Can the blood gas analyser results be believed? A prospective multicentre study comparing haemoglobin, sodium and potassium measurements by blood gas analysers and laboratory auto-analysers.

Katherine E Triplett1, Bradley A Wibrow1,2, Richard Norman3, Dana A Hince4, Liesel E Hardy5, Samantha Tan6, Kwok M Ho7, Matthew H Anstey1,2.   

Abstract

Blood gas analysers are point-of-care testing devices used in the management of critically ill patients. Controversy remains over the agreement between the results obtained from blood gas analysers and laboratory auto-analysers for haematological and biochemistry parameters. We conducted a prospective analytical observational study in five intensive care units in Western Australia, in patients who had a full blood count (FBC), urea, electrolytes and creatinine (UEC), and a blood gas performed within 1 h of each other during the first 24 h of their intensive care unit admission. The main outcome measure was to determine the agreement in haemoglobin, sodium, and potassium results between laboratory haematology and biochemistry auto-analysers and blood gas analysers. A total of 219 paired tests were available for haemoglobin and sodium, and 215 for potassium. There was no statistically significant difference between the results of the blood gas and laboratory auto-analysers for haemoglobin (mean difference -0.35 g/L, 95% confidence interval (CI) -1.20 to 0.51, P = 0.425). Although the mean differences between the two methods were statistically significant for sodium (mean difference 1.49 mmol/L, 95% CI 1.23-1.76, P < 0.0001) and potassium (mean difference 0.19 mmol/L, 95% CI 0.15-0.24, P < 0.0001), the mean biases on the Bland-Altman plots were small and independent of the magnitude of the measurements. The two methods of measurement for haemoglobin, sodium and potassium agreed with each other under most clinical situations when their values were within or close to normal range suggesting that routine concurrent blood gas and formal laboratory testing for haemoglobin, sodium and potassium concentrations in the intensive care unit is unwarranted.

Entities:  

Keywords:  Intensive care; haemoglobin; point-of-care testing; potassium; sodium

Mesh:

Substances:

Year:  2019        PMID: 31070468     DOI: 10.1177/0310057X19840046

Source DB:  PubMed          Journal:  Anaesth Intensive Care        ISSN: 0310-057X            Impact factor:   1.669


  3 in total

1.  Point of Care Testing of Serum Electrolytes and Lactate in Sick Children.

Authors:  Aashima Dabas; Shipra Agrawal; Vernika Tyagi; Shikha Sharma; Vandana Rastogi; Urmila Jhamb; Pradeep Kumar Dabla
Journal:  EJIFCC       Date:  2021-06-29

2.  Comparison of Hemoglobin Values Obtained by Arterial Blood Gas Analysis versus Laboratory Method during Major Head-and-Neck Surgeries.

Authors:  Sunil Rajan; Pulak Tosh; Maria Isaac; Niranjan Kumar Sasikumar; Avanthi Subramanian; Jerry Paul; Lakshmi Kumar
Journal:  Anesth Essays Res       Date:  2022-06-27

3.  Hyponatraemia despite isotonic maintenance fluid therapy: a time series intervention study.

Authors:  Milan Chromek; Åsa Jungner; Niclas Rudolfson; David Ley; Detlef Bockenhauer; Lars Hagander
Journal:  Arch Dis Child       Date:  2020-10-28       Impact factor: 3.791

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.