INTRODUCTION: Trans-femoral amputees are at risk of musculoskeletal problems that are in part caused by loading asymmetry during activities, such as prolonged standing, particularly on uneven or sloped ground. METHODS: Four prosthetic conditions were tested; microprocessor knee 'standing support' mode activated (ON) and deactivated (OFF), combined with a rigidly attached foot (RA) and with an articulating, hydraulic ankle-foot (HA). Five trans-femoral amputees and five able-bodied controls were measured using a motion capture system and a force plate while standing, facing down a 5° slope. Ground reaction force distributions and centre-of-pressure root-mean-square (COP RMS) were calculated as outcome measures. RESULTS: Compensatory kinematic adjustments were observed for RA conditions but not for HA conditions. HA-OFF reduced ground reaction force degree-of-asymmetry for all five amputees, compared to RA-OFF. RA-ON reduced ground reaction force degree-of-asymmetry for four amputees, compared to RA-OFF. In terms of balance, the HA conditions reduced the mean inter-limb COP RMS by 24-25% compared to equivalent RA conditions, while ON conditions reduced it by 9-11%, compared to equivalent OFF conditions. CONCLUSIONS: It is important to consider both prosthetic knee and ankle technologies when prescribing devices to trans-femoral amputees. The combination of hydraulic ankle and knee standing support technologies produced outcomes closest to normal biomechanics.
INTRODUCTION: Trans-femoral amputees are at risk of musculoskeletal problems that are in part caused by loading asymmetry during activities, such as prolonged standing, particularly on uneven or sloped ground. METHODS: Four prosthetic conditions were tested; microprocessor knee 'standing support' mode activated (ON) and deactivated (OFF), combined with a rigidly attached foot (RA) and with an articulating, hydraulic ankle-foot (HA). Five trans-femoral amputees and five able-bodied controls were measured using a motion capture system and a force plate while standing, facing down a 5° slope. Ground reaction force distributions and centre-of-pressure root-mean-square (COP RMS) were calculated as outcome measures. RESULTS: Compensatory kinematic adjustments were observed for RA conditions but not for HA conditions. HA-OFF reduced ground reaction force degree-of-asymmetry for all five amputees, compared to RA-OFF. RA-ON reduced ground reaction force degree-of-asymmetry for four amputees, compared to RA-OFF. In terms of balance, the HA conditions reduced the mean inter-limb COP RMS by 24-25% compared to equivalent RA conditions, while ON conditions reduced it by 9-11%, compared to equivalent OFF conditions. CONCLUSIONS: It is important to consider both prosthetic knee and ankle technologies when prescribing devices to trans-femoral amputees. The combination of hydraulic ankle and knee standing support technologies produced outcomes closest to normal biomechanics.
Impaired balance and a higher risk of falling are common problems amongst
amputees.[1-6] Studies have shown that up to
58% of lower limb amputees will trip or fall at least once a year.[6] During dynamic activities, such as walking, a lack of ankle dorsiflexion
motion can affect foot clearance, which has been correlated with the likelihood of falling.[7] Due to a loss of mass at the legs, lower limb amputees, and particularly
trans-femoral amputees (TFA), have a higher more decentralised centre-of-mass. Lower
limb amputees are less stable during static activities, such as standing, than
able-bodied controls.[2,8,9]Prosthetists are trained to align prostheses, so as to manipulate the position of the
components relative to the residual joints when standing and walking on level ground
to achieve satisfactory biomechanical performance. These adjustments influence the
external moments applied to the joints. However, external factors can have a
significant influence on the moments generated. For example, if the amputee is
standing on non-level ground, altered joint moments are generated at the ankle and
the proximal joints, due to changes in foot contact loading. Undesirable moments
generated can disturb balance and will have to be resisted through muscle action to
maintain a static equilibrium position, which would be tiring and uncomfortable for
the amputee.As a consequence, TFAs tend to increase their reliance on the sound limb for
support.[2,10,11] This asymmetry of loading has been linked to a number of
amputation associated comorbidities, including the prevalence of lower back
pain[12,13] and
osteoarthritis.[13-15] Studies have
cited back pain rates amongst lower limb amputees as between 48% and 71%[12,13,16-18] and up to 81% for TFAs, specifically.[19] This problem does not take long to materialise with 60% of lower limb
amputees reporting moderate to extreme back pain occurring within the first two
years after amputation.[19] The cited epidemiology of osteoarthritis among amputees varies between
sub-populations and specific joints[13-15,20-22] but can be approximated to two
to three times greater than that of the general population.[22]Recognising these needs, prosthetic design engineers have sought ways to provide
greater assistance during standing through advanced technology. Recent generations
of microprocessor prosthetic knees (MPKs) have incorporated ‘standing support’
functions, whereby when the limb detects the transition to standing activity, the
resistance to knee flexion is increased, allowing greater weight bearing on the
prosthetic side and, in turn, relieving excessive loading of the sound limb.
