Ross L Walenga1, Andrew H Babiskin1, Liang Zhao1. 1. Division of Quantitative Methods and Modeling, Office of Research and Standards, Office of Generic Drugs, Center for Drug Evaluation and Research, US Food and Drug Administration, Silver Spring, Maryland, USA.
Abstract
The development of generic, single-entity, drug-device combination products for orally inhaled drug products is challenging in part because of the complex nature of device design characteristics and the difficulties associated with establishing bioequivalence for a locally acting drug product delivered to the site of action in the lung. This review examines in silico models that may be used to support the development of generic orally inhaled drug products and how model credibility may be assessed.
The development of generic, single-entity, drug-device combination products for orally inhaled drug products is challenging in part because of the complex nature of device design characteristics and the difficulties associated with establishing bioequivalence for a locally acting drug product delivered to the site of action in the lung. This review examines in silico models that may be used to support the development of generic orally inhaled drug products and how model credibility may be assessed.
Status of Generic Orally Inhaled Drug Products in the United States
In a recent report, the Association for Accessible Medicines estimated that generic drugs produced $265 billion in savings in the United States in 2017 and accounted for 9 of 10 dispensed prescriptions.1 However, these benefits were not realized for the multibillion dollar market of complex drug–device combination orally inhaled drug products (OIDPs) because there were no approved generic drug products in the United States during this period. These increased costs are felt both by consumers and insurance providers, where the Centers for Disease Control has estimated that there are currently 27 million people in the United States2 with asthma and 16 million with chronic obstructive pulmonary disease.3 A generic drug developer that wishes to enter this market would need to design a product that demonstrates bioequivalence (BE) to the reference product in the multiple in vitro and in vivo studies currently recommended by the US Food and Drug Administration (FDA) while considering brand‐name drug product patent protections that may limit design decisions related to the device constituent parts of the drug–device combination product.4 In general, to obtain approval of an abbreviated new drug application (ANDA) for a generic drug, an ANDA applicant first must identify the previously approved drug product it seeks to duplicate, i.e., the reference listed drug (RLD), and must show, among other things, that the generic drug is bioequivalent to the RLD. A reference standard selected by the FDA is the specific drug product that the ANDA applicant must use in conducting any in vivo BE testing required to support approval of its ANDA. The reference standard, selected by the FDA, is ordinarily the RLD. For ease of the reader, this article will only use the term RLD when describing regulatory requirements and recommendations relating to BE. For more information regarding the distinction between an RLD and a reference standard, please consult the FDA's draft guidance for industry titled “Referencing Approved Drug Products in ANDA Submissions.” When final, this guidance will represent the FDA's current thinking on this topic.Considering that the majority of drug–device combination OIDPs are locally acting drug products, the FDA has established a weight‐of‐evidence approach for showing the BE of these products, which includes a combination of in vitro and in vivo studies, formulation sameness, and device similarity.5 This approach has been deemed necessary because the direct measurement of the drug concentration at the site of action is only possible during a surgical procedure,6, 7 rendering it impractical for a pivotal BE study. For metered dose inhalers (MDIs), the list of in vitro studies includes single actuation content at the beginning, middle, and end life stages; aerodynamic particle size distribution (APSD) at the beginning and end life stages; spray pattern at two different distances from the orifice; plume geometry; and priming and repriming. The in vivo studies include a fasting, single‐dose, two‐way crossover pharmacokinetic (PK) study in healthy adult subjects and either a clinical pharmacodynamic (PD) study in patients or a comparative clinical end point BE study in patients.5 The list of studies recommended for dry powder inhalers (DPIs) includes in vitro single actuation content at the beginning, middle, and end life stages and APSD at the beginning and end life stages as well as in vivo PK and either a PD or a comparative clinical end point BE study.5 Considering the device aspect of drug–device combination OIDPs, these products are recommended to provide a user interface that is similar to the RLD with similar operating principles to ensure therapeutic equivalence.8 The formulation for MDIs should be qualitatively and quantitatively the same as the RLD; for DPIs qualitative sameness is required, and if it is not quantitatively the same the differences should be justified with additional data.9 Taken together, a potential generic OIDP developer has to balance the appropriateness of passing all FDA‐recommended studies, producing a generic drug–device combination product that can be substituted for the RLD without additional training prior to use and/or without the intervention of a healthcare provider10 and observing any existing device patents that may exist. Meeting these standards may require many device or even formulation iterations before generating an approvable product.Considering the challenges that a potential complex generic OIDP developer faces, strategies are