| Literature DB >> 31057235 |
Neha Pangasa1,2, Jaspal Singh Dali1, Kavita Rani Sharma1, Mona Arya1, Anant Vikram Pachisia1,3.
Abstract
BACKGROUND AND AIMS: Truview evo2 has been found to improve the glottic view when compared with the Miller blade in pediatric population. However, there is limited literature comparing it with Macintosh laryngoscope in children. We thus aimed to assess and compare Truview evo2 with the Macintosh laryngoscope for orotracheal intubation in children with regards to time to intubate, laryngoscopic view, ease of intubation, and associated hemodynamic changes.Entities:
Keywords: CL grade; Macintosh; Truview evo2; time to intubation
Year: 2019 PMID: 31057235 PMCID: PMC6495625 DOI: 10.4103/joacp.JOACP_118_18
Source DB: PubMed Journal: J Anaesthesiol Clin Pharmacol ISSN: 0970-9185
Figure 1Truview evo2 with size “small” blade and camera attached to the eyepiece
Demographic variables in the two groups
| Group M ( | Group T ( | ||
|---|---|---|---|
| Age (years) | 5.4±2.1 | 4.5±2.1 | 0.081 |
| Weight (kg) | 16.1±4.9 | 14.7±5.0 | 0.162 |
| Height (cm) | 105.0±11.7 | 100.8±12.7 | 0.117 |
| Female/Male, n (%) | 9/16 (36/64) | 11/14 (44/56) | 0.282 |
Group M=Laryngoscopy-intubation with Macintosh laryngoscope; Group T=Laryngoscopy-intubation with Truview evo2 laryngoscope
Comparison of modified CL grade and intubation difficulty score between the groups
| Parameter | Group M ( | Group T ( | |||
|---|---|---|---|---|---|
| Modified CL Grade | |||||
| Percentage | Percentage | ||||
| I | 13 | 52 | 22 | 88 | 0.003 |
| IIa | 9 | 36 | 3 | 12 | 0.023 |
| IIb | 3 | 12 | 0 | 0 | 0.037 |
| III | 0 | 0 | 0 | 0 | |
| IV | 0 | 0 | 0 | 0 | |
| IDS | |||||
| 0 | 12 | 48 | 22 | 88 | 0.001 |
| 1 | 5 | 20 | 2 | 8 | 0.111 |
| 2 | 8 | 32 | 1 | 4 | 0.005 |
| Overall IDS Median (IQR) | 1 (0-2) | 0 (0-0) | |||
Modified CL grade=Modified Cormack and Lehane grading.[16]; IDS=Intubation Difficulty Score.[17]; IQR=Inter quantile range
Changes in heart rate during laryngoscopy and intubation
| HR | Group M | Group T | |
|---|---|---|---|
| Baseline | 122.8±22.2 | 120.4±20.5 | 0.346 |
| 0 minute | 123.8±21.4 | 116.6±18.4 | 0.102 |
| 1 minute | 129.1±20.9 | 124.6±18.6 | 0.210 |
| 2 minutes | 127.1±20.2 | 122.5±17.6 | 0.199 |
| 3 minutes | 120.3±29.4 | 120.9±17.1 | 0.467 |
| 4 minutes | 123.6±20.0 | 119.5±17.0 | 0.218 |
| 5 minutes | 122.3±19.8 | 118.2±16.7 | 0.213 |
Changes in systolic blood pressure during laryngoscopy and intubation
| SBP | Group M, | Group T, | |
|---|---|---|---|
| Baseline | 109.0±12.3 | 106±13.3 | 0.206 |
| 0 minute | 92.7±11.3 | 88.8±11.5 | 0.117 |
| 1 minute | 97.2±10.7 | 95.2±11.5 | 0.259 |
| 2 minutes | 91.7±20.4 | 92.8±10.7 | 0.404 |
| 3 minutes | 94.6±11.5 | 91.4±10.4 | 0.157 |
| 4 minutes | 93.9±11.3 | 90.0±10.5 | 0.109 |
| 5 minutes | 93.0±11.3 | 89.4±10.6 | 0.121 |
SBP=Systolic blood pressure; DBP=Diastolic blood pressure; df=Degree of freedom; Mean Diff=Mean difference; SE=Standard error of difference; CI=Confidence interval; Cohen’s δ =Depicts estimated effect size