Christopher N Schmickl1,2, Aravind Ajakumar Menon3,4, Rajanigandha Dhokarh5, Bhavna Seth3,6, Frank Schembri3.
Abstract
B-lines on lung ultrasound (US) are the hallmark of pulmonary edema. It is unknown if ultrasound machine settings or probe type matter. We created an in-vitro gelatin model. Using lung presets as baseline, five blinded investigators assessed the impact of 32 distinct settings on B-line visibility based on a Likert-Scale (LS) from 0 to10 (< 5 worse, > 5 better) separately for two probes. The experiment was then repeated in-vivo in a patient with known pulmonary edema. Based on a multivariable regression LS-ratings were similar when comparing the in-vitro versus in-vivo experiment (P = 0.16; partial R2 = 0.2%) and when using the curvilinear versus linear probe (P = 0.69; partial R2 = 0.02%) but significantly different across machine settings (P < 0.0001; partial R2 = 34.4%). Limited by its pilot character, our study suggests that (1) certain US-machine settings heavily impact B-line visibility, with no clear difference between probes; (2) in-vitro models are a valid and practical alternative to more challenging patient-based research; (3) there is significant potential to improve B-line visibility and thus diagnostic yield in the clinical setting by using lung presets, centering the focal zone at the pleural line and increasing the distal time gain compensation, most of which are (in our experience) rarely done.Entities:
Keywords: B-line; Lung ultrasound; Machine setting; in-vitro model; in-vivo validation
Mesh:
Substances:
Year: 2019 PMID: 31089845 PMCID: PMC7429193 DOI: 10.1007/s10877-019-00321-z
Source DB: PubMed Journal: J Clin Monit Comput ISSN: 1387-1307 Impact factor: 2.502