| Literature DB >> 31164192 |
Kaisu Rantakokko-Jalava1,2, Kati Hokynar3, Niina Hieta1,4, Anniina Keskitalo1,2, Pia Jokela5, Anna Muotiala6, T Sakari Jokiranta7, Rutta Kuusela8, Hannu Sarkkinen9, Janne Aittoniemi10, Tytti Vuorinen1,2, Antti J Hakanen1,2, Mirja Puolakkainen5.
Abstract
Since February 2019, over 160 Chlamydia trachomatis (CT) cases testing negative or equivocal by Aptima Combo 2 (AC2) but positive by Aptima CT test run with Panther instruments occurred in Finland. The AC2 test targets chlamydial 23S rRNA while the CT test targets 16S rRNA. Sequencing of 10 strains revealed a nucleotide substitution in 23S rRNA. The significance of this for the failure of the AC2 test to detect the variant is not yet known.Entities:
Keywords: Chlamydia trachomatis; molecular diagnostics; variant
Year: 2019 PMID: 31164192 PMCID: PMC6549458 DOI: 10.2807/1560-7917.ES.2019.24.22.1900298
Source DB: PubMed Journal: Euro Surveill ISSN: 1025-496X
Samples negative or equivocal for Chlamydia trachomatis by Aptima Combo 2 retested by Aptima CT in Turku, Finland, 11–29 March 2019 (n = 943)
| Qualitative result in the original AC2 test according to the instrument display | RLU in the original AC2 test | Number of samples tested by ACT test | Positive in the ACT | The AC2 test RLU values in the ACT positives | |
|---|---|---|---|---|---|
| Negative or equivocal for CT and negative for GC | ≤ 10 | 916 | 1 | 0.1 | 3 |
| 11–15 | 9 | 1 | 11 | 15 | |
| 16–19 | 0 | 0 | 0 | n.a. | |
| 20–84 | 15 | 13 | 87 | 20–49a | |
| Negative for CT and positive for GC | > 1,300 | 3 | 0 | 0 | n.a. |
AC2: Aptima Combo 2; ACT: Aptima CT; CT: Chlamydia trachomatis; GC: Neisseria gonorrhoeae; n.a.: not applicable; RLU: relative light units.
a One sample with RLU 49 was flagged as C. trachomatis equivocal; the others were reported as negative by the Panther instrument.
Samples re-analysed with Aptima CT test in HUSLAB, Finland, 6 March–30 April 2019 (n = 757)
| Qualitative result in the original AC2 test according to the instrument display | RLU in the original AC2 test | Number of samples tested by ACT test | Positive in the ACT | The AC2 test RLU values in the ACT positives | |
|---|---|---|---|---|---|
| Negative or equivocal for CT and negative for GC | ≤ 10 | 330 | 2 | 0.6 | 5–7 |
| 11–15 | 266 | 7 | 2.6 | 14–15 | |
| 16–19 | 71 | 13 | 18 | 16–19 | |
| 20–84 | 73 | 68a | 93 | 20–46 | |
| 85–250 | 3 | 3b | 100 | 89–97 | |
| Negative for CT and positive for GC | 48–1,492 | 14 | 1 | 7.1 | 1,492 |
AC2: Aptima Combo 2; ACT: Aptima CT; CT: Chlamydia trachomatis; GC: Neisseria gonorrhoeae; RLU: relative light units.
a Of 50 samples ≥ 25 RLU, one sample was flagged as equivocal in the original AC2 test.
b Of three samples with RLU between 85 and 99, three samples were flagged as equivocal in the original AC2 test.
Samples re-analysed with Aptima CT test in United Medix Laboratories Ltd, Finland, 1 March–7 May 2019 (n = 193)
| RLU in the original AC2 test | Number of samples tested by ACT | Positive in the ACT | The AC2 test RLU values in the ACT positives | |
|---|---|---|---|---|
| ≤ 10 | 51 | 0 | 0 | n.a. |
| 11–15 | 101 | 2 | 2.0 | 13, 14 |
| 16–20 | 10 | 3 | 30 | 17, 18, 20 |
| 21–84 | 31a | 31 | 100 | 21–33 |
| 85–250 | 0 | 0 | 0 | n.a. |
AC2: Aptima Combo 2; ACT: Aptima Chlamydia trachomatis; n.a.: not applicable; RLU: relative light units.
a All were flagged as negative by the Panther instrument.
FigureNumbers of detected AC2-negative or equivocal (RLU 20–84)/ACT-positive cases in hospital districts of Finland, mid-February–April 2019 (n = 180)