However, there is a lack of published scientific evidence to confirm the user
benefits of this functionality. At the ankle-foot complex, hydraulic ankles (HA)
have been developed, which exhibit viscoelastic behaviour like muscle. These allow a
variable equilibrium position and self-align so that they can comply with changes to
ground inclination without generating internal moments, reducing the requirement for
kinematic compensations compared to rigid ankle (RA) devices.[23]The purpose of this study was to investigate the efficacy of advanced prosthetic
componentry with respect to their effects on inter-limb load distribution and
balance ability. The study sought to evaluate MPK standing support functionality, as
well as to quantify the relative benefits of MPKs and hydraulic ankles, when
standing on sloped ground.
Methods
Prosthetic components and test conditions
The influences of two different prosthetic conditions were evaluated in this
study. One of these conditions was the ankle-foot device. This included an
energy-storage-and-return foot rigidly attached to the prosthetic pylon (RA –
EspritI, Endolite, Basingstoke, UK), while the other was a
hydraulic ankle with a torsional adaptor (HA – EchelonVTII, Endolite,
Basingstoke, UK). The second condition change was at the prosthetic knee. The
device was a microprocessor knee (Orion3III, Endolite, Basingstoke,
UK) with a standing support mode functionality. Upon detecting that the user is
standing, the hydraulic resistance to knee flexion is increased, to encourage
greater weight bearing on the prosthetic side without the knee buckling. The two
knee test conditions were standing support activated (ON) and deactivated (OFF).
No other changes were made to the knee that could affect kinematic or kinetic
parameters (e.g. prosthetic alignment) meaning that the differences observed
could be solely attributed to the effect of the standing support mode.
Consequently, there were four prosthetic conditions tested: (1) rigid ankle
without standing support (RA-OFF), (2) rigid ankle with standing support
(RA-ON), (3) hydraulic ankle without standing support (HA-OFF) and (4) hydraulic
ankle with standing support (HA-ON).
Participants
Five unilateral, TFA volunteered to participate in this study, each giving
informed, verbal consent. An ethical review ensured that the study protocol
complied with the tenets of the Declaration of Helsinki and the participants
were given the option of continued use of the advanced devices after the
completion of the study, should they wish to. Each participant was healthy, with
no comorbidities that might detrimentally influence balance control. They are
presented here as a case series, so as to eliminate inter-subject variability
due to external factors, such as prosthetic alignment. Each had a minimum of
four years’ experience with MPKs and both hydraulic and fixed ankle-feet
devices. The characteristics of each of the amputee participants are given in
Table 1.
Table 1.
Characteristics of the amputee participants.
ID
Sex
K level
SIGAM grade
Age (years)
Mass (kg)
Amputated side
Habitual ankle
Habitual knee
TF1
Male
K3
F
63
63
Left
EchelonVT
Orion3
TF2
Female
K2
E
48
48
Left
EchelonVT
Orion3
TF3
Male
K3
F
29
80
Right
EchelonVT
KX06
TF4
Male
K3
F
29
105
Left
EchelonVT
Orion3
TF5
Male
K3
F
39
90
Left
EchelonVT
Genium X3
Characteristics of the amputee participants.In addition to the amputee participants, five able-bodied participants
(27.4 ± 2.9 years, 66.8 ± 10.3 kg) volunteered to provide a comparison with
able-bodied biomechanics. Each of these participants gave informed, verbal
consent.
Gait lab setup
Body kinematics were captured using a Codamotion system (Charnwood Dynamics,
Leicestershire, UK) and ground kinetics were measured using a Kistler force
place (Kistler Group, Winterthur, Switzerland). The cameras collected data at a
frequency of 100 Hz and the force plate had an acquisition frequency of 500 Hz.
