Chemosaturation Percutaneous Hepatic Perfusion (CS-PHP) with Melphalan: Evaluation of 2D-Perfusion Angiography (2D-PA) for Leakage Detection of the Venous Double-Balloon Catheter.

Cornelia L A Dewald1, Timo C Meine1, Hinrich M B Winther1, Roman Kloeckner2, Sabine K Maschke1, Martha M Kirstein3, Arndt Vogel3, Frank K Wacker1, Bernhard C Meyer1, Julius Renne1, Jan B Hinrichs4.   

Abstract

PURPOSE: To evaluate the feasibility of 2D-perfusion angiography (2D-PA) for detecting leakage of the double-balloon catheter used for chemosaturation percutaneous hepatic perfusion (CS-PHP).
MATERIALS AND METHODS: Overall, 112 CS-PHP (09/2015-09/2018) in 52 patients were retrospectively screened for leakage alongside the double-balloon catheter on standard venograms. Finally, 18 procedures with visually detected leakage were included. Fifteen consecutive procedures without leakage served as control. To evaluate 2D-PA for leakage detection, the acquired digital subtraction venograms were post-processed. For each balloon, two different target ROIs were evaluated to assess a possible impact of localization and shape of the ROIs. Time to peak (TTP), peak density (PD), area under the curve (AUC), and ratios of target ROI/reference ROIs (PDtROI/PDREF; AUCtROI/AUCREF; and TTPtROI/TTPREF) were calculated.
RESULTS: Leakages were located as follows: 15/18 cranial and 3/18 caudal. At the cranial balloon both ROIs showed a significant decrease in PDtROI/PDREF and AUCtROI/AUCREF (ROI1: p < 0.0001; p < 0.0001; ROI2: p < 0.0001; p < 0.0001) and a significant increase in TTPtROI/TTPREF (ROI1: p = 0.0009; ROI2: p = 0.0003) after double-balloon correction. Following balloon adjustment, the 2D-PA ratios (PD and AUC) of the tested ROIs differed significantly (p < 0.05). The inter-individual comparison of the 2D-PA parameters of the group with leakage before balloon correction and the non-leakage group showed significantly different 2D-PA values for the cranial balloon in both ROIs (p < 0.05). No significant differences were found for the caudal balloon.
CONCLUSION: 2D-PA provides a feasible tool for detecting leakages alongside the cranial portion of the double-balloon catheter used in CS-PHP. The shape and position of the ROIs used to assess perfusion and flow have an impact on the measurements.

Entities:  

Keywords:  2D-perfusion angiography; Chemosaturation; Double-balloon catheter; Leakage; Percutaneous hepatic perfusion

Mesh:

Substances:

Year:  2019        PMID: 31089781     DOI: 10.1007/s00270-019-02243-4

Source DB:  PubMed          Journal:  Cardiovasc Intervent Radiol        ISSN: 0174-1551            Impact factor:   2.740


  6 in total

1.  2D-perfusion angiography for intra-procedural endovascular treatment response assessment in chronic mesenteric ischemia: a feasibility study.

Authors:  Annette Thurner; Anne Marie Augustin; Thorsten Alexander Bley; Ralph Kickuth
Journal:  BMC Med Imaging       Date:  2022-05-16       Impact factor: 2.795

2.  Percutaneous isolated hepatic perfusion (chemosaturation) with melphalan following right hemihepatectomy in patients with cholangiocarcinoma and metastatic uveal melanoma: peri- and post-interventional adverse events and therapy response compared to a matched group without prior liver surgery.

Authors:  C L A Dewald; L S Becker; S K Maschke; T C Meine; T A Alten; M M Kirstein; A Vogel; F K Wacker; B C Meyer; J B Hinrichs
Journal:  Clin Exp Metastasis       Date:  2020-10-09       Impact factor: 5.150

3.  Non-occlusive mesenteric ischemia (NOMI): evaluation of 2D-perfusion angiography (2D-PA) for early treatment response assessment.

Authors:  Lena S Becker; Klaus Stahl; Timo C Meine; Christian von Falck; Bernhard C Meyer; Cornelia L A Dewald; Nina Rittgerodt; Markus Busch; Sascha David; Frank Wacker; Jan B Hinrichs
Journal:  Abdom Radiol (NY)       Date:  2020-10

4.  2D-Perfusion Angiography Using Carbon Dioxide (CO2): A Feasible Tool to Monitor Immediate Treatment Response to Endovascular Therapy of Peripheral Arterial Disease?

Authors:  Cornelia L A Dewald; Lena S Becker; Sabine K Maschke; Timo C Meine; Bernhard C Meyer; Frank K Wacker; Jan B Hinrichs
Journal:  Cardiovasc Intervent Radiol       Date:  2020-12-16       Impact factor: 2.740

5.  Percutaneous Hepatic Perfusion (PHP) with Melphalan in Liver-Dominant Metastatic Uveal Melanoma: The German Experience.

Authors:  Cornelia L A Dewald; Mia-Maria Warnke; Roland Brüning; Martin A Schneider; Peter Wohlmuth; Jan B Hinrichs; Anna Saborowski; Arndt Vogel; Frank K Wacker
Journal:  Cancers (Basel)       Date:  2021-12-27       Impact factor: 6.639

Review 6.  Analysis of patient's X-ray exposure in hepatic chemosaturation procedures: a single center experience.

Authors:  Sebastian Ebel; Martin Reinhardt; Anne Bettina Beeskow; Felix Teske; Manuel Florian Struck; Rhea Veelken; Florian van Boemmel; Thomas Berg; Michael Moche; Matthias Gutberlet; Holger Gößmann; Timm Denecke
Journal:  BMC Med Imaging       Date:  2022-09-13       Impact factor: 2.795