Evaluating Propofol Concentration in Blood From Exhaled Gas Using a Breathing-Related Partition Coefficient.

Hao Dong1,2, Fengjiang Zhang3, Jing Chen1, Qiwen Yu1,2, Yinbo Zhong3, Jun Liu1,2, Min Yan3, Xing Chen1,2.   

Abstract

BACKGROUND: The anesthetic side effects of propofol still occur in clinical practice because no reliable monitoring techniques are available. In this regard, continuous monitoring of propofol in breath is a promising method, yet it remains infeasible because there is large variation in the blood/exhaled gas partial pressure ratio (RBE) in humans. Further evaluations of the influences of breathing-related factors on RBE would mitigate this variation.
METHODS: Correlations were analyzed between breathing-related factors (tidal volume [TV], breath frequency [BF], and minute ventilation [VM]) and RBE in 46 patients. Furthermore, a subset of 10 patients underwent pulmonary function tests (PFTs), and the parameters of the PFTs were then compared with the RBE. We employed a 1-phase exponential decay model to characterize the influence of VM on RBE. We also proposed a modified RBE (RBEM) that was not affected by the different breathing patterns of the patients. The blood concentration of propofol was predicted from breath monitoring using RBEM and RBE.
RESULTS: We found a significant negative correlation (R = -0.572; P < .001) between VM and RBE (N = 46). No significant correlation was shown between PFTs and RBE in the subset (N = 10). RBEM demonstrated a standard Gaussian distribution (mean, 1.000; standard deviation [SD], 0.308). Moreover, the predicted propofol concentrations based on breath monitoring matched well with the measured blood concentrations. The 90% prediction band was limited to within ±1 μg·mL.
CONCLUSIONS: The prediction of propofol concentration in blood was more accurate using RBEM than when using RBE and could provide reference information for anesthesiologists. Moreover, the present study provided a general approach for assessing the influence of relevant physiological factors and will inform noninvasive and accurate breath assessment of volatile drugs or metabolites in blood.

Entities:  

Year:  2020        PMID: 31124837     DOI: 10.1213/ANE.0000000000004225

Source DB:  PubMed          Journal:  Anesth Analg        ISSN: 0003-2999            Impact factor:   5.108


  1 in total

1.  Online Accurate Detection of Breath Acetone Using Metal Oxide Semiconductor Gas Sensor and Diffusive Gas Separation.

Authors:  Hao Dong; Libin Qian; Yaoxuan Cui; Xubin Zheng; Chen Cheng; Qingpeng Cao; Feng Xu; Jin Wang; Xing Chen; Di Wang
Journal:  Front Bioeng Biotechnol       Date:  2022-03-08
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.