needed to reduce the expected cost and time of development. In silico modeling offers a relatively cost‐effective means of accelerating generic OIDP development. Computational fluid dynamics (CFD) is a physics‐based modeling technique that can predict fluid and particle transport inside realistic geometries. The primary advantage of CFD in the context of generic OIDP development is its ability to consider device‐related characteristics such as the device geometry itself, APSD, spray angle, spray velocity, and orifice diameter while predicting regional deposition fraction of the initial dose in various regions of interest. As a coarse approximation, regional deposition fraction predictions can provide insight into the in vivo performance of a given drug–device combination. However, an additional technique is needed to predict local and systemic drug concentrations. Physiologically‐based pharmacokinetic (PBPK) modeling is a compartmental technique that can predict local and systemic PK based on several formulation and physiological input parameters. When CFD and PBPK are combined, it is possible to predict in vivo performance of an OIDP based on the device and formulation. This review provides an overview of research using CFD and/or PBPK to predict the performance of OIDPs, with consideration for how these methods may be useful for generic development as well as how model credibility may be assessed.
CFD Modeling of OIDPs
Conceptually, CFD is a combination of fluid mechanics, computer science, and mathematics for which there is a wide variety of applications, including aerospace, process engineering, automotive, power generation, sports, and biomedical.11 The Navier–Stokes equations of fluid motion are a set of nonlinear, coupled, partial differential equations that form the basis of CFD.11 Because of their complexity, analytical solutions of the Navier‐Stokes equations are only available in highly simplified cases.12 Consequently, to solve most real‐world problems, a numerical solution technique is necessary.11 To implement this technique, the domain of interest is decomposed into a set of elements called a “mesh.11 Most CFD solvers use a finite‐volume method in which the Navier–Stokes equations are integrated over each volume, which produces a set of coupled algebraic equations that are then solved numerically.11 Special consideration is given for the handling of turbulent flow regimes, where several different strategies are available that include the less computationally intensive Reynolds‐averaged Navier‐Stokes (RANS) methods or a large‐eddy simulation (LES) method that uses considerably more resources.11 Particle transport may be simulated via a Lagrangian approach in which individual particle trajectories are predicted by using the governing ordinary differential equations or the particles may be treated as a separate continuum using an Eulerian approach.11 Upon completion of a simulation, the results include local predictions of velocity, pressure, and temperature values as well as other values that may include species mole fractions, turbulence parameters, and particle trajectory data.Aerosolized medicine deposition in human lungs has been investigated using CFD methods for nearly 3 decades, with early work focusing on deposition predictions within a single idealized lung bifurcation.13, 14 Several recent reviews have summarized much of the work that has occurred since the early efforts of Gradon and Orlicki13 and Hofmann and Balásházy,14 which include a comprehensive review by Longest and Holbrook15 of all in silico models used to predict lung deposition. Recently, Longest et al.16 followed up the previous review15 with updates in whole‐lung modeling capabilities using CFD, a discussion on the role of CFD with inhaler design, the use of CFD for development of new respiratory drug‐delivery studies, CFD studies in special populations, and a discussion on the merging of CFD and PK models. Reviews by Wong et al.17 and Ruzycki et al.18 focus on the role of CFD within inhaler design, where as Wong et al.17 also describe discrete element modeling (DEM), a computational technique that considers particles as three‐dimensional (3D) domains as opposed to the standard Lagrangian method, which considers particles as point masses. A combined approach using CFD and DEM is particularly useful for predicting DPI product performance, where DEM is capable of capturing agglomeration and deagglomeration processes because of the interactions of carrier and active pharmaceutical ingredient (API) particles, as summarized in recent reviews by Tong et al.19 and Yang et al.20 In addition to the use of CFD for characterizing device performance, some recent reviews have focused on methodologies for capturing the influence of lung anatomy and physiology on particle deposition from OIDPs. The status of conducting airway modeling methodology as well as the application of local ventilation boundary conditions and turbulence model selection was reviewed by Lin et al.21 It was noted by Lin et al.21 that although some studies have used RANS turbulence models for the prediction of aerosolized medicine deposition, several studies using an LES method have demonstrated good predictive capability, including a study by Jayaraju et al.22 in which deposition predictions were improved by using an LES method as compared with a RANS model. Hofemeier et al.23 reviewed subacinar models and their use in predicting deposition of aerosols in the deep regions of the lung. Altogether, considering the breadth of reviews available that describe the use of CFD for predicting lung deposition, the portion of this review that addresses CFD models will focus on recent developments not included in these reviews, with a special focus on the utility of CFD for generic inhaler design and considerations of lung models for capturing intersubject variability.
