Small field correction factors determination for several active detectors using a Monte Carlo method in the Elekta Axesse linac equipped with circular cones.

Andrea Girardi1, Christian Fiandra, Francesca Romana Giglioli, Elena Gallio, Omar Hammad Ali, Riccardo Ragona.   

Abstract

A Monte Carlo (MC) method was used to determine small field output correction factors for several active detectors (Exradin A16, Exradin A26, PTW microLion, PTW microDiamond, Exradin W1 and IBA RAZOR) for an Elekta Axesse linac equipped with circular cones. MC model of the linac was built with the GamBet software, using the Penelope code system. The dose-to-water simulation for each cone, ranging from 5 to 30 mm of diameter size, was used to calculate field factors and the results were validated together with Gafchromic EBT3 film. Output factors (OFs) were measured with the active detectors and correction factors were determined using the MC results. The MC simulations agreed with films within 1.2%. OFs measured with Exradin W1 scintillator were in agreement within 0.8% with MC simulations. The Exradin A16 and A26 microchambers under-responded for small fields relative to the MC (-13.1% and  -4.6%, respectively). PTW microLion, IBA RAZOR and PTW microDiamond overestimated the output factor for the smallest field (+3.9%, +5.4 and  +7.1%, respectively). The present study pointed out that it is crucial to apply the appropriate correction factors in order to provide accurate measurements in small beams geometry. The results showed that the Exradin W1 can be used for very small field dosimetry without correction factors, which shall be contrariwise employed for other detectors.

Entities:  

Mesh:

Year:  2019        PMID: 31051491     DOI: 10.1088/1361-6560/ab1f26

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  3 in total

1.  Determination of correction factors in small MLC-defined fields for the Razor and microSilicon diode detectors and evaluation of the suitability of the IAEA TRS-483 protocol for multiple detectors.

Authors:  Andrew N McGrath; Samane Golmakani; Timothy J Williams
Journal:  J Appl Clin Med Phys       Date:  2022-06-02       Impact factor: 2.243

2.  The influence of small field output factors simulated uncertainties on the calculated dose in VMAT plans for brain metastases: a multicentre study.

Authors:  Stefania Clemente; Maria Daniela Falco; Elisabetta Cagni; Cinzia Talamonti; Mafalda Boccia; Eva Gino; Elena Lorenzini; Federica Rosica; Serenella Russo; Alessandro Alparone; Daniele Zefiro; Christian Fiandra
Journal:  Br J Radiol       Date:  2021-01-22       Impact factor: 3.039

3.  Feasibility study of using Stereotactic Field Diode for field output factors measurement and evaluating three new detectors for small field relative dosimetry of 6 and 10 MV photon beams.

Authors:  Attia Gul; Shigekazu Fukuda; Hideyuki Mizuno; Nakaji Taku; M Basim Kakakhel; Sikander M Mirza
Journal:  J Appl Clin Med Phys       Date:  2020-10-19       Impact factor: 2.243

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.