These data were used to calculate means and standard deviations of kinematic and
kinetic parameters.A six-degree-of-freedom (6 DoF) marker model was used to track the movement of
body segments.[24,25] For the amputee volunteers, the lateral and medial
prosthetic knee pivots replaced the femoral epicondyle virtual markers defining
the knee axis. At the ankle, the lateral and medial pivot points of the
hydraulic body replaced the malleoli virtual markers defining the ankle axis.
There is precedence in previous prosthetics literature for adapting the marker
model for amputee participants in this way.[26,27] The rigidly attached ESR
foot and the hydraulic foot have similar geometry (aside from the hydraulic
body), enabling a likewise approximation of the ankle joint axis between the
feet tested (Figure 1).
Figure 1.
The two prosthetic ankle-foot devices used in this study; a rigidly
attached, energy-storage-and-return foot (RA – Esprit, left) and a
hydraulic ankle-foot (HA – EchelonVT, right). The red circles
indicate the equivalent locations used to define the ‘ankle’ axis
with virtual markers.
The two prosthetic ankle-foot devices used in this study; a rigidly
attached, energy-storage-and-return foot (RA – Esprit, left) and a
hydraulic ankle-foot (HA – EchelonVT, right). The red circles
indicate the equivalent locations used to define the ‘ankle’ axis
with virtual markers.
Data collection
Each participant wore tight-fitting shorts and t-shirt for the data collection
session to reduce marker mounting movement artefacts and to avoid marker
occlusions. The testing took place on a 5° ramp, with the force plate integrated
so that its upper surface was flush with that of the surrounding walkway. Facing
down the ramp, each participant was asked to step onto the force plate (one foot
contacting at a time). This procedure was repeated until three ‘clean’ trials
had been completed on each limb. A clean trial was defined as one where the
entirety of the footprint of the tested limb was within the boundary of the
force plate and none of the footprint of the contralateral foot contacted the
plate. Data were recorded for 14 seconds per trial, so three repetitions on each
limb meant 42 s of data in total. This was repeated for each of the four
prosthetic conditions. Prior to data collection with each new prosthetic
condition, a period of 30 minutes acclimatisation was permitted for the
participant to become accustomed to the changes in behaviour of the limb. The
well-being of the participants was paramount so testing would only begin once
both the participant and a senior prosthetist were satisfied with their ability
to perform the protocol safely with the specific prosthetic condition. Both
force and marker data were captured so that any kinematic compensation could be
observed. The order in which these conditions were performed was randomised. Two
experimenters were present: the lead experimenter, who collected the data, and
the assistant experimenter, who would change the knee condition. Both
participants and the lead experimenter were blinded to the knee condition. Such
blinding was not possible for the foot condition.
Data processing and analysis
To exclude initial force spikes and variability resulting from the movement onto
the plate, data were only used after the heel marker velocity was below 30 mm/s.
This was used to define the point at which the foot was ‘static’ on the force
plate. All kinematic and kinetic measurements were calculated ‘per second’ to
observe the change and variability of parameters over time. Therefore, for each
tested condition, there were 42 individual measurements, from which the mean and
standard deviation values were calculated.Degree-of-asymmetry (DOA) was used in order to quantify the difference between
sound and prosthetic limbs in terms of ground reaction force (GRF). The GRF was
measured with respect to the sound limb and with respect to the prosthetic limb.
As shown in equation (1), DOA is calculated as the ratio of the difference of
these two parameters, to their sum. A value of zero indicates perfect symmetry,
a positive value shows the parameter is greater on the sound limb and a negative
value shows it’s greater on the prosthetic limb.Since there was only a single force plate available, trials that measured the
sound limb and those that measured the prosthetic limb were recorded
asynchronously, rather than simultaneously. Therefore, in order to calculate
DOA, the mean values were used for each prosthetic condition.In order to quantify the effect of the prosthetic condition on balance ability,
the centre-of-pressure root mean square (COP RMS) was calculated based on the
method used by Feick et al.[28] used in a previous amputee study. Briefly, the COP RMS calculates the
mean displacement of COP from its overall mean position. Better balance ability
is thought to be indicated by a lower COP deviation and consequently a reduced
COP RMS. This calculation is given in equation (2), where N is
the total number of samples within each second of the collected data,
is the vector position of COP of the nth
sample, is the mean vector position of COP of all N
samples, while denotes the scalar distance of nth COP
position from the mean.
