 | 23 | LP Vari-Flex |
| 4 | M | 74 | 160 | 52 | Diabetes | Left | 12 | 7 | 20 | FlexFoot Balance |
| 5 | M | 67 | 170 | 61 | Diabetes | Left | 11 | 11 | 26 | Aspire |
| 6 | M | 67 | 164 | 57 | Trauma | Left | 10 | 22 | 27 | LP Vari-Flex |
| 7 | F | 40 | 153 | 89 | Tumor | Left | 9 | 5 | 22 | Aspire |
| 8 | M | 42 | 172 | 84 | Trauma | Right | 19 | 32 | 21 | LP Vari-Flex |
| 9 | M | 35 | 177 | 70 | Trauma | Left | 15 | 5 | 25 | LP Vari-Flex |
Figure
1.An experimental prosthesis used in this study: (a) sagittal view and (b) coronal view.
The spatiotemporal parameters of gait under coronal angulation and translation alignment changes.
| Gait parameters | Prosthetic/ non-prosthetic side | Angulation | Significance | ||||
|---|---|---|---|---|---|---|---|
| Step time (s) | Prosthetic side | 0.60 ± 0.08 | 0.59 ± 0.06 | 0.56 ± 0.05 | 0.57 ± 0.06 | 0.58 ± 0.04 | N.S. |
| Non-prosthetic side | 056 ± 0.09 | 0.56 ± 0.04 | 0.56 ± 0.05 | 0.56 ± 0.03 | 0.60 ± 0.06 | N.S. | |
| Step length (m) | Prosthetic side | 0.59 ± 0.14 | 0.63 ± 0.09 | 0.62 ± 0.08 | 0.62 ± 0.11 | 0.60 ± 0.12 | N.S. |
| Non-prosthetic side | 0.62 ± 0.07 | 0.60 ± 0.05 | 0.59 ± 0.05 | 0.60 ± 0.05 | 0.57 ± 0.06 | N.S. | |
| Single limb support time (s) | Prosthetic side | 0.40 ± 0.07 | 0.40 ± 0.03 | 0.41 ± 0.04 | 0.40 ± 0.04 | 0.43 ± 0.04 | N.S. |
| Non-prosthetic side | 0.46 ± 0.07 | 0.45 ± 0.05 | 0.44 ± 0.04 | 0.44 ± 0.03 | 0.45 ± 0.03 | N.S. | |
| Cadence (steps/min) | 105 ± 11 | 106 ± 8 | 107 ± 10 | 107 ± 7 | 103 ± 7 | N.S. | |
| Gait speed (m/s) | 1.04 ± 0.13 | 1.10 ± 0.16 | 1.08 ± 0.16 | 1.09 ± 0.12 | 1.00 ± 0.15 | N.S. | |
| Step width (mm) | 169 ± 52 | 167 ± 35 | 136 ± 41 | 114 ± 48 | 112 ± 47 | N.S. | |
| Translation | |||||||
| 10 mm medial translation | 5 mm medial translation | Baseline | 5 mm lateral translation | 10 mm lateral translation | |||
| Step time (s) | Prosthetic side | 0.55 ± 0.05 | 0.58 ± 0.05 | 0.56 ± 0.05 | 0.57 ± 0.04 | 0.57 ± 0.03 | N.S. |
| Non-prosthetic side | 0.55 ± 0.04 | 0.57 ± 0.05 | 0.56 ± 0.05 | 0.56 ± 0.05 | 0.58 ± 0.05 | N.S. | |
| Step length (m) | Prosthetic side | 0.62 ± 0.11 | 0.66 ± 0.10 | 0.62 ± 0.08 | 0.65 ± 0.08 | 0.62 ± 0.11 | N.S. |
| Non-prosthetic side | 0.59 ± 0.07 | 0.60 ± 0.05 | 0.59 ± 0.05 | 0.63 ± 0.05 | 0.59 ± 0.05 | N.S. | |
| Single limb support time (s) | Prosthetic side | 0.41 ± 0.02 | 0.40 ± 0.04 | 0.41 ± 0.04 | 0.41 ± 0.05 | 0.41 ± 0.05 | N.S. |
| Non-prosthetic side | 0.44 ± 0.03 | 0.44 ± 0.04 | 0.44 ± 0.04 | 0.43 ± 0.02 | 0.44 ± 0.03 | N.S. | |
| Cadence (steps/min) | 107 ± 7 | 106 ± 9 | 107 ± 10 | 107 ± 7 | 105 ± 8 | N.S. | |
| Gait speed (m/s) | 1.08 ± 0.16 | 1.10 ± 0.14 | 1.08 ± 0.16 | 1.13 ± 0.10 | 1.05 ± 0.15 | N.S. | |
| Step width (mm) | 156 ± 45 | 153 ± 55 | 136 ± 41 | 138 ± 49 | 132 ± 44 | N.S. | |
N.S., no significant differences.
Figure
2.(a) Mean coronal SRM across participants in stance and (b) mean maximum varus moment across participants with standard deviation under angulation changes. An asterisk (*) indicates significant differences (P < 0.05).
Abd: abduction; Add: adduction; SRM: socket reaction moment.
Figure
3.(a) Mean coronal SRM across participants in stance and (b) mean maximum varus moment across participants with standard deviation under translation changes. An asterisk (*) indicates significant differences (P < 0.05).
Lat: lateral translation; Med: medial translation; SRM: socket reaction moment.
Figure
4.Mean SRM under each alignment condition. The shaded area indicates standard deviations during stance (per % stance).
SRM: socket reaction moment.
Perception of coronal angulation and translation alignment changes.
| Perception | Angulation | Significance: N.S. | ||
|---|---|---|---|---|
| 6° adduction | 3° adduction | 3° abduction | 6° abduction | |
| Perceived correctly | 4 | 4 | 6 | 6 |
| Perceived incorrectly | 2 | 0 | 1 | 1 |
| Uncertain | 3 | 5 | 2 | 2 |
| Translation | ||||
| 10-mm medial translation | 5-mm medial translation | 5-mm lateral translation | 10-mm lateral translation | |
| Perceived correctly | 3 | 3 | 6 | 4 |
| Perceived incorrectly | 2 | 1 | 1 | 2 |
| Uncertain | 4 | 5 | 2 | 3 |
N.S., no significant differences. Note: Each number indicates the number of responses in perception of coronal alignment changes among the participants.
Figure
5.An illustration explaining the horizontal and vertical displacement of the prosthetic foot under angulation alignment changes.