Contact analysis of gap formation at dental implant-abutment interface under oblique loading: A numerical-experimental study.

Yiting He1, Alex Fok2, Conrado Aparicio2, Wei Teng1.   

Abstract

PURPOSE: To develop numerical and experimental methods for investigating the formation of micro-gaps and the change in contact area at the implant-abutment interface of two different connector designs under oblique cyclic loading.
MATERIALS AND METHODS: Samples (n = 10 per group) of two-piece implant systems with the conical connection (group A) and the external hexagonal connection (group B) were subjected to cyclic loading with increasing load amplitudes up to 220 N. After loading, the samples were scanned using micro-CT, with silver nitrate as a high-contrast penetrant, and the level of leakage was assessed using a discrete scoring system. Three-dimensional finite element (FE) analyses of the two implant systems were also conducted to reveal the micro-gap formation process, especially bridging of the internal abutment screw space. The experimental and numerical results for the bridging load were then compared.
RESULTS: 90% of the samples in group A showed leakage into the internal implant space at a load of around 100 N; while over 80% of those in group B did so at a load of around 40 N. This agreed with the FE analysis, which showed bridging of the internal implant space at loads similar to those measured for the two implant systems. Residual gaps of less than 1.49 μm were predicted for group A only after unloading.
CONCLUSIONS: The FE-predicted loads for bridging agreed well with those found experimentally for leakage to occur. The conical connection showed more resistance against formation of micro-gaps at the implant-abutment interface than the external hexagonal connection. Although the minimum load required to bridge the internal implant space was within the range of human biting force, the relation between bacterial invasion and micro-gaps needs further research.
© 2019 Wiley Periodicals, Inc.

Entities:  

Keywords:  finite element analysis; implant connection design; implant-abutment micro-gap; micro-computed tomography

Mesh:

Substances:

Year:  2019        PMID: 31087510     DOI: 10.1111/cid.12792

Source DB:  PubMed          Journal:  Clin Implant Dent Relat Res        ISSN: 1523-0899            Impact factor:   3.932


  3 in total

1.  A comparative biomechanical study of original and compatible titanium bases: evaluation of screw loosening and 3D-crown displacement following cyclic loading analysis.

Authors:  Rimantas Ožiūnas; Jurgina Sakalauskienė; Darius Jegelevičius; Gintaras Janužis
Journal:  J Adv Prosthodont       Date:  2022-04-27       Impact factor: 1.989

2.  Effects of cementless fixation of implant prosthesis: A finite element study.

Authors:  Hyeonjong Lee; Soyeon Park; Kung-Rock Kwon; Gunwoo Noh
Journal:  J Adv Prosthodont       Date:  2019-12-18       Impact factor: 1.904

Review 3.  Review of the Mechanical Behavior of Different Implant-Abutment Connections.

Authors:  Ana Sofia Vinhas; Carlos Aroso; Filomena Salazar; Paula López-Jarana; José Vicente Ríos-Santos; Mariano Herrero-Climent
Journal:  Int J Environ Res Public Health       Date:  2020-11-23       Impact factor: 3.390

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.