Statistical analysis
Statistically significant differences were identified using paired t-tests. For
each amputee, data were compared between foot conditions with the same knee
condition (i.e. RA-OFF vs. HA-OFF) and between knee conditions with the same
foot condition (i.e. RA-OFF vs RA-ON). The normality of the data was
investigated with Shapiro–Wilk tests. Where the data were found to not be
normally distributed, the significance of any identified changes was further
validated using non-parametric Wilcoxon tests. Due to the large number of
comparisons being made, in order to avoid a type I error, a false discovery rate
(FDR) controlling method was employed. This method was chosen ahead of the
Bonferroni correction as it is less susceptible to type II errors. All
statistical analyses were conducted using R statistical software (The R
Foundation, Vienna, Austria).
Results
Kinematic compensations
In terms of joint kinematics, while there were some significant differences for
individuals between OFF and ON conditions (e.g. RA-OFF vs. RA-ON or HA-OFF vs.
HA-ON), no one of these differences was consistent across all the amputees.
There were, however, consistent joint kinematic differences when standing with
an RA and with an HA (Figure
2). In order to achieve ‘foot-flat’, with the RA-OFF, there was a
mean increase in knee flexion of 5.4° (p < 0.001) and a mean increase in hip
flexion of 3.1° (p < 0.001), compared to the HA-OFF condition. Differences
were also observed with standing support mode active across ankle conditions,
with RA-ON exhibiting a mean increase in knee flexion of 4.7° (p < 0.001) and
a mean increase in hip flexion of 2.3° (p < 0.001), compared to HA-ON.
Figure 2.
The posture of TF3 when standing with (a) a rigid ankle-foot (RA) and
(b) a hydraulic ankle-foot (HA). Red lines illustrate body segment
orientations.
The posture of TF3 when standing with (a) a rigid ankle-foot (RA) and
(b) a hydraulic ankle-foot (HA). Red lines illustrate body segment
orientations.
Kinetics: Bodyweight distribution
Figure 3(a) shows the GRF
distribution for the rigid ankle conditions (RA-OFF vs. RA-ON). Four of the
amputees showed a significant increase in the GRF under the prosthetic foot of
up to 22% with RA-ON compared to RA-OFF (p < 0.01), while the other one did
not show a change. All five saw a decrease in GRF under the sound foot of
between 4% and 13% with RA-ON (p ≤ 0.03). Four of the five amputees had a DOA
closer to zero with RA-ON compared to RA-OFF (Figure 3(b)), with the fifth showing
little change (TF4).
Figure 3.
The mean (a) ground reaction force (GRF) and (c) ground reaction
force component parallel to the ground (GRFx) for each participant,
under the prosthetic foot (solid) and the sound foot (striped) when
‘standing support’ was switched off (grey) and on (black). The error
bars indicate ± one standard deviation. Values were measured with a
rigid ankle (RA). Significant changes are marked with asterisks. The
background horizontal lines and shaded areas indicate the mean ± one
standard deviation ranges for the GRFs of the dominant (solid line)
and non-dominant (dashed line) limbs of the able-bodied (AB) control
participants. Also shown are the degree-of-asymmetry values for (b)
GRF and (d) GRFx, for each of the participants when ‘standing
support’ was switched off (horizontal stripes) and on (vertical
stripes).
