Types of laryngeal electrodes
| Laryngeal nerve monitoring with surface laryngeal electrode | |||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|
| Types | Embedded in tracheal tube | Stick on surface electrodes on endotracheal tubes | |||||||||
| Brand | Medtronic Xomed (Jacksonville, FL)1 | NIM contact reinforced EMG tube (Jacksonville, FL)2 | NIM Flex reinforced EMG tube (Jacksonville, FL)2 | Medtronic Trivantage (Jacksonville, FL)2 | Inomed Medizintechnik (GmbH, Emmendingen, Teningen Germany)3 | Neurosign Magstim Inc (Wales, UK)4 | Lantern Neurosign Magstim (Wales UK)5 | Dragonfly Neurovision Medical (Ventura, CA)6 | Laryngeal Electrode, TechnoMed Europe (Amerikalaan, Netherlands)7 | Laryngeal Surface Invotec International (Jacksonville, FL)8 | |
| Sizes | Tube size ID/OD (mm) | Sizes Tube Fit I.D (mm) | |||||||||
| Age (yo)1 | 2 sizes | 2 sizes | 3 sizes | 2 sizes | 3 sizes | 2 sizes | |||||
| +5.0/7.8 | 10-12 | 6/9 | 6/8.4 | 5.0/6.5 | 32×29 mm 6-7 | 6-7 | 4-5 | 6-7.5 | 4-5 | 6-7.5 | |
| 6.0/8.8 | 13-15 | 7/10.5 | 6.5/8.9 | 6.0/8.2 | 32×37 mm 7-9.5 | 8-9 | 6-7 | 8-9.5 | 6-7 | ||
| 7.0/9.8 | >16 | 8/11.5 | 7/9.7 | 7.0/9.5 | 8-9 | 8-9 | |||||
| 8.0/10.8 | >16 | 7.5/10.3 | 8.0/10.7 | ||||||||
| 8/10.8 | 9.0/12 | ||||||||||
| 8.5/11.2 | |||||||||||
| Tracheal tube use | NA | NA | NA | NA | * | Any standard tracheal tube | Any standard | Any endotracheal Tube | * | Non-silicon tracheal tube | |
| tracheal tube or armoured tracheal tube | (non-silicon) Longer than 8 hours not recommended | ||||||||||
| FDA recall9 and field safety notice | Class 2 device recall for NIM EMG tracheal tube in 2016 due to exposed wire. | Class 2 Device recall for Dragonfly 2 Channel laryngeal surface electrode (Neurovision) in 2010 due to migration of adhesive creating an electrical short. Use of such defective electrodes could lead to failure in monitoring.9 | |||||||||
| Class 1 device recall for NIM Trivantage EMG tube in 2013 due to cuff leakage in the intraoperative cuff deflation necessitating reinflation or replacement of the deflated tube to ensure the continued ventilation of the patient. | |||||||||||
| Voluntary recall of NIM Flex EMG tube by Medtronic in 2013 due to potential inner lumen herniation constricting inner lumen of tracheal tube. | |||||||||||
| Class 2 device recall for NIM EMG tube in 2004 due to reports of wires out of channel leading to patients’ injuries | |||||||||||
| Complication Reported in literature | Perforation of cuff of EMG tube (Medtronic NIM Response®Nerve Integrity Monitoring System) from distortion of electrodes from use with a lightwand.10 | Separation of electrode sheath and kink at cable entry point in locally prepared EMG tube resulted in obstruction during removal of tube in 16 months old.14 | |||||||||
| Perforation of cuff of EMG tube (Medtronic Xomed) from dislocated electrodes from use with a glidescope and Rüsch stylet.11 | |||||||||||
| Ineffective ventilation resulting from cuff herniation of reinforced EMG tube (Medtronic Xomed, Nerve Integrity monitoring).12 | |||||||||||
| Tracheal Tear in a case of underlying tracheomalacia after using an NIM EMG tube (Medtronic).13 | |||||||||||
+prototype investigative, * No mention, NA Not applicable
1. White WM, Randolph GW, Hartnick CJ, Cunningham MJ. Recurrent laryngeal nerve monitoring during thyroidectomy and related cervical procedures in the pediatric population. Arch Otolaryngol Head Neck Surg 2009;135:88-94.
2. www. MedtronicENT-TechComms.com. Medtronic NIM EMG Endotracheal tube.
3. www. Inomed.com
4. www.mundinc.com/uploads/LE.pdf (magstim)
5. www.neurosignsurgical.com
6. www.neurovisionmedical.com
7. www. 510(k) summary www.technomed-europe.com
8. www.invotec.net/products/laryngealSurfaceElectrodes
9. www.accessdata.fda.gov
10. Kim HS, Park KS, Kang MH, Park CD. Damage to the cuff of EMG tube at endotracheal intubation by using a lightwand -A case report. Korean J Anesthesiol 2010;59 Suppl:S17-20.
11. de Kant-Pariama E, Preckel B, Schlack WS, van Dijkum EJ. Cuff perforation by dislocated electrodes of an electromyogram tube. Anesth Analg 2012;115:1473.
12. Oysu C1, Demir K. Life-threatening complication of recurrent laryngeal nerve monitoring with EMG reinforced silicone ETT. J Craniofac Surg 2011;22:2419-21.
13. Joshi M, Mardakh S, Yarmush J, Kamath H, Schianodicola J, Mendoza E. Intraoperatively Diagnosed Tracheal Tear after Using an NIM EMG ETT with Previously Undiagnosed Tracheomalacia. Case Rep Anesthesiol 2013;2013:568373.
14. McCormack J, Purdy R. Airway complication related to an electromyography tracheal tube. Paediatr Anaesth 2008;18:572-3.