Literature DB >> 31037539

A two-centre validation study of sterility test of corneal storage media with elimination of interfering antimicrobials in compliance with the European Pharmacopoeia.

R Vignola1, R Mistò2, L Giurgola3, F Pateri2, A Pocobelli1, J D'Amato Tóthová4.   

Abstract

The purpose of this study was to validate the sterility test of corneal culture and deswelling/transport media using a device for removal of antimicrobials before incubation in BACTEC™ automated system in two Italian Eye Banks. Corneal culture medium, TISSUE-C, and deswelling/transport medium, CARRY-C, were inoculated with 10-100 cfu of six European Pharmacopoeia (EP) reference strains and either treated with medical device RESEP for removal of antimicrobials (RESEP+ group) or left untreated (RESEP- group) before injection into the BACTEC Plus bottles. The same steps were repeated in the absence of inocula with tryptone soy broth samples as negative controls, and the inocula were also directly injected in the BACTEC™ bottles as growth controls. All the samples were incubated in BACTEC™ automated system for 7 days, and the time to detection of microbial growth was recorded automatically. At both the Eye Banks, in the RESEP+ groups, microbial growth was detected in 100% of samples. In the RESEP- group, the method sensitivity ranged from 66.7 ± 21.1 to 88.9 ± 6.4% for TISSUE-C samples while for CARRY-C samples the method sensitivity ranged from 94.5 ± 5.1 to 100%. The method specificity corresponded to 100% for all the groups at both Eye Banks. This two-centre validation study showed that the use of RESEP increased the sensitivity of sterility test using BACTEC™ automated system up to 100% and, consequently, allowed validation of the method for sterility testing of corneal storage and deswelling/transport media according to the EP requirements. The test could not be validated without the use of RESEP.

Entities:  

Keywords:  Antimicrobials removal neutralization; Corneal culture media; Corneal storage and transplantation; European Pharmacopoeia sterility test; Eye banking; Microbiological testing

Year:  2019        PMID: 31037539     DOI: 10.1007/s10561-019-09766-7

Source DB:  PubMed          Journal:  Cell Tissue Bank        ISSN: 1389-9333            Impact factor:   1.522


  1 in total

1.  Killing efficacy of a new hypothermic corneal storage medium against the micro-organisms frequently found in human donor cornea intended for transplantation.

Authors:  Laura Giurgola; Claudio Gatto; Claudia Honisch; Orietta Rossi; Eugenio Ragazzi; Jana D'Amato Tothova
Journal:  BMJ Open Ophthalmol       Date:  2021-12-17
  1 in total