The mean (a) ground reaction force (GRF) and (c) ground reaction
force component parallel to the ground (GRFx) for each participant,
under the prosthetic foot (solid) and the sound foot (striped) when
‘standing support’ was switched off (grey) and on (black). The error
bars indicate ± one standard deviation. Values were measured with a
rigid ankle (RA). Significant changes are marked with asterisks. The
background horizontal lines and shaded areas indicate the mean ± one
standard deviation ranges for the GRFs of the dominant (solid line)
and non-dominant (dashed line) limbs of the able-bodied (AB) control
participants. Also shown are the degree-of-asymmetry values for (b)
GRF and (d) GRFx, for each of the participants when ‘standing
support’ was switched off (horizontal stripes) and on (vertical
stripes).The breakdown of the GRF into linear components provided further insight. A
particular focus was given to the axis parallel to the standing surface (GRFx),
which was the direction that was most influenced by the adoption of compensatory
standing postures. As Figure
3(c) shows, all five amputee participants increased load bearing
under the prosthetic foot (p < 0.001) and reduced loading under the sound
foot (p < 0.001), within the ranges 17–54% and 12–50%, respectively, with
RA-ON. Three of the amputees had DOA values closer to zero with RA-ON compared
to RA-OFF, while the other two (TF2 and TF4) showed greater prosthetic side than
sound side loading (Figure
3(d)).The effects of standing support were less pronounced within the hydraulic ankle
conditions (HA-OFF vs. HA-ON). In terms of GRF, no consistent pattern across
subjects was observed, with two presenting a significant decrease in prosthetic
loading with HA-ON (p ≤ 0.02) but the others presenting no change. For the HA-ON
condition compared to HA-OFF, two amputees presented a significant decrease in
GRF loading on the sound limb (p ≤ 0.007), one presented a trend towards a
significant decrease (p = 0.052), one saw a significant increase (p = 0.03) and
the final participant presented no change.Figure 4(a) shows both
the GRF distributions when using an RA and an HA, with standing support off
(RA-OFF vs HA-OFF). GRF under the prosthetic foot increased significantly for
four of the amputees by 7–24% (p < 0.001) with HA-OFF. TF4 presented a
significant 3% reduction in GRF with HA-OFF (p = 0.02) but since, for this
participant, prosthetic loading was higher than sound side loading, this change
improved the DOA (Figure
4(b)). Under the sound limb, significant decreases were observed for
three of the amputees, ranging from 4 to 20% (p < 0.001) for the HA-OFF
condition. All five amputees presented a DOA closer to zero for the HA-OFF
condition, compared to RA-OFF (Figure 4(b)).
Figure 4.
The mean (a) ground reaction force (GRF) and (c) ground reaction
force component parallel to the ground (GRFx) for each
participantunder the prosthetic foot (solid) and the sound foot
(striped) when using a rigid ankle-foot (grey) and a hydraulic
ankle-foot (black). The error bars indicate ± one standard
deviation. Values were measured with ‘standing support’ off.
Significant changes are marked with asterisks. The background
horizontal lines and shaded areas indicate the mean ± one standard
deviation ranges for the GRF/GRFx of the dominant (solid line) and
non-dominant (dashed line) limbs of the able-bodied (AB) control
participants. Also shown are the degree-of-asymmetry values for (b)
GRF and (d) GRFx, for each of the participants when using a rigid
ankle-foot (horizontal stripes) and a hydraulic ankle-foot (vertical
stripes).
The mean (a) ground reaction force (GRF) and (c) ground reaction
force component parallel to the ground (GRFx) for each
participantunder the prosthetic foot (solid) and the sound foot
(striped) when using a rigid ankle-foot (grey) and a hydraulic
ankle-foot (black). The error bars indicate ± one standard
deviation. Values were measured with ‘standing support’ off.
Significant changes are marked with asterisks. The background
horizontal lines and shaded areas indicate the mean ± one standard
deviation ranges for the GRF/GRFx of the dominant (solid line) and
non-dominant (dashed line) limbs of the able-bodied (AB) control
participants. Also shown are the degree-of-asymmetry values for (b)
GRF and (d) GRFx, for each of the participants when using a rigid
ankle-foot (horizontal stripes) and a hydraulic ankle-foot (vertical
stripes).Once again, GRFx provided further interesting results (Figure 4(c)). During the HA-OFF
condition, the prosthetic side increases were between 14% and 99% for four of
the amputees (p < 0.01), while sound side significant decreases were between
14% and 53% (p < 0.001), for four of the five participants. Three of the five
amputees had a DOA closer to zero with the HA-OFF condition, while the other two
exhibited greater prosthetic side loading than sound side loading (Figure 4(d)).With standing support on (RA-ON vs. HA-ON), the change of prosthetic foot didn’t
show a consistent trend in terms of GRF. During HA-ON, prosthetic side GRF only
increased for a single amputee by 7% (p < 0.001) while two showed a decrease
of 3–12% (p < 0.001). Equally mixed results were observed on the sound side –
two amputees showed a significant decrease with HA-ON of 7–17% (p < 0.001)
while two showed significant increases of 6–13% (p < 0.001).
