Vivek Y Reddy1, Petr Neuzil2, Tom de Potter3, Jan van der Heyden4, Selma C Tromp5, Benno Rensing4, Eva Jiresova2, Libor Dujka2, Veronika Lekesova2.   

Abstract

BACKGROUND: Patients with high stroke risk and atrial fibrillation who are unsuitable to oral anticoagulants (OACs) require other stroke prevention strategies. A novel permanent coil filter directly placed into both common carotid arteries (CCAs) was designed to capture emboli >1.4 mm in diameter.
OBJECTIVES: The multicenter, nonrandomized, first-in-human clinical CAPTURE (Carotid Artery Implant for Trapping Upstream Emboli for Preventing Stroke in Atrial Fibrillation Patients) trial sought to determine the feasibility and safety of bilateral CCA filter placement.
METHODS: Eligible patients had atrial fibrillation, CHA2DS2-VASc (Congestive heart failure, Hypertension, Age 75 years, Diabetes, Stroke/transient ischemic attack, Vascular disease, Age 65 to 74 years, Sex category) ≥2, OAC unsuitability, CCA size 4.8 to 9.8 mm, and no carotid stenosis >30%. Under ultrasound guidance, after direct transcutaneous carotid puncture with a 24-gauge needle, a motorized unit expels the filter to unfurl in the artery. Patients received aspirin/clopidogrel for 3 months, and aspirin thereafter. Primary endpoints were: 1) procedural success-bilateral, properly positioned CCA filters; and 2) 30-day incidence of major adverse events-death, stroke, major bleeding, filter migration, CCA thrombus, or stenosis. Carotid ultrasounds were conducted post-procedure, pre-discharge, at 1 week, and at 1, 3, 6, and 12 months.
RESULTS: At 3 centers, 25 patients were enrolled: age 71 ± 9 years, CHA2DS2-VASc = 4.4 ± 1.0, prior embolism in 48%. Procedure success was 92% (23 of 25 patients); 1 patient had unilateral deployment. There were no device/procedure-related major adverse events; minor puncture site hematomas/edema occurred in 5 of 25 (20%). After 6-month mean follow-up, asymptomatic thrombi were detected in 4 patients (1 bilateral, 4 unilateral), adjudicated as captured (n = 3), unclassified (n = 2), or in situ (n = 0). In all patients, the thrombi dissolved with subcutaneous heparin. In 1 patient, 2 device/procedure-unrelated minor strokes occurred.
CONCLUSIONS: Permanent carotid filter placement for stroke prophylaxis is technically feasible and safe. (Carotid Artery Implant for Trapping Upstream Emboli for Preventing Stroke in Atrial Fibrillation Patients [CAPTURE]; NCT03571789).
Copyright © 2019 The Authors. Published by Elsevier Inc. All rights reserved.

Entities:  

Keywords:  atrial fibrillation; carotid filter; common carotid artery; embolic protection; oral anticoagulation; stroke prevention

Year:  2019        PMID: 31085320     DOI: 10.1016/j.jacc.2019.04.035

Source DB:  PubMed          Journal:  J Am Coll Cardiol        ISSN: 0735-1097            Impact factor:   24.094


  1 in total

Review 1.  Carotid Atherosclerosis in Patients with Atrial Fibrillation.

Authors:  Zhaojia Wang; Panagiotis Korantzopoulos; Tong Liu
Journal:  Curr Atheroscler Rep       Date:  2019-11-29       Impact factor: 5.113

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.