Literature DB >> 3106892

Incendiary characteristics of a new laser-resistant endotracheal tube.

D M Hayes, D M Gaba, R L Goode.   

Abstract

Endotracheal tube fires are the most frequent complication of laser surgery of the larynx. We investigated the incendiary characteristics of a new laser-resistant endotracheal tube, compared to polyvinyl chloride (PVC) and red rubber (RR) tubes, with different gas mixtures. A CO2 laser was focused on each tube and the probability of fire was assessed with oxygen fractions (FiO2) of 30%, 40%, 50%, 70%, and 100%, the balance being N2 or N2O. The laser-resistant tube could not be penetrated by a single laser pulse, regardless of laser energy or gas composition. PVC and RR tubes were readily ignited, the probability of ignition being related to oxygen concentration and laser energy. Nitrous oxide readily supported combustion. We suggest that laser surgery of the larynx can be most safely performed with a laser-resistant endotracheal tube and with gas mixtures containing the minimal safe O2 concentration (without nitrous oxide).

Entities:  

Mesh:

Substances:

Year:  1986        PMID: 3106892     DOI: 10.1177/019459988609500109

Source DB:  PubMed          Journal:  Otolaryngol Head Neck Surg        ISSN: 0194-5998            Impact factor:   3.497


  3 in total

Review 1.  Airway fire during tracheostomy: prevention strategies for surgeons and anaesthetists.

Authors:  M L Rogers; R W Nickalls; E T Brackenbury; F D Salama; M G Beattie; A G Perks
Journal:  Ann R Coll Surg Engl       Date:  2001-11       Impact factor: 1.891

2.  Nitrous oxide is contraindicated in endoscopic surgery.

Authors:  M Sosis
Journal:  Can J Anaesth       Date:  1987-09       Impact factor: 5.063

3.  New Metal Shield for the Endotracheal Tube during Laser Endolaryngeal Surgery.

Authors:  Nasser-Nagieb Mohamed; Mohammad-Waheed El-Anwar
Journal:  Iran J Otorhinolaryngol       Date:  2018-09
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.