Predicting the Influence of Device and Formulation Parameters on Regional Lung Deposition Using CFD
To adequately compare a potential generic MDI to the RLD, a CFD simulation must be able to accurately characterize the spray, which requires consideration of the potential impact from all device and formulation differences. Spray characterization of MDIs using CFD is challenging because of the several physical processes that include droplet formation,24 the large temperature gradient at the device orifice,25 high initial velocity values,26 and rapid evaporation of the propellant.24 Several studies have considered MDI delivery to the lungs using CFD, with varying physical treatments.27, 28, 29, 30, 31, 32, 33, 34, 35, 36, 37 Proper characterization of an MDI spray requires the careful selection of orifice boundary conditions because this process is crucial to understanding device differences. There are no known studies that directly simulate the fluid behavior inside the MDI canister through atomization. One group has used theoretical mathematical formulations to produce both velocity and APSD boundary conditions at the orifice, indicating good agreement with experimental values using this approach.27, 28 Recently, this same group has developed a theoretical method for predicting velocity and APSD boundary conditions for mixtures of ethanol and 1,1,1,2‐tetrafluoroethane (HFA‐134a), as illustrated in Figure 1.38 The method is based on work by Fletcher39 and Clark,40 where the behavior of constituents in the metering and expansion chambers, as shown in Figure 1
a, is modeled using a homogeneous frozen flow model that does not permit evaporation in these regions but rather allows isentropic expansion.38, 41 Upon exit from the spray orifice, atomization is predicted using the model developed by Gavtash et al.42 as displayed in Figure 1
b, where the initial output is treated as a flat sheet using the linear instability sheet analysis framework as developed by Senecal et al.43 which predicts growth of the wave‐induced stabilities that are responsible for droplet breakup. Further development and application of methods such as these will be beneficial for future CFD studies seeking to compare a generic MDI to the corresponding RLD because they may directly capture the influence of device and formulation changes on APSD and other in vitro metrics.