Centre-of-pressure: Balance ability
Figure 5 shows the mean
inter-limb COP RMS for each prosthetic condition, presenting all five amputees
as a cohort. The addition of standing support resulted in a 9% decrease in COP
RMS with a rigid ankle (RA-OFF vs. RA-ON, not significant) and an 11% decrease
with the hydraulic ankle (HA-OFF vs. HA-ON, p = 0.02). The change from a rigid
ankle to a hydraulic one resulted in a 24% decrease in COP RMS with standing
support off (RA-OFF vs. HA-OFF, p < 0.001) and a 25% decrease with standing
support on (RA-ON vs. HA-ON, p < 0.001). The prosthetic condition that
presented the closest COP RMS to that of the able-bodied controls was the HA-ON
condition – in fact, no statistically significant difference was found between
this condition and the able-bodied control participants.
Figure 5.
The cohort, inter-limb mean COP RMS for each prosthetic condition –
‘standing support’ off (striped), ‘standing support’ on (solid),
rigid ankle-foot (grey), hydraulic ankle-foot (black). The same
measure for the able-bodied controls is also shown (white).
Significant changes are marked with asterisks.
The cohort, inter-limb mean COP RMS for each prosthetic condition –
‘standing support’ off (striped), ‘standing support’ on (solid),
rigid ankle-foot (grey), hydraulic ankle-foot (black). The same
measure for the able-bodied controls is also shown (white).
Significant changes are marked with asterisks.
Discussion
If an amputee is not confident about the weight-bearing capability of their
prosthetic limb, they may use compensatory movements to offload the device,
affecting balance, and resulting in asymmetrical loading of their joints, which may
lead to lower back pain. Prosthetic knee manufacturers have explored variable
flexion resistance and stance phase microprocessor-control strategies as methods to
mitigate these risks. With respect to microprocessor prosthetic knees, there is a
wealth of evidence relating to their functionality during walking tasks,[29-35] but noticeably less evidence
relating to non-walking activities of daily living, such as gait termination, gait
initiation and standing. This is in spite of the high occurrence of transitionary
activities and the low bouts of steady-state walking that have been known to occur
in daily life.[36,37] The goal of this study was to investigate aspects of quiet
standing, which may be more frequent and potentially more risky forprosthetic knees
and feet than simply level, steady-state walking. During a period of quiet standing
on a declined surface the kinematic compensations, load distribution and the COP
trajectory were observed – measures which are thought to be related to standing
stability and safety.[28,38]Kinematic compensations were only observed when the conventional, RA prosthetic foot
was worn (Figure 2).
Conventional prosthetic feet are rigidly attached at the ankle and rely on the
deflection of keel and/or heel and toe springs to mimic the dorsiflexion and
plantarflexion movements of biological feet during walking. These springs act
antagonistically, meaning they act to return the foot and shank to a fixed
equilibrium point, usually defined during static alignments on flat ground. These
forces are transmitted through the socket interface into the body. If an amputee is
standing, rather than walking, on sloped ground, kinematic compensations are
commonly used to achieve a ‘foot-flat’ and unload excessive socket forces. These
compensations move the locations of the joint centres, altering the joint moments,
which, when excessive, require TFAs to ‘pull back’ with hip extension with their
residuum inside their socket to obtain normal posture. HAs can be mechanically
modelled as series spring-damper systems, the ankle mechanism enables the joint to
have a variable equilibrium position and thus ‘self-align’ within the range of
hydraulic movement. This permits a more natural posture with reduced kinematic
compensations.The resulting kinetic effects were apparent in the inter-limb load distribution
analysis. The participants in this study tended to increase load on the prosthetic
limb and reduce the load borne on the sound side, when the HA was worn, rather than
the RA (Figure 4). This was
most apparent when standing support mode was off when four of the five participants
presented a DOA closer to 0, indicating improved symmetry with the HA (Figure 4(b)).The addition of standing support made the difference in inter-limb load distribution
between the RA and the HA less pronounced. This suggests that, in spite of kinematic
compensations, the amputees were still able to load bear on the prosthetic side
while wearing the RA. It should be noted that the manufacturers stateI
that standing support can activate, even with the knee flexed. That said, the
addition of standing support had a greater effect on the RA (i.e. the RA-OFF vs.
RA-ON comparison), than it did on the HA (i.e. the HA-OFF vs. HA-ON comparison).
