A Method to Detect Presence of Chest Compressions During Resuscitation Using Transthoracic Impedance.

Jason Coult, Jennifer Blackwood, Thomas D Rea, Peter J Kudenchuk, Heemun Kwok.   

Abstract

OBJECTIVE: Interruptions in chest compressions during treatment of out-of-hospital cardiac arrest are associated with lower likelihood of successful resuscitation. Real-time automated detection of chest compressions may improve CPR administration during resuscitation, and could facilitate application of next-generation ECG algorithms that employ different parameters depending on compression state. In contrast to accelerometer sensors, transthoracic impedance (TTI) is commonly acquired by defibrillators. We sought to develop and evaluate the performance of a TTI-based algorithm to automatically detect chest compressions.
METHODS: Five-second TTI segments were collected from patients with out-of-hospital cardiac arrest treated by one of four defibrillator models. Segments with and without chest compressions were collected prior to each of the first four defibrillation shocks (when available) from each case. Patients were divided randomly into 40% training and 60% validation groups. From the training segments, we identified spectral and time-domain features of the TTI associated with compressions. We used logistic regression to predict compression state from these features. Performance was measured by sensitivity and specificity in the validation set. The relationship between performance and TTI segment length was also evaluated.
RESULTS: The algorithm was trained using 1859 segments from 460 training patients. Validation sensitivity and specificity were >98% using 2727 segments from 691 validation patients. Validation performance was significantly reduced using segments shorter than 3.2 s.
CONCLUSIONS: A novel method can reliably detect the presence of chest compressions using TTI. These results suggest potential to provide real-time feedback in order to improve CPR performance or facilitate next-generation ECG rhythm algorithms during resuscitation.

Entities:  

Mesh:

Year:  2019        PMID: 31144648      PMCID: PMC7235095          DOI: 10.1109/JBHI.2019.2918790

Source DB:  PubMed          Journal:  IEEE J Biomed Health Inform        ISSN: 2168-2194            Impact factor:   5.772


  30 in total

1.  Compression depth estimation for CPR quality assessment using DSP on accelerometer signals.

Authors:  Sven O Aase; Helge Myklebust
Journal:  IEEE Trans Biomed Eng       Date:  2002-03       Impact factor: 4.538

Review 2.  Part 5: Adult Basic Life Support and Cardiopulmonary Resuscitation Quality: 2015 American Heart Association Guidelines Update for Cardiopulmonary Resuscitation and Emergency Cardiovascular Care.

Authors:  Monica E Kleinman; Erin E Brennan; Zachary D Goldberger; Robert A Swor; Mark Terry; Bentley J Bobrow; Raúl J Gazmuri; Andrew H Travers; Thomas Rea
Journal:  Circulation       Date:  2015-11-03       Impact factor: 29.690

Review 3.  Part 7: Adult Advanced Cardiovascular Life Support: 2015 American Heart Association Guidelines Update for Cardiopulmonary Resuscitation and Emergency Cardiovascular Care.

Authors:  Mark S Link; Lauren C Berkow; Peter J Kudenchuk; Henry R Halperin; Erik P Hess; Vivek K Moitra; Robert W Neumar; Brian J O'Neil; James H Paxton; Scott M Silvers; Roger D White; Demetris Yannopoulos; Michael W Donnino
Journal:  Circulation       Date:  2015-11-03       Impact factor: 29.690

4.  Interruptions of chest compressions during emergency medical systems resuscitation.

Authors:  Terence D Valenzuela; Karl B Kern; Lani L Clark; Robert A Berg; Marc D Berg; David D Berg; Ronald W Hilwig; Charles W Otto; Daniel Newburn; Gordon A Ewy
Journal:  Circulation       Date:  2005-08-22       Impact factor: 29.690

Review 5.  A review of the performance of artifact filtering algorithms for cardiopulmonary resuscitation.

Authors:  Yushun Gong; Bihua Chen; Yongqin Li
Journal:  J Healthc Eng       Date:  2013       Impact factor: 2.682

6.  Automatic detection of chest compressions for the assessment of CPR-quality parameters.

Authors:  U Ayala; T Eftestøl; E Alonso; U Irusta; E Aramendi; S Wali; J Kramer-Johansen
Journal:  Resuscitation       Date:  2014-04-15       Impact factor: 5.262

7.  Suppression of the cardiopulmonary resuscitation artefacts using the instantaneous chest compression rate extracted from the thoracic impedance.

Authors:  E Aramendi; U Ayala; U Irusta; E Alonso; T Eftestøl; J Kramer-Johansen
Journal:  Resuscitation       Date:  2011-12-21       Impact factor: 5.262

8.  Association Between Chest Compression Interruptions and Clinical Outcomes of Ventricular Fibrillation Out-of-Hospital Cardiac Arrest.

Authors:  Tom F Brouwer; Robert G Walker; Fred W Chapman; Rudolph W Koster
Journal:  Circulation       Date:  2015-08-07       Impact factor: 29.690

9.  Ventricular fibrillation waveform measures combined with prior shock outcome predict defibrillation success during cardiopulmonary resuscitation.

Authors:  Jason Coult; Heemun Kwok; Lawrence Sherman; Jennifer Blackwood; Peter J Kudenchuk; Thomas D Rea
Journal:  J Electrocardiol       Date:  2017-08-01       Impact factor: 1.438

10.  A new paradigm for human resuscitation research using intelligent devices.

Authors:  Charles F Babbs; Andre E Kemeny; Weilun Quan; Gary Freeman
Journal:  Resuscitation       Date:  2008-03-07       Impact factor: 5.262

View more
  2 in total

1.  Automatic Detection of Ventilations During Mechanical Cardiopulmonary Resuscitation.

Authors:  Xabier Jaureguibeitia; Unai Irusta; Elisabete Aramendi; Pamela C Owens; Henry E Wang; Ahamed H Idris
Journal:  IEEE J Biomed Health Inform       Date:  2020-01-17       Impact factor: 5.772

2.  Monitoring chest compression rate in automated external defibrillators using the autocorrelation of the transthoracic impedance.

Authors:  Sofía Ruiz de Gauna; Jesus María Ruiz; Jose Julio Gutiérrez; Digna María González-Otero; Daniel Alonso; Carlos Corcuera; Juan Francisco Urtusagasti
Journal:  PLoS One       Date:  2020-09-30       Impact factor: 3.240

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.