Cherenkov emission-based external radiotherapy dosimetry: I. Formalism and feasibility.

Yana Zlateva1, Bryan R Muir2, Issam El Naqa3, Jan P Seuntjens1.   

Abstract

PURPOSE: Cherenkov emission (CE)-based external beam dosimetry is envisioned to involve the detection of CE directly in water with placement of a high-resolution detector out of the field, avoiding perturbations encountered with traditional dosimeters. In this work, we lay out the groundwork for its implementation in the clinic and motivate CE-based dosimeter design efforts. To that end, we examine a formalism for broad-beam in-water CE-based dosimetry of external radiotherapy beams, design and test a Monte Carlo (MC) simulation framework for the calculation of CE-to-dose conversion factors used by the formalism, and demonstrate the experimental feasibility of this method.
METHODS: The formalism is conceptually analogous to ionization-based dosimetry and employs CE-to-dose conversion factors, k C θ ± δ θ , including only and all CE generated within polar angles θ ± δθ on beam axis. The EGSnrc user code SPRRZnrc is modified to calculate k C θ ± δ θ , as well as CE spectral and angular distributions. The modified code is tested with monoenergetic parallel electrons on a thin water slab. Detector configurations are examined for broad 6-22 MeV electron beams from a BEAMnrc TrueBeam model, with a focus on θ ± δ θ = 90 ∘ ± 90 ∘ (4π detection), 90 ∘ ± 5 ∘ , and 42 ∘ ± 5 ∘ ( θ = 42 ∘ is the CE angle of relativistic electrons in water). We perform a relative experimental validation at 90 ∘ with electron beams, using a simple detector design with spherical optics and geometrical optics approximation of the sensitive volume, which spans the water tank. Due to transient charged particle equilibrium, broad photon beams are generally less sensitive to beam quality, depth, and angle.
RESULTS: For 0.1-50 MeV electrons on a thin water slab, the code outputs CE photon spectral density per unit mass (calculated from dose and k C θ ± δ θ ) and angle in agreement with theory within ±0.03% and ± 0 . 01 ∘ , respectively, corresponding to the output precision. The 42 ∘ configuration was found impractical due to detection considerations. Detection at 90 ∘ ± δ θ for small δθ exhibited beam quality dependence of the same order as well as strong superficial depth dependence. A 4π configuration ameliorates these effects. A more practical approach may employ a large numerical aperture. In comparing with literature, we find that these effects are less pronounced for broad photon beams in water, as expected. Measured relative k C 90 ∘ ± δ θ at small δθ were within 1% of simulated factors (relative to their local average) for percent-depth CE (PDC) >50%. At other depths, deviations were in accordance with signal-to-noise, known detector limitations, and approximations. It was found that the CE spectrum is beam quality and depth invariant, while for electron beams the CE angular distribution is strongly dependent on beam quality and depth. However, the uncertainty of CE and PDC measurement at 90 ∘ ± δ θ detection for small δθ due to ± 0 . 1 ∘ deviations around δθ was shown to be ≤1% and <0.1% (k = 1), respectively. The robustness to expected detector setup variations was found to result in ≤1% (k = 1) local uncertainty contribution for PDC >50%.
CONCLUSIONS: Based on our MC and experimental studies, we conclude that the CE-based method is promising for high-resolution, perturbation-free, three-dimensional dosimetry in water, with specific applications contingent on comprehensive detector development and characterization.
© 2019 American Association of Physicists in Medicine.

Entities:  

Keywords:  Cerenkov; Cherenkov; Monte Carlo; dosimetry formalism; experimental validation; external radiotherapy beams

Mesh:

Year:  2019        PMID: 31034637     DOI: 10.1002/mp.13414

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  3 in total

1.  Direct in-water radiation dose measurements using Cherenkov emission corrected signals from polarization imaging for a clinical radiotherapy application.

Authors:  Émily Cloutier; Luc Beaulieu; Louis Archambault
Journal:  Sci Rep       Date:  2022-06-10       Impact factor: 4.996

2.  Theoretical lateral and axial sensitivity limits and choices of molecular reporters for Cherenkov-excited luminescence in tissue during x-ray beam scanning.

Authors:  Ethan P M LaRochelle; Brian W Pogue
Journal:  J Biomed Opt       Date:  2020-11       Impact factor: 3.170

3.  Step-size effect on calculated photon and electron beam Cherenkov-to-dose conversion factors.

Authors:  Yana Zlateva; Bryan R Muir; Issam El Naqa; Jan Seuntjens
Journal:  Phys Med       Date:  2020-09-08       Impact factor: 2.685