Microwave ablation zones are larger than they macroscopically appear - Reevaluation based on NADH vitality staining ex vivo.

Beatrice Geyer1, Franz G M Poch1, Ole Gemeinhardt2, Christina A Neizert1, Stefan M Niehues3, Janis L Vahldiek3, Robert Klopfleisch4, Kai S Lehmann1.   

Abstract

BACKGROUND: Animal liver is established as an ex vivo model for studies on hepatic microwave ablation (MWA). Macroscopically visible color changes in the ablation zone are used to assess cell destruction and confirm successful ablation ex vivo.
OBJECTIVE: Macroscopy and histology of MWA zones regarding cell viability in ex vivo porcine livers were compared in this study.
METHODS: Six MWA were performed in porcine livers post mortem. A 14-G antenna and microwave generator (928 MHz; 9.0 kJ) were used. MWA were cut at the maximum cross section in vertical alignment to the antenna. NADH-diaphorase staining determined cell vitality. Macroscopic and microscopic ablation zones were statistically analyzed.
RESULTS: Histology showed two distinct ablation zones: central white zone (WZH) with no cell viability and peripheral red zone (RZH) with partial cell viability. However, the macroscopically visible WZM was significantly smaller than the microscopic WZH with an area difference of 43.1% (p < 0.05) and a radius difference of 21.2% (1.6 mm; p < 0.05). Macroscopy and histology showed a very high correlation for the complete lesion area (WZH/M+RZH/M; r = 0.9; p = 0.001).
CONCLUSIONS: The avital central zone is significantly larger as the macroscopically visible WZ which is commonly used to assess successful ablation in MWA ex vivo studies. Irreversible cell destruction can be underestimated in macroscopic evaluation.

Entities:  

Keywords:  Liver; NADH-diaphorasezzm321990; animal model; ex vivo; microwave ablation (MWA); viability staining

Mesh:

Substances:

Year:  2019        PMID: 31156148     DOI: 10.3233/CH-190583

Source DB:  PubMed          Journal:  Clin Hemorheol Microcirc        ISSN: 1386-0291            Impact factor:   2.375


  2 in total

1.  Exploring Patterns of Dynamic Size Changes of Lesions after Hepatic Microwave Ablation in an In Vivo Porcine Model.

Authors:  Keno K Bressem; Janis L Vahldiek; Christoph Erxleben; Franz Poch; Seyd Shnaiyen; Beatrice Geyer; Kai S Lehmann; Bernd Hamm; Stefan M Niehues
Journal:  Sci Rep       Date:  2020-01-21       Impact factor: 4.379

2.  Influence of interapplicator distance on multibipolar radiofrequency ablation during physiological and interrupted liver perfusion in an in vivo porcine model.

Authors:  F G M Poch; C A Neizert; B Geyer; O Gemeinhardt; L Bruder; S M Niehues; J L Vahldiek; K K Bressem; M E Kreis; K S Lehmann
Journal:  Sci Rep       Date:  2020-10-01       Impact factor: 4.379

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.