Literature DB >> 31125671

Why low-voltage shock impedance measurements fail to reliably detect insulation breaches in transvenous defibrillation leads.

Charles D Swerdlow1, John E Porterfield2, Anil G Kottam2, Mark W Kroll3.   

Abstract

BACKGROUND: Implantable cardioverter-defibrillators (ICDs) use low-voltage measures of shock impedance (LVSZ) to monitor integrity of leads.
OBJECTIVE: To determine the separation distance between conductors required for LVSZ to detect insulation breaches that produce short circuits during shocks, causing failed defibrillation.
METHODS: We simulated in-pocket insulation breaches between the ICD generator (CAN) and cables to the distal coil of 10 leads from 2 manufacturers. The ICD and lead were placed in an electrolyte bath. Polystyrene sheets were used to control the breach-CAN separation. We determined both the maximum lead-CAN separation for shorts during 800 V shocks and the shock strength at which shorts occurred for a fixed separation. We also calculated breach impedance and measured it using a low-voltage instrument.
RESULTS: The maximum breach-CAN separation for shorting was 350-500 μm for all leads. The minimum shock strength to short varied from 650 to 771 V (24-32 J). LVSZ never triggered a warning, even with no separation between the cable's inner insulation and the CAN. Using low-voltage pulses, breach impedance was measured at approximately 500-1000 Ω.
CONCLUSION: LVSZ is insensitive to insulation breaches that cause life-threatening, shorted shocks. The explanation likely relates to impedance differences between ionic conduction during LVSZ measurements and free-electron conduction in plasma discharges.
Copyright © 2019 Heart Rhythm Society. Published by Elsevier Inc. All rights reserved.

Entities:  

Keywords:  Defibrillation; Impedance; Implantable cardioverter-defibrillator; Insulation; Leads

Mesh:

Year:  2019        PMID: 31125671     DOI: 10.1016/j.hrthm.2019.05.021

Source DB:  PubMed          Journal:  Heart Rhythm        ISSN: 1547-5271            Impact factor:   6.343


  1 in total

1.  High shocking and pacing impedances due to defibrillation lead calcification.

Authors:  Robert G Hauser; Jay Sengupta; Susan Casey; Chuen Tang; Larissa I Stanberry; Raed Abdelhadi
Journal:  J Interv Card Electrophysiol       Date:  2019-12-18       Impact factor: 1.900

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.