Literature DB >> 31057257

Selection of laryngeal electrodes for intraoperative laryngeal nerve monitoring.

Hwan Ing Hee1.   

Abstract

Entities:  

Year:  2019        PMID: 31057257      PMCID: PMC6495620          DOI: 10.4103/joacp.JOACP_138_18

Source DB:  PubMed          Journal:  J Anaesthesiol Clin Pharmacol        ISSN: 0970-9185


× No keyword cloud information.
Madam, Neurophysiological monitoring for localization of the laryngeal nerve is achieved through surface contact between laryngeal muscles and specialized laryngeal electrode. These electrodes are commercially available as disposable surface adhesives or embedded within an electromyograph (EMG) tracheal tube. Besides understanding the structure and function,[1] it is prudent for anesthesiologists to be cognizant of commercially available options to allow discerning selection for the best fit that is essential for its function [Table 1]. Electrode embedded EMG tracheal tubes are commercially available in sizes from 6 to 9 mm internal diameter (I.D.). Compared to the earlier Medtronic Xomed tracheal tube, the newer TriVantage tracheal tube has a smaller outer diameter that is comparable with standard tracheal tubes. Its possible availability in 5 mm I.D. further improves its suitability for smaller adults and children. However, lack of ½ sizes with EMG tubes remains a major shortcoming for paediatric use. The other option is commercial surface laryngeal electrode adhesives which are available in sizes that fit tracheal tube ranging from 4 to 10 mm I.D, enabling utility in younger or smaller patients. These electrodes are designed to adhere onto preselected tracheal tubes of specified I.D. as recommended by manufacturers. To optimize the fit with the true vocal cords to allow detection of bilateral laryngeal responses, selected tracheal tube are prepared by adhering the surface adhesives to the tube at appropriate shaft position. Compared to the typical flush design of other laryngeal electrodes, the newer bulb design of Neurosign's lantern laryngeal electrode has the advantage of ensuring contact with the vocal cords even if tracheal tube position is suboptimal. Adhesion of contact electrodes onto the tracheal tube shaft nevertheless increases the rigidity and adds to its overall outer diameter. For children, the wider selection of commercial brands and sizes makes adhesive electrodes a feasible option though there is still lack of adhesive electrodes fitting tracheal tube sizes smaller than 4-mm I.D. When larger electrode than recommended is used, care must be taken to minimize overlapping of electrodes for proper functioning.
Table 1

Types of laryngeal electrodes

Laryngeal nerve monitoring with surface laryngeal electrode

TypesEmbedded in tracheal tubeStick on surface electrodes on endotracheal tubes
BrandMedtronic Xomed (Jacksonville, FL)1NIM contact reinforced EMG tube (Jacksonville, FL)2NIM Flex reinforced EMG tube (Jacksonville, FL)2Medtronic Trivantage (Jacksonville, FL)2Inomed Medizintechnik (GmbH, Emmendingen, Teningen Germany)3Neurosign Magstim Inc (Wales, UK)4Lantern Neurosign Magstim (Wales UK)5Dragonfly Neurovision Medical (Ventura, CA)6Laryngeal Electrode, TechnoMed Europe (Amerikalaan, Netherlands)7Laryngeal Surface Invotec International (Jacksonville, FL)8
SizesTube size ID/OD (mm)Sizes Tube Fit I.D (mm)
Age (yo)12 sizes2 sizes3 sizes2 sizes3 sizes2 sizes
+5.0/7.810-126/96/8.45.0/6.532×29 mm 6-76-74-56-7.54-56-7.5 7.5-10
6.0/8.813-157/10.56.5/8.96.0/8.232×37 mm 7-9.58-96-78-9.56-7
7.0/9.8>168/11.57/9.77.0/9.58-98-9
8.0/10.8>167.5/10.38.0/10.7
8/10.89.0/12
8.5/11.2
Tracheal tube useNANANANA*Any standard tracheal tubeAny standardAny endotracheal Tube*Non-silicon tracheal tube
tracheal tube or armoured tracheal tube(non-silicon) Longer than 8 hours not recommended
FDA recall9 and field safety noticeClass 2 device recall for NIM EMG tracheal tube in 2016 due to exposed wire.Class 2 Device recall for Dragonfly 2 Channel laryngeal surface electrode (Neurovision) in 2010 due to migration of adhesive creating an electrical short. Use of such defective electrodes could lead to failure in monitoring.9
Class 1 device recall for NIM Trivantage EMG tube in 2013 due to cuff leakage in the intraoperative cuff deflation necessitating reinflation or replacement of the deflated tube to ensure the continued ventilation of the patient.
Voluntary recall of NIM Flex EMG tube by Medtronic in 2013 due to potential inner lumen herniation constricting inner lumen of tracheal tube.
Class 2 device recall for NIM EMG tube in 2004 due to reports of wires out of channel leading to patients’ injuries
Complication Reported in literaturePerforation of cuff of EMG tube (Medtronic NIM Response®Nerve Integrity Monitoring System) from distortion of electrodes from use with a lightwand.10Separation of electrode sheath and kink at cable entry point in locally prepared EMG tube resulted in obstruction during removal of tube in 16 months old.14
Perforation of cuff of EMG tube (Medtronic Xomed) from dislocated electrodes from use with a glidescope and Rüsch stylet.11
Ineffective ventilation resulting from cuff herniation of reinforced EMG tube (Medtronic Xomed, Nerve Integrity monitoring).12
Tracheal Tear in a case of underlying tracheomalacia after using an NIM EMG tube (Medtronic).13

+prototype investigative, * No mention, NA Not applicable

1. White WM, Randolph GW, Hartnick CJ, Cunningham MJ. Recurrent laryngeal nerve monitoring during thyroidectomy and related cervical procedures in the pediatric population. Arch Otolaryngol Head Neck Surg 2009;135:88-94.

