In Vitro Evaluation of Mechanical Properties of Segmented Esophageal Self-Expandable Metal Stents: Innovative Test Methods Are Needed.

Martin Bürger1, Till Herbold2, Sven Lange3, Felix Berlth4, Patrick Sven Plum4, Christoph Schramm1, Robert Kleinert4, Tobias Goeser1, Christiane Josephine Bruns4, Seung-Hun Chon4.   

Abstract

Background: Self-expanding metal stents (SEMSs) in different geometric shapes are well established treatment options in diseases of the esophagus. Mechanical properties and stent design may have an impact on patient comfort, migration rate, and removability. In this in vitro study, we evaluated mechanical properties of three segmented SEMSs (segSEMSs) for the esophagus with regard to distinct stent sections. Materials and
Methods: Radial forces were measured using a testing method distinguishing between circumferential radial and local radial force. The center parts of the segSEMSs were measured for circumferential radial forces without being affected by the flared ends. Axial forces were measured at 20° bending.
Results: Circumferential radial force measurements over the full stent length showed substantial differences against measurements of the center parts of the stents as the flared ends falsify test results by up to 53%. Although circumferential radial forces of the center parts were about the same (<10% variances) for all segSEMSs, local radial forces showed considerable differences of up to 26%. One segSEMS showed high axial forces, whereas the other two only needed half of the force (up to 53%) to be bent to 20°.
Conclusion: Flared ends of segSEMSs have a substantial impact on radial force measurements and therefore alter test results, confirmed by our separated center part test of segSEMSs. Our innovative setup whereby we compressed the stent in an asymmetric manner (local radial force) and evaluated sections of stents separately, indeed revealed differences to circumferential measurements, leading to a more in-depth knowledge of stent characteristics.

Entities:  

Keywords:  esophageal cancer; esophageal stent; in vitro mechanical properties; segmented self-expanding metal stent

Mesh:

Year:  2019        PMID: 31161950     DOI: 10.1089/lap.2019.0173

Source DB:  PubMed          Journal:  J Laparoendosc Adv Surg Tech A        ISSN: 1092-6429            Impact factor:   1.878


  1 in total

1.  Author's Reply: Outcome of Self-Expanding Metal Stents in the Treatment of Anastomotic Leaks After Ivor Lewis Esophagectomy.

Authors:  Patrick Sven Plum; Felix Berlth; Till Herbold; Seung-Hun Chon
Journal:  World J Surg       Date:  2019-09       Impact factor: 3.352

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.