PBPK Modeling Coupled With CFD to Predict Drug Absorption in the Lungs
It is widely considered that regional deposition values for locally acting OIDPs provide a good estimate of drug delivery. However, the precise characterization of the rate and extent of drug delivered to the site of action for these products requires lung tissue concentration measurements. Although it is possible to directly measure unbound drug in lung tissue interstitial fluid using a microdialysis technique, it must be done during open chest surgery, making it an impractical technique for comparing two drug products in a BE study.6, 7 In the absence of direct lung tissue concentration measurements, PBPK modeling is useful for understanding the differences in drug product performance because it is capable of predicting lung tissue concentrations for locally acting OIDPs. A PBPK model is a compartmental approach in which various tissues in the body (e.g., brain, lung, kidney, etc.) are represented by separate compartments with corresponding tissue volume and blood flow rate values, and it represents a more mechanistic approach when compared with a simple two‐compartment PK model.78 When paired with accurate regional deposition data, a PBPK model may be capable of predicting lung tissue concentrations with enough accuracy that results may be useful for drug development. In addition, PBPK is capable of simultaneously predicting pulmonary and gastrointestinal tract absorption, which is useful for understanding the competing effects of absorption from these two regions on systemic plasma concentration values. When paired with a validated CFD model capable of predicting regional deposition with a reasonable amount of accuracy, the combination of CFD and PBPK represents a fully in silico approach to OIDP development. This may be especially useful in early drug development, when many device and formulation changes are expected yet the firm may not wish to invest too much because of the perceived risk.Lung PBPK modeling for delivery of locally acting OIDPs is a relatively unexplored area when compared with PBPK modeling of solid oral dosage forms. PBPK modeling from a pulmonary toxicology perspective has a somewhat longer history, with published work available as early as 200879 and with several studies published afterward. A recent review by Bäckman et al.6 provides an overview of available lung PBPK models for pharmaceutical drug delivery and of those developed for private use by industry. Key considerations according to model structure and parameterization for the models described were divided into aerosol deposition, dissolution, nonabsorptive clearance, and absorptive clearance.6 The only commercially available PBPK software that was capable of addressing all four of these areas as identified by Bäckman et al.6 was Gastroplus (Simulations Plus, Rochester, NY), where other available packages such as the SimCyp Simulator (Certara USA, Princeton, NJ) and PK‐SIM (Bayer AG, Leverkusen, Germany) reduced dissolution to first‐order processes, which limited their ability to model region‐specific absorption. One notable example of a custom PBPK model was developed by Boger and Fridén,80 who used PBPK combined with PD modeling to predict values of forced expiratory volume after 1 second after the administration of albuterol sulfate using either an OIDP or an oral tablet, where the predicted values of forced expiratory volume after 1 second show reasonable agreement with the clinical data.Only a few studies have used the Pulmonary Compartmental Absorption & Transit (PCAT) model in Gastroplus to model absorption from OIDPs.70, 81, 82, 83 The effect of different carrier properties on systemic PK was investigated by Wu et al.81 who measured APSD from two formulations of albuterol sulfate using the Cyclocaps capsules Teva UK, Harlow, UK in an Aerolizer device, where glass beads were used as carrier particles as opposed to lactose because of the ease of surface modification. Values of maximum plasma concentration (Cmax) and FPF were increased by factors of 1.20 and 1.36, respectively, for the formulation where the glass bead surface was modified as opposed to the formulation with the unmodified glass beads.81 The predictive power of deposition and permeability estimates provided by the Gastroplus PCAT model when compared with in vivo deposition data and experimentally determined alveolar permeability data was tested by Salar‐Behzadi et al.82 where predictions of Cmax and area under the time‐concentration curve from time 0 to time t from Turbuhaler (budesonide DPI) when compared with available PK data were greatly improved by the addition of experimental data. Bäckman et al.83 used Gastroplus PCAT to model the exposure of a poorly soluble investigational compound, AZD5423, using a number of delivery methods, including different nebulizers and DPI devices, where it was found that total lung deposition values from an in vitro model were not alone predictive of differences in PK metrics, but when deposition pattern and dissolution were also considered, the predictions matched well with available PK data.Although most studies have used empirical or simplified analytical models to predict lung deposition for lung PBPK modeling of pharmaceutical drugs, two have used CFD.70, 84 A CFD model using a Lagrangian particle‐tracking method was used to predict emitted dose, MMAD, and FPF of four amiloride hydrochloride (HCi) formulations from an Aerosolizer