C F Garcia, L M Weiss, J Lowder, C Komoroske, M P Link, R Levy, R A Warnke.
Abstract
A quantitation method for lymphocyte subsets in immunoperoxidase-stained frozen tissue sections was compared with flow cytometry in 23 cases of non-Hodgkin's lymphoma. Close correlations were obtained, demonstrating the accuracy of the technic. Weak intensity of fluorescence and fragility of the tumor cells during the fluorescence-activated cell sorter (FACS) analyses were the most likely explanations for a number of the discrepancies observed. The tissue quantitation method was precise, particularly at low values, where it was better than the FACS. A simpler and faster estimation method employing categories within 10 percentage units was also tested in this study; this method correlated as well with the FACS as the quantitation method and gave the best interobserver correlations.Entities:
Mesh:
Substances:
Year: 1987 PMID: 3103419 DOI: 10.1093/ajcp/87.4.470
Source DB: PubMed Journal: Am J Clin Pathol ISSN: 0002-9173 Impact factor: 2.493