Literature DB >> 31079818

Adverse Events Associated with the Use of Guide Extension Catheters during Percutaneous Coronary Intervention: Reports from the Manufacturer and User Facility Device Experience (MAUDE) database.

Yuefeng Chen1, Arhum A Shah1, Evan Shlofmitz1, Nauman Khalid1, Anees Musallam1, Jaffar M Khan1, Micaela Iantorno1, Deepakraj Gajanana1, Toby Rogers2, Hayder Hashim1, Nelson L Bernardo1, Ron Waksman3.   

Abstract

BACKGROUND/
PURPOSE: We aimed to assess the reported complications and event modes for the GuideLiner and Guidezilla extension catheters. METHODS/MATERIALS: The US Food and Drug Administration (FDA) Manufacturer and User Facility Device Experience (MAUDE) database was queried for reported events.
RESULTS: Of the 65 cases with reported GuideLiner-related issues, 15 (23%) involved the inability to pass equipment through or damage to percutaneous coronary intervention (PCI) devices in the GuideLiner catheter, 38 (58%) involved GuideLiner catheter fracture, 9 (14%) involved coronary artery dissection, 2 (3%) involved coronary artery perforation, and 1 (1.5%) involved thrombus formation in the catheter. Of the 408 cases with reported Guidezilla-related issues, 53 (13%) involved inability to pass or damaged PCI devices into the Guidezilla catheter, 117 (29%) involved inability to advance the Guidezilla catheter to the target lesion, 59 (14%) involved kinked Guidezilla catheter, mostly because of partial or complete catheter fracture upon further investigation, 164 (40%) involved a broken Guidezilla catheter, 10 (2.5%) involved coronary artery dissection, 2 (0.5%) involved coronary artery perforation, 1 (0.2%)involved aortic dissection, 1 (0.2%) involved thrombosis formation, and 1 (0.2%) involved no-reflow phenomenon.
CONCLUSIONS: Findings from the MAUDE database highlight the complications and modes of events associated with the use of GuideLiner and Guidezilla extension catheters.
SUMMARY: To assess the reported complications and event modes for the GuideLiner and Guidezilla extension catheters, the US Food and Drug Administration (FDA) Manufacturer and User Facility Device Experience (MAUDE) database was queried. There were more reports on Guidezilla-related events during the search period. The events for both extension catheters mainly involved the inability to pass equipment through or damage to percutaneous coronary intervention (PCI) devices in the extension catheter, extension catheter fracture, coronary artery dissection and perforation and, occasionally, the death of the patients.
Copyright © 2019 Elsevier Inc. All rights reserved.

Entities:  

Keywords:  Complications; Guide extension catheter; Percutaneous coronary intervention

Mesh:

Year:  2019        PMID: 31079818     DOI: 10.1016/j.carrev.2019.02.016

Source DB:  PubMed          Journal:  Cardiovasc Revasc Med        ISSN: 1878-0938


  3 in total

1.  Stroke thrombectomy catheter for aspiration of refractory or inaccessible clot in acute myocardial infarction.

Authors:  Jacek Klaudel; Dariusz Surman; Krzysztof Pawłowski; Wojciech Trenkner
Journal:  Postepy Kardiol Interwencyjnej       Date:  2022-04-11       Impact factor: 1.065

2.  Application of a rapid exchange extension catheter technique in type B2/C nonocclusive coronary intervention via a transradial approach.

Authors:  Hong-Chao Wang; Wei Lu; Zi-Han Gao; Ya-Nan Xie; Jie Hao; Jin-Ming Liu
Journal:  World J Clin Cases       Date:  2021-04-26       Impact factor: 1.337

3.  A simple yet novel solution to prevent stent stripping in Guidezilla use.

Authors:  Kerim Esenboga; Kaan Akın; Nil Ozyuncu; Turkan Seda Tan; Durmus Eralp Tutar
Journal:  Anatol J Cardiol       Date:  2020-09       Impact factor: 1.596

  3 in total