device, where predicted values showed reasonable agreement with impactor‐based emitted dose data, although MMAD values were underpredicted.70 These predicted values were used as inputs for a Gastroplus PCAT model of nebulized amiloride HCl, where the predictions agreed reasonably well with available PK data85 in terms of Cmax and time to maximum plasma concentration, as illustrated in Figure 4, especially for the case in which the absorption rate constant in the pulmonary region was reduced by 50%.70 However, the model did overpredict area under the time‐concentration curve from time 0 to time ∞ (AUC0‐∞) because of its inability to capture the trough between the double peaks, where the double peaks are present because of competing absorption processes in the lung and the gastrointestinal tract.70 After the validation case with nebulized amiloride HCl, the model was used to predict PK profiles for the four DPI‐based formulations, where significant differences were observed.70
Although CFD and PBPK are useful tools for understanding the mechanisms associated with OIDP delivery, model credibility should be established for these tools to be used effectively, particularly for regulatory decision making. Validation for PBPK models typically involves comparing systemic PK end points, such as Cmax and area under the time‐concentration curve from time 0 to time t after administration of OIDPs. If PK data from an intravenously administered drug are available, this can be useful for predicting clearance parameters. Wu et al.81, Salar‐Behzadi et al.82 and Bäckman et al.83 used intravenous data to estimate clearance, whereas others did not, presumably because they were unavailable.70, 84 Some studies used data from an oral dosage form of the same drug to build an oral model with the purpose of increasing confidence in the lung model.70, 81 A primary limitation for validating lung PBPK models is that there is a lack of lung tissue concentration data, so it is generally not possible to validate the model against the true metrics of interest. Until this issue is solved either via collection of human lung tissue concentration data or some other approach, lung PBPK model validation will likely be limited to comparison with systemic plasma concentration values, which does not ensure that local concentration predictions are accurate.For CFD, several studies have validated the model via comparison of predicted regional deposition values with in vitro or in vivo data.29, 31, 34, 47, 69, 87, 88, 89 For comparison with in vitro data, a replica of the computational geometry may be generated using rapid prototyping, and regional deposition after actuation of the device into the geometry may be quantified using high‐performance liquid chromatography.31 Validation with in vivo data uses data collected via gamma scintigraphy with radiolabeled aerosols, where two‐dimensional images of deposition locations are used to estimate central and peripheral deposition amounts.47, 69 As shown in Figure 3, the regional deposition predictions of the Respimat Soft Mist inhaler drug delivery from Tian et al.47 were compared against gamma scintigraphy data. Another approach has been to use particle image velocimetry to experimentally measure local velocity values in an in vitro replica of the in silico model and then to compare the in silico predictions to the in vitro data on a qualitative basis.68, 90 Validating against particle image velocimetry data may be especially useful if deposition predictions are also validated against experimental deposition data.
Future Directions
The Office of Generic Drugs (OGD) in the Center for Drug Evaluation and Research at the FDA has funded two grants related to the in silico modeling of OIDPs prior to fiscal year 2018 using funds obtained from the Generic Drug User Fee Amendments regulatory science program.91 In 2014, a grant titled “A Predictive Multiscale Computational Tool for Simulation of Lung Absorption and Pharmacokinetics and Optimization of Pulmonary Drug Delivery” was awarded to CFD Research Corporation for the development of an integrated CFD‐PBPK lung model, which has been previously described in this article.84, 86, 91 The key gap that this grant addressed is the lack of PBPK models for OIDPs that take full advantage of CFD for its ability to both predict deposition and to capture the effects of mucociliary clearance. In addition, a grant titled “A Cluster‐Based Assessment of Drug Delivery in Asthmatic Small Airways” was awarded to the University of Iowa in 2016.91 This grant is currently ongoing, and the purpose of the project is to use CFD to explore intersubject variability of small‐airway delivery to asthmatic patients from OIDPs using a cluster‐based approach to identifying different asthma subgroups.91 As described earlier in this article, because in vitro and in vivo methods are unable to measure drug‐delivery variability in the small airways, it is believed that CFD may be a useful tool for quantifying this variability, where the results of this grant are expected to advance this capability.There are several examples in the literature demonstrating the usefulness of CFD for OIDP development, especially with respect to guiding device changes.63, 64, 65 However, several issues remain unsolved and must be addressed if CFD is to be capable of fully capturing the effects of device differences on OIDP performance. The definition of orifice boundary conditions remains a key challenge for MDI models