Standing support reduced sound limb loading, in both resultant GRF and the GRFx
component, for all five amputees when using the RA (Figure 3(a) and (c)), compared to only two reductions in
resultant GRF when using the HA.A consistent observation between the HA and RA was the effect on GRFx. Four of the
five amputees had increased GRFx with the HA compared to the RA, when standing
support mode was off (Figure
3(c)). It is postulated that this observation can be explained by the
reduction in compensatory mechanisms when using the HA. With ankle compliance to the
slope, a more natural, upright posture is achieved, positioning the whole body
centre-of-mass in the same location as that of an able-bodied participant. The
result of this is that the contribution of GRFx to bodyweight support is closer to
being biomimetic.Figure 5 shows the mean,
inter-limb value of COP RMS – the outcome measure for balance ability – for the
amputees as a cohort, for each prosthetic condition. This was compared to that of
able-bodied control participants. The effect of the HA was greater than that of
standing support – leading to a 24–25% reduction in COP RMS when the HA was worn,
compared to the 9–11% reduction produced by the presence of standing support.
Self-alignment of the ankle to the slope had a greater influence on balance than the
ability to weight-bear on a flexed knee. That said, overall the results closest to
healthy biomechanics were achieved when both functionalities were used together. In
fact, for this prosthetic condition, the COP RMS was not significantly different
from that of the able-bodied control participants.The case study design was chosen to eliminate inter-participant differences, such as
participant demographic or physiology – any compounding factors that may affect
their ability to balance. This provided useful insights into the effects of
different prosthetic technologies. However, this approach does have limitations and
may have still been influenced by characteristics of the individual participants.
For example, in this study, each of the participants had previous experience of
advanced prosthetic devices. The study did not address how long it would take to
develop these observed benefits for a new user, who had no previous experience of
the advanced technology. Nevertheless, the consistent trends observed across all
participants (i.e. reduction in sound limb GRF with RA-ON compared to RA-OFF)
highlight the promise of the chosen outcome measures to indicate clinically
meaningful changes across a much larger cohort of amputees. Future work will build
on the current dataset and expand to a larger sample to more definitively evaluate
the efficacy of advanced prosthetic technology.Another area for future study will be to examine microprocessor ankle-feet (MPF).
Given that these devices are able to comply with gradients, like hydraulic ankles,
and many now provide a standing support function, like MPKs, it is likely that MPFs
will further enhance inter-limb loading and standing balance. For TFAs, the effect
of the combination of standing support provided by both the knee and the ankle would
be of interest, particularly for bilateral amputees, where sound limb reliance is
not an option. Furthermore, MPFs would also permit a parallel study with a cohort of
trans-tibial amputees (TTA).
Conclusion
Both prosthetic knee and ankle technologies must be considered when assessing the
optimum clinical outcomes for a TFA. While the prosthetic knee has traditionally
been the focus in TFA literature,[29-35] the ankle-foot plays an
important role as it is the point of contact with the surrounding environment. Both
microprocessor-controlled standing support mode at the knee and hydraulic
self-alignment of the ankle joint have been shown to contribute to more symmetrical
weight distribution and improved balance ability. Indeed, the combination of the two
technologies exhibited the best results that were the closest to normal standing
biomechanics.
Manufacturers’ documentation
I. https://www.blatchford.co.uk/endolite/espritII. https://www.blatchford.co.uk/endolite/echelonvtIII. https://www.blatchford.co.uk/endolite/orion3
Authors: D M Ehde; D G Smith; J M Czerniecki; K M Campbell; D M Malchow; L R Robinson Journal: Arch Phys Med Rehabil Date: 2001-06 Impact factor: 3.966
Authors: D M Ehde; J M Czerniecki; D G Smith; K M Campbell; W T Edwards; M P Jensen; L R Robinson Journal: Arch Phys Med Rehabil Date: 2000-08 Impact factor: 3.966
Authors: David F Rusaw; Rasmus Alinder; Sigurd Edholm; Karin L L Hallstedt; Jessika Runesson; Cleveland T Barnett Journal: Sci Rep Date: 2021-04-15 Impact factor: 4.379
Authors: Michael McGrath; Laura A Gray; Beata Rek; Kate C Davies; Zoe Savage; Jane McLean; Alison Stenson; Saeed Zahedi Journal: PLoS One Date: 2022-09-02 Impact factor: 3.752