N C Ikoro, D A Johnson, P P Antich.
Abstract
Clinical dosimetry data are presented for the lower-energy x-ray beam of a Varian Clinac 1800 linear accelerator. This beam has comparable characteristics to single energy linear accelerators with the same stated 6-MV x-ray energy. The nominal beam energy was found to be 5.3 +/- 0.3 MV on the central axis. Beam quality expressed in terms of half-value layer in water was found to vary by less than 10% over the entire field. The surface doses are only slightly, but consistently, larger than those reported in the literature for other 6-MV linacs. Dosimetric results presented include central axis percentage depth dose (% DD) and tissue-maximum ratio (TMR), surface and buildup doses, output factors, and inverse square law applicability. The flatness and symmetry characteristics are within the manufacturer's specifications for both large and small fields.Entities:
Mesh:
Year: 1987 PMID: 3104744 DOI: 10.1118/1.596113
Source DB: PubMed Journal: Med Phys ISSN: 0094-2405 Impact factor: 4.071