because experimentally measured values of velocity, temperature, and APSD are only available downstream. If possible, a mechanistic model of atomization using CFD or another modeling strategy may be useful for solving this problem, where work by Gavtash and colleagues may represent a key step in this direction.27, 28 Further refinement of heat‐transfer modeling approaches may improve the accuracy of CFD‐based MDI models, especially for solution‐based MDIs, where the evaporation of ethanol in droplets from these formulations is sensitive to local temperature changes. Although not discussed at length in this review, evaporation model selection is another key issue for solution‐based MDIs, for which a key consideration is the characterization of nonideal mixture behavior.92, 93 With respect to OIDPs that use the Respimat Soft Mist inhaler device and for any potential generic OIDPs that reference that device, the modeling approaches described in the literature have shown promise for accurately predicting regional deposition.45, 46, 47, 48 From the perspective of generic OIDP development, future work to characterize the differences in device characteristics such as orifice diameter and chamber volume and device metrics such as spray angle, spray velocity, and spray duration would be useful for understanding how product differences may influence regional deposition. For DPIs, several studies have used CFD with a Lagrangian particle‐tracking approach to demonstrate how such models may be used to guide device changes.63, 64, 65 However, Lagrangian particle tracking requires several assumptions to account for agglomeration and deagglomeration of API and carrier particles. The combination of CFD and DEM modeling is capable of directly modeling these processes and with further validation may allow for the enhanced precision of formulation change effects on deposition and FPF. To this end, the OGD has recently funded two grants using funds obtained from the Generic Drug User Fee Amendments for the development of CFD‐DEM methodologies applicable to DPIs.94 Improvements in this area are expected to involve the refinement of models, especially with respect to the complete characterization of electrostatic forces and the generation of experimental data that may be used to validate the accuracy of agglomeration and deagglomeration modeling on a more direct basis than deposition‐based validation is capable of. In addition, as noted in the review by Yang et al.20 a study by Kaialy et al.95 has demonstrated that particle shape can affect DPI performance. Both new grants will consider the effect of particle shape on various performance metrics by using DEM to model either cylindrical particles or several spherical particles packed together to approximate irregular shapes. Another area of improvement to be addressed is the effect of humidity, where one grant will explore the effect of humidity on the triboelectric charge of dry powders.The intersubject variability of OIDP regional deposition in human airways has been considered,21, 34, 46, 76, 77 but a key challenge is the difficulty in establishing model credibility, especially with respect to small‐airway deposition predictions. Future work is needed to enhance confidence in the predictive capability of these models. In addition, to fully understand how device and formulation differences may affect OIDP performance, the interaction of these differences with intersubject human airway variability must also be understood. Several studies have combined OIDP device and realistic human airway geometries,29, 33, 34, 36, 45, 76, 77 but these studies tend to have a strong focus on either device or human airway differences without much consideration for interactions between the two geometries. A study that considers how device and formulation differences affect not only the mean regional deposition values but also the intersubject variability would be useful from a generic perspective. Another consideration for future models that consider human airway intersubject variability is the appropriateness of the rigid airway wall assumption applied by most models because it is known that human airway diameter and length change dynamically during inhalation and exhalation. Although several models have considered the effects of moving airways either indirectly76, 77 or directly,96 the impact of this assumption for regional deposition predictions as it relates to device type (i.e., MDI, DPI, etc.) is not well understood.Several promising PBPK models of locally acting OIDPs have been developed that have shown the ability to predict systemic plasma concentration values with reasonable accuracy. However, examples of lung PBPK modeling in the literature are few, and more examples are needed to explore the potential utility of these models with respect to guiding formulation and device changes. Validation based on systemic absorption is the current approach, which is particularly useful if the purpose of the model is to assess the probability of passing a PK study with a proposed generic OIDP and the RLD. However, if the purpose of the model is to assess the local BE of a generic OIDP in relation to the RLD, the key metric of interest is lung tissue concentration at different regions of the lung. In this scenario, the ideal experimental data for validation would be unbound drug in human lung tissue interstitial fluid. The difficulty of obtaining this type of data via microdialysis is very