2. www. MedtronicENT-TechComms.com. Medtronic NIM EMG Endotracheal tube.

3. www. Inomed.com

4. www.mundinc.com/uploads/LE.pdf (magstim)

5. www.neurosignsurgical.com

6. www.neurovisionmedical.com

7. www. 510(k) summary www.technomed-europe.com

8. www.invotec.net/products/laryngealSurfaceElectrodes

9. www.accessdata.fda.gov

10. Kim HS, Park KS, Kang MH, Park CD. Damage to the cuff of EMG tube at endotracheal intubation by using a lightwand -A case report. Korean J Anesthesiol 2010;59 Suppl:S17-20.

11. de Kant-Pariama E, Preckel B, Schlack WS, van Dijkum EJ. Cuff perforation by dislocated electrodes of an electromyogram tube. Anesth Analg 2012;115:1473.

12. Oysu C1, Demir K. Life-threatening complication of recurrent laryngeal nerve monitoring with EMG reinforced silicone ETT. J Craniofac Surg 2011;22:2419-21.

13. Joshi M, Mardakh S, Yarmush J, Kamath H, Schianodicola J, Mendoza E. Intraoperatively Diagnosed Tracheal Tear after Using an NIM EMG ETT with Previously Undiagnosed Tracheomalacia. Case Rep Anesthesiol 2013;2013:568373.

14. McCormack J, Purdy R. Airway complication related to an electromyography tracheal tube. Paediatr Anaesth 2008;18:572-3.

Types of laryngeal electrodes +prototype investigative, * No mention, NA Not applicable 1. White WM, Randolph GW, Hartnick CJ, Cunningham MJ. Recurrent laryngeal nerve monitoring during thyroidectomy and related cervical procedures in the pediatric population. Arch Otolaryngol Head Neck Surg 2009;135:88-94. 2. www. MedtronicENT-TechComms.com. Medtronic NIM EMG Endotracheal tube. 3. www. Inomed.com 4. www.mundinc.com/uploads/LE.pdf (magstim) 5. www.neurosignsurgical.com 6. www.neurovisionmedical.com 7. www. 510(k) summary www.technomed-europe.com 8. www.invotec.net/products/laryngealSurfaceElectrodes 9. www.accessdata.fda.gov 10. Kim HS, Park KS, Kang MH, Park CD. Damage to the cuff of EMG tube at endotracheal intubation by using a lightwand -A case report. Korean J Anesthesiol 2010;59 Suppl:S17-20. 11. de Kant-Pariama E, Preckel B, Schlack WS, van Dijkum EJ. Cuff perforation by dislocated electrodes of an electromyogram tube. Anesth Analg 2012;115:1473. 12. Oysu C1, Demir K. Life-threatening complication of recurrent laryngeal nerve monitoring with EMG reinforced silicone ETT. J Craniofac Surg 2011;22:2419-21. 13. Joshi M, Mardakh S, Yarmush J, Kamath H, Schianodicola J, Mendoza E. Intraoperatively Diagnosed Tracheal Tear after Using an NIM EMG ETT with Previously Undiagnosed Tracheomalacia. Case Rep Anesthesiol 2013;2013:568373. 14. McCormack J, Purdy R. Airway complication related to an electromyography tracheal tube. Paediatr Anaesth 2008;18:572-3. Complications and Food and Drug Administration recall of EMG electrodes from market had been reported and are also summarized in Table 1. Reported complications of use of embedded EMG tracheal tubes included cuff perforation from dislocated or distorted electrodes during stylet-assisted intubation,[2] cuff herniation,[3] and tracheal tear.[4] Airway complication from electrode sheath malfunction has also been reported in locally prepared electromyography tube in paediatric use.[5] As recurrent laryngeal nerve monitoring is still relatively new. Knowledge of available sizing of electrodes for the best fit and awareness of problem from user experience are another important practical consideration for selection of the most appropriate laryngeal electrodes

Financial support and sponsorship

Nil.

Conflicts of interest

There are no conflicts of interest.
  5 in total

1.  Life-threatening complication of recurrent laryngeal nerve monitoring with EMG reinforced silicone ETT.

Authors:  Cagatay Oysu; Kadri Demir
Journal:  J Craniofac Surg       Date:  2011-11       Impact factor: 1.046

2.  Airway complication related to an electromyography tracheal tube.

Authors:  Jon McCormack; Robert Purdy
Journal:  Paediatr Anaesth       Date:  2008-06       Impact factor: 2.556

3.  Cuff perforation by dislocated electrodes of an clectromyogram tube.

Authors:  Esther de Kant-Pariama; Benedikt Preckel; Wolfgang S Schlack; Els J M Nieveen van Dijkum
Journal:  Anesth Analg       Date:  2012-12       Impact factor: 5.108

4.  Intraoperatively Diagnosed Tracheal Tear after Using an NIM EMG ETT with Previously Undiagnosed Tracheomalacia.

Authors:  Minal Joshi; Simon Mardakh; Joel Yarmush; H Kamath; Joseph Schianodicola; Ernesto Mendoza
Journal:  Case Rep Anesthesiol       Date:  2013-03-27

5.  The neural integrity monitor electromyogram tracheal tube: Anesthetic considerations.

Authors:  Glen Atlas; Marshall Lee
Journal:  J Anaesthesiol Clin Pharmacol       Date:  2013-07
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.