High-precision measurements of δ2H, δ18O and δ17O in water with the aid of cavity ring-down laser spectroscopy.

Anna Pierchala1, Kazimierz Rozanski1, Marek Dulinski1, Zbigniew Gorczyca1, Michal Marzec1, Robert Czub1.   

Abstract

A thorough evaluation of measurement uncertainty together with control of short-term and long-term precision of measurements should be a basis of any successful quality assurance/quality control (QA/QC) strategy aimed at maintaining a high quality of the analytical process. Here we present the results of a comprehensive assessment of the analytical performance of a Picarro L2140-i CRDS laser spectrometer analysing δ2H, δ18O and δ17O in water. The assessment is based on results obtained during 15 months of continuous operation of this instrument (February 2017 to May 2018). The short-term precision of measured and derived quantities was 0.11, 0.036, 0.028, 0.23 ‰ and 11 per meg, for δ2H, δ18O, δ17O, d-excess and Δ17O, respectively, and is comparable to the precision reported by the manufacturer. The long-term precision of the L2140-i, defined as standard uncertainty of the time series of 153 analyses of a laboratory standard conducted throughout 15 months, was roughly two times lower (0.24, 0.053, 0.038, 0.37 ‰ and 21 per meg, for δ2H, δ18O, δ17O, d-excess and Δ17O). In-depth assessment of the measurement uncertainty of a single analysis revealed that assigned uncertainty of the calibration standards is an important component of the uncertainty budget, especially in case of δ2H analysis.

Entities:  

Keywords:  Cavity ring-down spectroscopy; QA/QC; hydrogen-2; isotope measurements; methods and equipment; oxygen-17; oxygen-17 excess; oxygen-18; uncertainty

Mesh:

Substances:

Year:  2019        PMID: 31037964     DOI: 10.1080/10256016.2019.1609959

Source DB:  PubMed          Journal:  Isotopes Environ Health Stud        ISSN: 1025-6016            Impact factor:   1.675


  1 in total

1.  Faster and more precise isotopic water analysis of discrete samples by predicting the repetitions' asymptote instead of averaging last values.

Authors:  Nico Hachgenei; Véronique Vaury; Guillaume Nord; Lorenzo Spadini; Céline Duwig
Journal:  MethodsX       Date:  2022-03-03
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.