| Literature DB >> 31031034 |
Toru Kameda1, Naohisa Kamiyama2, Hideo Kobayashi3, Yuko Kanayama2, Nobuyuki Taniguchi4.
Abstract
We evaluated the influence of settings on an ultrasound machine on the configuration of a single B-line in a healthy model and analyzed the frequency spectrum. We also devised simple experimental models that generated B-line-like artifacts and evaluated the influence of the machine settings on the configuration. Visualization of B-lines was affected by the spatial compound imaging, the focal zone and the frequency. The spectra of both the B-line and non-B-line region at the same depth had the same center frequency and bandwidth. B-line-like artifact was generated by a spindle-shaped juice sac of a mandarin orange, an edible string-shaped glucomannan gel, glass beads and glass plates. Visualization of B-line-like artifacts was also affected by these machine settings. Our study indicated that the physical basis of some B-lines is multiple reverberations. B-line-like artifacts provide clues for solving key issues, such as the physical basis of B-lines, the sonographic-pathologic correlation in B-lines and the effects of machine settings.Entities:
Keywords: B-line; B-line–like artifact; Experimental model; Lung ultrasound; Machine setting; Reverberation
Mesh:
Year: 2019 PMID: 31031034 DOI: 10.1016/j.ultrasmedbio.2019.03.003
Source DB: PubMed Journal: Ultrasound Med Biol ISSN: 0301-5629 Impact factor: 2.998