Single-port thoracoscopic wedge resection using the Endo GIA Radial Reload: outcome of 15 cases.

Atsushi Sano1.   

Abstract

BACKGROUND: We previously described the technique for single-port thoracoscopic lung wedge resection using the Endo GIATM Radial Reload surgical stapler. The GIA Radial Reload has a curved cut line that is perpendicular to the direction of instrument insertion, facilitating the approach to the lung tissue behind the lesion. Herein, we report the outcomes of 15 cases using this procedure.
METHODS: Between August 2016 and February 2018, 15 patients underwent single-port thoracoscopic wedge resection using the Endo GIA Radial Reload. Single-port thoracoscopic surgery was performed through 3-5-cm incisions. For the first stapler, we used a GIA Radial Reload cartridge. For the second and subsequent staplers, we used the GIA Radial Reload cartridge or a straight cartridge based on the direction of the cut.
RESULTS: In one patient, we extended the incision up to 7 cm and performed video-assisted thoracotomy because the lesion was too small to find. In the other 14 patients, the procedure was completed through a 3-5-cm port. In one patient, we added a suture for air leakage detected intraoperatively. Postoperative air leakage occurred in two patients. These three patients had emphysema based on computed tomography. In the remaining 12 patients, no air leakage was found and the chest tube was removed on postoperative day 1.
CONCLUSIONS: This procedure is suitable for resecting small nodules, because palpation with two fingers is easy. The GIA Radial Reload cartridge may not be a good choice for emphysematous lung, because air leakage sometimes occurs.

Entities:  

Keywords:  Device; operative procedure; thoracoscopy

Year:  2019        PMID: 31179094      PMCID: PMC6531696          DOI: 10.21037/jtd.2019.03.26

Source DB:  PubMed          Journal:  J Thorac Dis        ISSN: 2072-1439            Impact factor:   2.895


  7 in total

1.  Uniportal VATS wedge pulmonary resections.

Authors:  Gaetano Rocco; Antonio Martin-Ucar; Eliseo Passera
Journal:  Ann Thorac Surg       Date:  2004-02       Impact factor: 4.330

2.  Thoracoscopic reoperation for recurrent pneumothorax after single-incision thoracoscopic surgery.

Authors:  Atsushi Sano; Mitsuaki Kawashima
Journal:  Asian J Endosc Surg       Date:  2015-11

3.  Single-port thoracoscopic lung wedge resection using the Endo GIA Radial Reload.

Authors:  Atsushi Sano; Takuma Yotsumoto
Journal:  Surg Today       Date:  2017-07-25       Impact factor: 2.549

4.  Single-port thoracoscopic surgery for a hudge hydatid cyst.

Authors:  Ilhan Ocakcioglu
Journal:  Gen Thorac Cardiovasc Surg       Date:  2016-05-05

5.  Uniportal vs standard three-port VATS technique for spontaneous pneumothorax: comparison of post-operative pain and residual paraesthesia.

Authors:  Rajwinder Singh Jutley; Mohammed Wesam Khalil; Gaetano Rocco
Journal:  Eur J Cardiothorac Surg       Date:  2005-07       Impact factor: 4.191

6.  Single-port video-assisted thoracic surgery for early lung cancer: initial experience in Japan.

Authors:  Kyoji Hirai; Shingo Takeuchi; Jitsuo Usuda
Journal:  J Thorac Dis       Date:  2016-03       Impact factor: 2.895

7.  Is finger palpation at operation indispensable for pulmonary metastasectomy in colorectal cancer?

Authors:  Jun Nakajima; Tomohiro Murakawa; Takeshi Fukami; Atsushi Sano; Miki Sugiura; Shinichi Takamoto
Journal:  Ann Thorac Surg       Date:  2007-11       Impact factor: 4.330