Literature DB >> 31042994

A 4-channel, vector network analyzer microwave imaging prototype based on software defined radio technology.

Paul Meaney1, Alexander Hartov1, Selaka Bulumulla2, Timothy Raynolds1, Cynthia Davis3, Florian Schoenberger1, Sebastian Richter1, Keith Paulsen1.   

Abstract

We have implemented a prototype 4-channel transmission-based, microwave measurement system built on innovative software defined radio (SDR) technology. The system utilizes the B210 USRP SDR developed by Ettus Research that operates over a 70 MHz-6 GHz bandwidth. While B210 units are capable of being synchronized with each other via coherent reference signals, they are somewhat unreliable in this configuration and the manufacturer recommends using N200 or N210 models instead. For our system, N-series SDRs were less suitable because they are not amenable to RF shielding required for the cross-channel isolation necessary for an integrated microwave imaging system. Consequently, we have configured an external reference that overcame these limitations in a compact and robust package. Our design exploits the rapidly evolving technology being developed for the telecommunications environment for test and measurement tasks with the higher performance specifications required in medical microwave imaging applications. In a larger channel configuration, the approach is expected to provide performance comparable to commercial vector network analyzers at a fraction of the cost and in a more compact footprint.

Year:  2019        PMID: 31042994      PMCID: PMC6483785          DOI: 10.1063/1.5083842

Source DB:  PubMed          Journal:  Rev Sci Instrum        ISSN: 0034-6748            Impact factor:   1.523


  29 in total

1.  Three-dimensional microwave imaging of realistic numerical breast phantoms via a multiple-frequency inverse scattering technique.

Authors:  Jacob D Shea; Panagiotis Kosmas; Susan C Hagness; Barry D Van Veen
Journal:  Med Phys       Date:  2010-08       Impact factor: 4.071

2.  Dielectric food data for microwave sterilization processing.

Authors:  T Ohlsson; N E Bengtsson
Journal:  J Microw Power       Date:  1975-03

3.  Ultrawideband temperature-dependent dielectric properties of animal liver tissue in the microwave frequency range.

Authors:  Mariya Lazebnik; Mark C Converse; John H Booske; Susan C Hagness
Journal:  Phys Med Biol       Date:  2006-03-21       Impact factor: 3.609

4.  A large-scale study of the ultrawideband microwave dielectric properties of normal breast tissue obtained from reduction surgeries.

Authors:  Mariya Lazebnik; Leah McCartney; Dijana Popovic; Cynthia B Watkins; Mary J Lindstrom; Josephine Harter; Sarah Sewall; Anthony Magliocco; John H Booske; Michal Okoniewski; Susan C Hagness
Journal:  Phys Med Biol       Date:  2007-04-23       Impact factor: 3.609

5.  Image accuracy improvements in microwave tomographic thermometry: phantom experience.

Authors:  P M Meaney; K D Paulsen; M W Fanning; D Li; Q Fang
Journal:  Int J Hyperthermia       Date:  2003 Sep-Oct       Impact factor: 3.914

6.  A large-scale study of the ultrawideband microwave dielectric properties of normal, benign and malignant breast tissues obtained from cancer surgeries.

Authors:  Mariya Lazebnik; Dijana Popovic; Leah McCartney; Cynthia B Watkins; Mary J Lindstrom; Josephine Harter; Sarah Sewall; Travis Ogilvie; Anthony Magliocco; Tara M Breslin; Walley Temple; Daphne Mew; John H Booske; Michal Okoniewski; Susan C Hagness
Journal:  Phys Med Biol       Date:  2007-10-01       Impact factor: 3.609

7.  Electromagnetic breast imaging: results of a pilot study in women with abnormal mammograms.

Authors:  Steven P Poplack; Tor D Tosteson; Wendy A Wells; Brian W Pogue; Paul M Meaney; Alexander Hartov; Christine A Kogel; Sandra K Soho; Jennifer J Gibson; Keith D Paulsen
Journal:  Radiology       Date:  2007-03-30       Impact factor: 11.105

8.  Dielectrical spectroscopy of canine myocardium during acute ischemia and hypoxia at frequency spectrum from 100 kHz to 6 GHz.

Authors:  Serguei Y Semenov; Robert H Svenson; Vitaly G Posukh; Alexei G Nazarov; Yuri E Sizov; Alexander E Bulyshev; Alexander E Souvorov; Wendy Chen; Jackie Kasell; George P Tatsis
Journal:  IEEE Trans Med Imaging       Date:  2002-06       Impact factor: 10.048

9.  Microwave thermal imaging: initial in vivo experience with a single heating zone.

Authors:  P M Meaney; M W Fanning; K D Paulsen; D Lit; S A Pendergrass; Q Fang; K L Moodie
Journal:  Int J Hyperthermia       Date:  2003 Nov-Dec       Impact factor: 3.914

10.  Surface wave multipath signals in near-field microwave imaging.

Authors:  Paul M Meaney; Fridon Shubitidze; Margaret W Fanning; Maciej Kmiec; Neil R Epstein; Keith D Paulsen
Journal:  Int J Biomed Imaging       Date:  2012-04-10
View more
  1 in total

1.  Low Cost, High Performance, 16-Channel Microwave Measurement System for Tomographic Applications.

Authors:  Paul Meaney; Alexander Hartov; Timothy Raynolds; Cynthia Davis; Sebastian Richter; Florian Schoenberger; Shireen Geimer; Keith Paulsen
Journal:  Sensors (Basel)       Date:  2020-09-22       Impact factor: 3.576