Literature DB >> 31067530

MRI Metal Artifact Reduction Sequence for Auditory Implants: First Results with a Transcutaneous Bone Conduction Implant.

Wilhelm Wimmer1,2, Arsany Hakim3, Claus Kiefer4, Manuela Pastore-Wapp4, Lukas Anschuetz1,2, Marco D Caversaccio1,2, Franca Wagner5.   

Abstract

OBJECTIVE: Magnetic resonance imaging (MRI) is often limited in patients with auditory implants because of the presence of metallic components and magnets. The aim of this study was to evaluate the clinical usefulness of a customized MRI sequence for metal artifact suppression for patients with implants in the temporal bone region, specifically patients with a transcutaneous bone conduction implant.
METHODS: Two whole head specimens were unilaterally implanted with a transcutaneous bone conduction implant. MRI examinations with and without a primarily self-build sequence (SEMAC-VAT WARP) for metal artifact suppression were performed. The diagnostic usefulness of the acquired MRI scans was rated independently by two neuroradiologists. The sequence was also used to acquire postimplantation follow-up MRI in a patient with a transcutaneous bone conduction implant.
RESULTS: The customized SEMAC-VAT WARP sequence significantly improved the diagnostic usefulness of the postimplantation MRIs. The image acquisition time was 12 min and 20 s for the T1-weighted and 12 min and 12 s for the T2-weighted MRI. There was good agreement between the two blinded raters (Cohen's κ = 0.61, p < 0.001).
CONCLUSION: The sequence for metal artifact reduction optimized in Bern enables MRI at 1.5 T in patients with active transcutaneous bone conduction implants without sacrificing diagnostic imaging quality. Particularly on the implanted side, imaging of intracranial and supra- and infratentorial brain pathologies is clinically more valuable than standard diagnostic MRI without any artifact reduction sequences.
© 2019 S. Karger AG, Basel.

Entities:  

Keywords:  Artifact reduction; Bonebridge system; Magnetic resonance imaging; SEMAC-VAT; WARP

Mesh:

Substances:

Year:  2019        PMID: 31067530     DOI: 10.1159/000500513

Source DB:  PubMed          Journal:  Audiol Neurootol        ISSN: 1420-3030            Impact factor:   1.854


  3 in total

Review 1.  Recent advances in MRI of the head and neck, skull base and cranial nerves: new and evolving sequences, analyses and clinical applications.

Authors:  Philip Touska; Steve E J Connor
Journal:  Br J Radiol       Date:  2019-09-24       Impact factor: 3.039

2.  Multicenter Results With an Active Transcutaneous Bone Conduction Implant in Patients With Single-sided Deafness.

Authors:  Alexander M Huber; Bernd Strauchmann; Marco D Caversaccio; Wilhelm Wimmer; Thomas Linder; Nicola De Min; John-Martin Hempel; Marlene Pollotzek; Henning Frenzel; Frauke Hanke; Christof Röösli
Journal:  Otol Neurotol       Date:  2022-02-01       Impact factor: 2.311

Review 3.  Safety of active auditory implants in magnetic resonance imaging.

Authors:  Guy Fierens; Nina Standaert; Ronald Peeters; Christ Glorieux; Nicolas Verhaert
Journal:  J Otol       Date:  2021-01-06
  3 in total