Nazmiye Şen1, Ibrahim Bülent Şermet1, Nezahat Gürler2.
Abstract
PURPOSE: Limited data is available regarding the differences for possible microleakage problems and fitting accuracy of zirconia versus titanium abutments with various connection designs. The purpose of this in vitro study was to investigate the effect of connection design and abutment material on the sealing capability and fitting accuracy of abutments.Entities:
Keywords: Abutment; Dental implant; Fit accuracy; Implant-abutment connection design; Microbial sealing; Zirconia
Year: 2019 PMID: 31080571 PMCID: PMC6491357 DOI: 10.4047/jap.2019.11.2.105
Source DB: PubMed Journal: J Adv Prosthodont ISSN: 2005-7806 Impact factor: 1.904
Characteristics of the implants and abutments used
| Test groups | Abutments (Lot no.) | Implants (Lot no.) | Material | Connection design | Implant diameter × Length (mm) | Manufacturer |
|---|---|---|---|---|---|---|
| IC-Ti | Procera Esthetic | Nobel Parallel | Titanium | Internal conical connection (Titanium screw-retained) | 4.3 × 10.0 | Nobel Biocare, Göteborg, Sweden |
| Abutment CC RP | CC RP | |||||
| 1.5 mm #5 | (12103619) | |||||
| (13060524) | ||||||
| IC-Zr | Esthetic Abutment CC | Nobel Parallel | Zirconia | Internal conical connection (Titanium screw-retained) | 4.3 × 10.0 | Nobel Biocare, Göteborg, Sweden |
| RP 1.5 mm | CC RP | |||||
| (12098745) | (12103619) | |||||
| IT-Ti | Procera Esthetic | Replace Select | Titanium | Internal tri-channel connection (Titanium screw-retained) | 4.3 × 10.0 | Nobel Biocare, Göteborg, Sweden |
| Abutment NobRpl RP | Straight TiU RP | |||||
| 1 mm #10 | (12101424) | |||||
| (13061407) | ||||||
| IT-Zr | Esthetic Abutment | Replace Select | Zirconia | Internal tri-channel connection (Titanium screw-retained) | 4.3 × 10.0 | Nobel Biocare, Göteborg, Sweden |
| NobRpl RP 1 mm | Straight TiU RP | |||||
| (12098515) | (12101424) | |||||
| EH-Ti | Procera Esthetic | NOBELSPEEDY | Titanium | External hexagon (Titanium screw-retained) | 4.0 × 10.0 | Nobel Biocare, Göteborg, Sweden |
| Abutment BmkSyst RP | Groovy RP | |||||
| 1 mm #10 | (12107497) | |||||
| (12106471) | ||||||
| EH-Zr | Esthetic Abutment | NOBELSPEEDY | Zirconia | External hexagon (Titanium screw-retained) | 4.0 × 10.0 | Nobel Biocare, Göteborg, Sweden |
| BmkSyst RP 1 mm | Groovy RP | |||||
| (12096030) | (12107497) |
Study design
Fig. 1Tested abutments.
Mean ± SD values of leaked colonies
| CFUs/1 μL of Broth | ||||
|---|---|---|---|---|
| Mean Test Groups (torqued at 35 N/cm) (N = 30, n = 5) | Positive Control Group (torqued at 10 N/cm) (N = 6, n = 1) | |||
| Connection Design, Abutment Material | Internal conical connection (IC) | IC-Zr | (6.7 ± 0.9) × 102a | 65 × 102 |
| IC-Ti | (11.0 ± 1.3) × 102b | 112 × 102 | ||
| Internal tri-channel connection (IT) | IT-Zr | (13.7 ± 0.9) × 102b | 149 × 102 | |
| IT-Ti | (10.5 ± 0.3) × 102b | 91 × 102 | ||
| External hexagonal connection (EH) | EH-Zr | (17.4 ± 2.6) × 102b | 176 × 102 | |
| EH-Ti | (26.9 ± 4.5) × 102c | 120 × 102 | ||
Identical letters imply no significant difference (P > .05) according to Mann-Whitney U tests with Bonferroni-Holm correction.
Mean ± SD values of marginal misfit at the IAI
| Marginal Misfit (μm) | Connection Design, Abutment Material | |||||
|---|---|---|---|---|---|---|
| Internal conical connection (IC) | Internal tri-channel connection (IT) | External hexagonal connection (EH) | ||||
| IC-Zr (n = 5) | IC-Ti (n = 5) | IT-Zr (n = 5) | IT-Ti (n = 5) | EH-Zr (n = 5) | EH-Ti (n = 5) | |
| Mesial | 4.0 ± 0.9b | 2.0 ± 0.6a | 9.6 ± 1.3c | 5.4 ± 0.3b | 22.7 ± 9.1d | 10.5 ± 7.4c |
| Distal | 2.7 ± 0.5a | 5.3 ± 1.2b | 17.1 ± 2.6d | 11.3 ± 0.9c | 19.5 ± 6.8d | 12.7 ± 5.3c |
| Buccal | 2.9 ± 1.6a | 1.8 ± 0.3a | 15.2 ± 1.9c,d | 8.9 ± 1.1b,c | 16.8 ± 8.5d | 15.4 ± 2.7d |
| Palatinal | 3.5 ± 1.1a,b | 2.3 ± 0.7a | 6.5 ± 0.4b | 12.0 ± 1.7c | 18.9 ± 3.8d | 10.3 ± 5.5c |
Identical letters imply no significant difference (P > .05) according to Mann-Whitney U tests with Bonferroni-Holm correction.
Fig. 2Scanning electron micrographs of marginal misfit at implant-abutment interface (original magnification × 2000).
(A) IC-Ti; (B) IC-Zr; (C) IT-Ti; (D) IT-Zr; (E) EH-Ti; (F) EH-Zr and (original magnification × 100).
(a) IC-Ti; (b) IC-Zr; (c) IT-Ti; (d) IT-Zr; (e) EH-Ti, and (f) EH-Zr.