Literature DB >> 31102046

Duration of ultrasonic activation causing secondary fractures during the removal of the separated instruments with different tapers.

Hakan Arslan1, Ezgi Doğanay Yıldız2, Gizem Taş3, Nuray Akbıyık3, Hüseyin Sinan Topçuoğlu4.   

Abstract

OBJECTIVES: The aim of the present study was to determine the effect of taper (.08, .06, and .04) of separated K3XF instruments on duration taken for the secondary fracture formation during ultrasonic activation.
MATERIALS AND METHODS: Ten 25/.08 K3XF (SybronEndo, Orange, CA, USA), ten 25/.06 K3XF, and ten 25/.04 K3XF instruments were used for the study. The apical 5 mm of the instruments was cut to simulate the fragments in root canals. Fragments of the instruments were sandwiched between two straight dentin blocks. An ultrasonic tip was used to cause a secondary fracture of the fragment. The time needed for the secondary fracture was recorded for each instrument. The data were statistically analyzed using the Kruskal-Wallis H test (alpha = 0.05).
RESULTS: Secondary fractures occurred in all instruments. In the .08 taper group, secondary fractures took longer than in the case of the .06 and the .04 taper groups (P < 0.05). There were no significant differences between the .06 and the .04 taper groups in terms of the time required for the occurrence of a secondary fracture (P > 0.05).
CONCLUSIONS: In the .08 taper group, secondary fracture took longer time than in the case of the .06 and the .04 taper groups due to its larger cross-sectional area involved. CLINICAL RELEVANCE: Typically, when removing separated instruments, a much lower power setting is chosen. The purpose of this in vitro study was to determine which tapered files were more resilient to secondary fracture, thus allowing a higher power setting to be chosen. Thus, the results of the present study cannot be used in clinical practice. If the clinician knows the taper of the broken file, the clinician should be very careful with regard to secondary fractures when using ultrasonics to remove the separated smaller tapered instruments.

Entities:  

Keywords:  K3XF; Nickel titanium; Secondary fracture; Separated file; Taper; Ultrasonics

Mesh:

Substances:

Year:  2019        PMID: 31102046     DOI: 10.1007/s00784-019-02936-7

Source DB:  PubMed          Journal:  Clin Oral Investig        ISSN: 1432-6981            Impact factor:   3.573


  14 in total

1.  Evaluation of an ultrasonic technique to remove fractured rotary nickel-titanium endodontic instruments from root canals: clinical cases.

Authors:  Jeff R Ward; Peter Parashos; Harold H Messer
Journal:  J Endod       Date:  2003-11       Impact factor: 4.171

2.  Evaluation of an ultrasonic technique to remove fractured rotary nickel-titanium endodontic instruments from root canals: an experimental study.

Authors:  Jeff R Ward; Peter Parashos; Harold H Messer
Journal:  J Endod       Date:  2003-11       Impact factor: 4.171

3.  Perforation risks associated with the use of Masserann endodontic kit drills in mandibular molars.

Authors:  Oguz Yoldas; Haluk Oztunc; Cemal Tinaz; Nazan Alparslan
Journal:  Oral Surg Oral Med Oral Pathol Oral Radiol Endod       Date:  2004-04

4.  K3 Endo, ProTaper, and ProFile systems: breakage and distortion in severely curved roots of molars.

Authors:  Matthew T Ankrum; Gary R Hartwell; John E Truitt
Journal:  J Endod       Date:  2004-04       Impact factor: 4.171

5.  Factors associated with the removal of fractured NiTi instruments from root canal systems.

Authors:  Ya Shen; Bin Peng; Gary Shun-Pan Cheung
Journal:  Oral Surg Oral Med Oral Pathol Oral Radiol Endod       Date:  2004-11

6.  A retrospective clinical study of incidence of root canal instrument separation in an endodontics graduate program: a PennEndo database study.

Authors:  Mian K Iqbal; Meetu R Kohli; Jessica S Kim
Journal:  J Endod       Date:  2006-09-25       Impact factor: 4.171

7.  Evaluation of the efficiency of a new file removal system in comparison with two conventional systems.

Authors:  Yoshitsugu Terauchi; Le O'Leary; Izumi Kikuchi; Mami Asanagi; Takatomo Yoshioka; Chihiro Kobayashi; Hideaki Suda
Journal:  J Endod       Date:  2007-03-12       Impact factor: 4.171

8.  Comparison of the time required to create secondary fracture of separated file fragments by using ultrasonic vibration under various canal conditions.

Authors:  Yoshitsugu Terauchi; Le O'Leary; Takatomo Yoshioka; Hideaki Suda
Journal:  J Endod       Date:  2013-10       Impact factor: 4.171

9.  Instrument separation analysis of multi-used ProTaper Universal rotary system during root canal therapy.

Authors:  Jintao Wu; Gang Lei; Ming Yan; Yan Yu; Jinhua Yu; Guangdong Zhang
Journal:  J Endod       Date:  2011-04-02       Impact factor: 4.171

10.  Success of ultrasonic technique in removing fractured rotary nickel-titanium endodontic instruments from root canals and its effect on the required force for root fracture.

Authors:  Hasan Shahabinejad; Amirreza Ghassemi; Lida Pishbin; Arash Shahravan
Journal:  J Endod       Date:  2013-04-16       Impact factor: 4.171

View more