Fully Automated Determination of Phosphatidylethanol 16:0/18:1 and 16:0/18:2 in Dried Blood Spots.

Marc Luginbühl1, Stefan Gaugler2, Wolfgang Weinmann1.   

Abstract

PURPOSE: Direct alcohol markers are widely applied during abstinence monitoring, driving aptitude assessments and workplace drug testing. The most promising direct alcohol marker was found to be phosphatidylethanol (PEth). Compared to other markers it shows a long window of detection due to accumulation in blood. To facilitate and accelerate the determination of PEth in DBS, we developed a fully automated analysis approach.
METHODS: The validated and novel online-SPE-LC-MS/MS method with automated sample preparation using a CAMAG DBS-MS 500 system reduces manual sample preparation to an absolute minimum, only requiring calibration and quality control DBS.
RESULTS: During the validation process, the method showed a high extraction efficiency (>88%), linearity (correlation coefficient >0.9953), accuracy and precision (within ±15%) for the determination of PEth 16:0/18:1 and PEth 16:0/18:2. Within a run time of about 7 min, the two monitored analogs could be baseline separated. A method comparison in liquid whole blood of 28 authentic samples from alcohol use disorder patients showed a mean deviation of less than 2% and a correlation coefficient of >0.9759. The comparison with manual DBS extraction showed a mean deviation of less than 8% and a correlation coefficient of >0.9666.
CONCLUSIONS: The automated analysis of PEth in DBS can provide a fast and accurate solution for abstinence monitoring. In contrast to the manual extraction of PEth in DBS, no laborious sample preparation is required with this automated approach. Furthermore, the application of the internal standard by a spray module can compensate for extraction bias and matrix effects.
© The Author(s) 2019. Published by Oxford University Press. All rights reserved. For Permissions, please email: journals.permissions@oup.com.

Entities:  

Mesh:

Substances:

Year:  2019        PMID: 31062845     DOI: 10.1093/jat/bkz035

Source DB:  PubMed          Journal:  J Anal Toxicol        ISSN: 0146-4760            Impact factor:   3.367


  2 in total

1.  Measurement of the alcohol biomarker phosphatidylethanol (PEth) in dried blood spots and venous blood-importance of inhibition of post-sampling formation from ethanol.

Authors:  Olof Beck; Maria Mellring; Christian Löwbeer; Sabina Seferaj; Anders Helander
Journal:  Anal Bioanal Chem       Date:  2021-02-15       Impact factor: 4.142

2.  Blood Phosphatidylethanol Concentrations Following Regular Exposure to an Alcohol-Based Mouthwash.

Authors:  Gary M Reisfield; Scott A Teitelbaum; Joseph T Jones; Dana Mason; Max Bleiweis; Ben Lewis
Journal:  J Anal Toxicol       Date:  2021-11-09       Impact factor: 3.367

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.