Filling accuracy and imprecision of commercial evacuated sodium citrate coagulation tubes.

Giuseppe Lippi1, Gian Luca Salvagno1, Vanja Radišić Biljak2, Ana-Katarina Kralj2, Ivona Kuktić2, Matteo Gelati1, Ana-Maria Šimundić2.   

Abstract

Current recommendations advocate that blood tubes for coagulation testing should be filled not less than 90% of their nominal filling volume, since under- or over-filling >10% may generate unreliable results of some hemostasis assays. This study was hence aimed to explore filling accuracy and precision of commercial blood tubes. Between-lot variations of 3 different lots (20 tubes per lot) of 3.2% citrate blood tubes manufactured by Becton Dickinson, Greiner and Kima were studied. One additional lot from each manufacturer was assessed in triplicate (three series of 20 tubes), to assess within-lot variation. All tubes were first weighed empty and then filled with distilled water by a syringe, under ideal filling conditions. Filled tubes were weighed again, in duplicate. For each 20 tubes series, mean bias (deviation from the ideal tube filling volume) and imprecision (coefficient of variation; CV%) were calculated. All biases were within ±10%. Within-lot and between-lot variation in filling volume was acceptable, and comprised between 0.4 and 2.4%. Greiner tubes were the most accurate (bias, -1.0 to 2.4%), followed by Kima (bias, -7.8 to -5.9%) and Becton Dickinson (bias, -9.6 to 3.3%) tubes. The highest between-lot difference was noted for Becton Dickinson tubes (up to 12.9%), followed by Greiner and Kima tubes (up to 3.4 and 1.8%, respectively). Although coagulation tubes filling accuracy was within ±10% for all three tested manufacturers, the overall bias was found to be variable among manufacturers and lots. Major effort shall be made by blood tube manufacturers for improving standardization of their products.

Entities:  

Keywords:  Quality; blood collection; blood tubes ; coagulation; errors

Mesh:

Substances:

Year:  2019        PMID: 31035815     DOI: 10.1080/00365513.2019.1609696

Source DB:  PubMed          Journal:  Scand J Clin Lab Invest        ISSN: 0036-5513            Impact factor:   1.713


  1 in total

1.  Breakfast can Affect Routine Hematology and Coagulation Laboratory Testing: An Evaluation on Behalf of COLABIOCLI WG-PRE-LATAM.

Authors:  Maria Elena Arredondo; Eduardo Aranda; Rubén Astorga; Lorena Michele Brennan-Bourdon; Marise Danielle Campelo; Silvia Flores; Claudio Medel; Ignacio Manríquez; Patricia Ochoa; Beatriz Varela; Carlos Vega Salinas; Gabriel Lima-Oliveira
Journal:  TH Open       Date:  2019-12-17
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.