challenging because it may only be collected during a surgical procedure, rendering the regular use of such data for validation impractical. However, a few limited data sets may still prove useful for establishing overall credibility of lung PBPK modeling. Another area of potential improvement for lung PBPK models is mucociliary clearance methodology, where most models use a compartmental approach. Kannan and colleagues have addressed this issue in part by developing a Q3D methodology that is capable of local absorption prediction with much greater precision.84, 86 A Generic Drug User Fee Amendments–funded continuation of this work has recently been announced by the OGD, which will involve the expansion of the Q3D model to include respiratory airways and the application of the Q3D methodology to other lung geometries.94To date, CFD and PBPK models have generally established model credibility via comparison of predicted values with experimental data. However, there is no currently accepted standard for assessing how closely predicted values should match experimental data for a given application. One proposed methodology is verification, validation, and uncertainty quantification, as described by Pathmanathan et al.97 and Pathmanathan and Gray.98 A key concept described by this group is context of use, which refers to the level of evidence required to establish model credibility.97 The level of risk for a given model is an interaction of the model's influence in a given decision‐making process and also the consequence of the decision,97 which is illustrated in Figure 5. Higher risk models would require higher evidentiary standards if model credibility is to be reliably assessed. Once the level of evidence required is established for a given application, the group proposes that the model be assessed according to verification, validation, and uncertainty quantification.98 As defined by Pathmanathan and Gray,98
verification refers to the numerical accuracy of the model with respect to the mathematical formulation of the problem of interest, validation refers to accuracy of the model as compared with reality, and uncertainty quantification explores the influence of parameter sensitivity of the results of interest. Many of the ideas described by Pathmanathan et al.97 and Pathmanathan and Gray98 are echoed in the recently published standards released by the American Society of Mechanical Engineers (ASME) for establishing the credibility of medical devices, which have been referred to as ASME V&V 40.99 The ASME V&V 40 standards provide more specific details on how verification, validation, and uncertainty quantification methodology may be applied.99 Although none of the ideas expressed in the literature described here97, 98 or in ASME V&V 40 are specifically endorsed by the OGD or the Center for Drug Evaluation and Research for the review of OIDPs, they provide a useful means for developing a rigorous model that may prove useful for OIDP development.
Figure 5
Qualitative description of the interaction between model influence and decision consequence on model risk.
Qualitative description of the interaction between model influence and decision consequence on model risk.
Conclusions
There is a current lack of generic OIDPs on the US market, where increased costs and a reduction in accessibility are associated with this situation. To aid with the development of generic OIDPs, CFD and PBPK modeling may be useful for reducing the number of device and formulation changes by quantifying the influence of specific modifications. Research involving the CFD modeling of OIDPs has been ongoing for several decades, with much of the focus on the development of geometric lung models and improvements in regional deposition prediction accuracy. More precise modeling of device‐specific and formulation‐specific influences on OIDP performance with CFD is more recent, where most studies have focused on modeling device and formulation effects of DPIs. Other efforts have investigated the influence of orifice boundary conditions on MDI performance and have developed an experimentally validated model for products using the Respimat device. CFD models may also be useful for assessing intersubject variability of small‐airway deposition, but only limited work in this area has been completed thus far. PBPK models have been recently developed for the delivery of locally acting OIDPs to the lung, with a few examples in the literature. The current models may be useful for predicting the probability of passing a PK study that compares a potential generic OIDP with the RLD. Further enhancements in model precision and credibility may be needed to use PBPK models for comparing the local deposition of two OIDPs. To establish model credibility for both CFD and PBPK, researchers have typically used experimentally derived in vitro and in vivo data to compare against model predictions. The more rigorous verification, validation, and uncertainty quantification methodology may be useful for enhancing the credibility of CFD and PBPK models to be used during the OIDP development process. As model credibility and predictability are continually improved, it is expected that CFD and PBPK will increasingly occupy a more significant role in the generic OIDP development process, where expected benefits include reductions in time and cost.
Funding
No funding was received for this work.
Conflict of Interest
The authors declared no competing interests for this work.