Literature DB >> 31070117

Accuracy of otoacoustic emissions, and automated and diagnostic auditory brainstem responses, in high-risk infants.

W Khaimook1, D Pantuyosyanyong1, P Pitathawatchai1.   

Abstract

OBJECTIVES: This study aimed to compare the diagnostic reliabilities of transient evoked otoacoustic emissions, automated auditory brainstem responses and brainstem auditory evoked responses for detecting hearing loss, and to use the information regarding hearing level of automated auditory brainstem responses for planning rehabilitation.
METHODS: A total of 144 high-risk infants (288 ears) completed the 3 hearing tests. The sensitivity and specificity of otoacoustic emissions and automated auditory brainstem responses were compared using the chi-square test.
RESULTS: Automated auditory brainstem response was the most reliable test of hearing levels, with a sensitivity of 91.7 per cent and specificity of 92.1 per cent; the sensitivity of otoacoustic emissions was 78.7 per cent and the specificity was 88.8 per cent.
CONCLUSION: Automated auditory brainstem responses have acceptably high sensitivity and specificity. Additionally, the hearing level from automated auditory brainstem responses can help the screeners explain to the parents the importance of further diagnosis and rehabilitation.

Entities:  

Keywords:  Auditory Brainstem Responses; Data Accuracy; Infants

Mesh:

Year:  2019        PMID: 31070117     DOI: 10.1017/S0022215119000872

Source DB:  PubMed          Journal:  J Laryngol Otol        ISSN: 0022-2151            Impact factor:   1.469


  1 in total

1.  Neonatal Hearing Screening: Prevalence of Unilateral and Bilateral Hearing Loss and Associated Risk Factors.

Authors:  Susan Sabbagh; Marzieh Amiri; Maryam Khorramizadeh; Zahra Iranpourmobarake; Mansoureh Nickbakht
Journal:  Cureus       Date:  2021-06-26
  1 in total