His bundle has a shorter chronaxie than does the adjacent ventricular myocardium: Implications for pacemaker programming.

Marek Jastrzębski1, Paweł Moskal2, Agnieszka Bednarek2, Grzegorz Kiełbasa2, Pugazhendhi Vijayaraman3, Danuta Czarnecka2.   

Abstract

BACKGROUND: Strength-duration curves for permanent His bundle (HB) pacing are potentially important for pacemaker programming.
OBJECTIVES: We aimed to calculate strength-duration curves and chronaxies of the HB and of the adjacent right ventricular (RV) working myocardium and to analyze zones of selective HB capture and battery current drain when pacing at different pulse durations (PDs).
METHODS: Consecutive patients with permanent HB pacing were studied. The RV and HB capture thresholds were assessed at several PDs. Battery current drain and zones of selective HB capture at PDs of 0.1, 0.2, 0.4, and 1.0 ms were determined.
RESULTS: In the whole group (N = 127), the HB chronaxie was shorter than the RV chronaxie. This difference was driven by patients with selective HB pacing (HB chronaxie 0.47 ms vs RV chronaxie 0.79 ms). The strength-duration curve for the HB had a lower rheobase and its steep portion started at shorter PDs, thus creating wider distance-zone of programmable selective HB pacing-between the HB and RV strength-duration curves at shorter PDs. The battery current drain was lower with pacing at PDs of 0.1-0.4 ms vs 1.0 ms. Chronaxie-adjusted PDs offered the lowest current drain.
CONCLUSION: For the first time, the strength-duration curves for permanent selective and nonselective HB pacing were determined. Selective HB capture and battery longevity can be promoted by shorter PDs (0.2 ms). Longer PDs (1.0 ms) offer greater safety margin for RV capture and may be preferable if simultaneous RV capture during HB pacing is desired.
Copyright © 2019 Heart Rhythm Society. Published by Elsevier Inc. All rights reserved.

Entities:  

Keywords:  Chronaxie; Current drain; His bundle pacing; Selective capture; Strength-duration

Year:  2019        PMID: 31181375     DOI: 10.1016/j.hrthm.2019.06.001

Source DB:  PubMed          Journal:  Heart Rhythm        ISSN: 1547-5271            Impact factor:   6.343


  2 in total

1.  In Vitro/Ex Vivo Investigation of Modified Utah Electrode Array to Selectively Sense and Pace the Sub-Surface Cardiac His Bundle.

Authors:  Ankur R Shah; Muhammad S Khan; Annie M Hirahara; Matthias Lange; Ravi Ranjan; Derek J Dosdall
Journal:  ACS Biomater Sci Eng       Date:  2020-05-07

2.  Crossing of strength-duration curves with His bundle pacing and impact of pacing mode on thresholds.

Authors:  Elise Bakelants; Alwin Zweerink; Haran Burri
Journal:  HeartRhythm Case Rep       Date:  2020-11-28
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.